Cautions/Warnings 

1.

 

Power (ALWAYS check the Chiller Parameter tag and verify operational voltage before plugging in chiller) 

a.

 

Only connect North Slope Chillers to a properly grounded circuit 

b.

 

Never connect your North Slope Chiller to voltage that is outside the acceptable voltage range. 
Connecting your chiller to any voltage outside of these ranges will damage the compressor. 

i.

 

110V model will operate on voltages between 100 ~ 130V 

ii.

 

230V model will operate on voltages between 200 ~ 250V 

c.

 

Only use power sources that are 60Hz, running the chiller at 50Hz will void the warranty and will 
damage the compressor 

2.

 

Never run the pump dry, only turn on chiller after the fluid reservoir has been filled. Running the chiller without 
fluid will permanently damage the pump and void the warranty 

3.

 

This chiller is air cooled and must have good ventilation. For proper function there must be a minimum of 1 foot 
clearance on both sides and 2 foot clearance at the back of the chiller 

 

 

 

Содержание NSC0250-FROST

Страница 1: ...tage outside of these ranges will damage the compressor i 110V model will operate on voltages between 100 130V ii 230V model will operate on voltages between 200 250V c Only use power sources that are...

Страница 2: ...3000 BTU hr 878 kW 90F ambient and 45F fluid Refrigerant R 410a Refrigerant Charge 77 lbs Precision 0 5 F 0 3 C Metering Device Capillary Protection Overcurrent protection for compressor flow alarm a...

Страница 3: ...Model NSC0250 FROST Owners Manual 3 P a g e www northslopechillers com 866 826 2993 3130 S 1030 W Suite 1 Salt Lake City UT 84119 Chiller overview FRONT BACK...

Страница 4: ...S 1030 W Suite 1 Salt Lake City UT 84119 Installation 1 Open package to check for damage and to verify that all components are there 2 Fill the reservoir 3 Connect inlet and outlet hoses 4 Plug in Pow...

Страница 5: ...by pressing any button but the code will remain until the alarm condition is fixed Check ambient temperature 1 Press button to display ambient temperature 2 D1 indicator light will flash while ambient...

Страница 6: ...ng parameter changes Parameter List Parameter Description Range Factory Default Notes Constant Temperature Mode Intelligent Mode F0 Temperature Setting F8 F9 45 F Constant Temperature F1 Temperature D...

Страница 7: ...the operation and performance of the Products themselves Only the Manufacturer is authorized to make any warranty or representation and the customer may not rely on any other warranty or representati...