Mobile injector unit E 380

Miele Company Ltd., Fairacres, Marcham Road, Abingdon, Oxon, OX14 1TW, Tel (01235) 554455,
Fax (01235) 554477, Internet: http://www.miele.co.uk /

Alteration rights reserved / 2104

^

Use the spanner to secure the jets into the
positions required. The overview above gives an
example of possible positions.  Start from the
centre and work your way outwards in ever
increasing circles.

a

E 351 jets, 4.0 x 160 mm

b

E 352 jets, 6.0 x 220 mm

^

Screw the jet for flushing out the powder dispenser
in positon

c

and secure it to the spray arm with a

cross-tip screwdriver.

The mobile unit is supplied with two sizes of clamps,
holders and jets. Make sure they are put together in
the correct order.

^

Push the clamp into the holder as shown above.

^

Then squeeze the clamp together and feed the
injector jet through the holes in the clamp.

Items supplied:

1 water feed pipe

14 x E 351 jets, 4.0 x 160 mm

1 seal

14 x E 353 clips for 4.0 x 160 mm jets, consisting of clamp

1 mobile unit E 380

18 x E 352 jets, 6.0 x 220 mm

1 jet for flushing out powder dispenser

18 x E 354 clips for 6.0 x 220 mm jets, consisting of clamp

,

Carefully read the Warning and Safety instructions in the Operating Instructions for your washer-disinfector as well as any instructions provided by the glassware

manufacturer.

Any broken glassware can cause serious injury.

All glassware should be properly inspected and tested for cleanliness. Do not rely on a visual check alone.

Do not let the individual pieces of
glassware touch each other.

Place the laboratory glassware over the jets:

Push the clips upwards, if necessary, to
prevent the tip of the jet from contacting
the bottom of the glassware.

Glassware must not sit higher than the
top of the water feed pipe (with the wa-
ter feed adapter pressed downwards).

Glassware must not extend beyond the
sides of the mobile unit.

^

Fit the seal in the holder in the mobile unit.

^

Then fit the water feed pipe into the holder. When
doing so make sure that the groove in the water
feed pipe lines up with the notch in the holder. The
hot air drying connection points to the back of the
mobile unit.

^

Tighten the nut by hand.

Installation:

Tools required:

Spanner SW 9, cross-tip screwdriver

Areas of application:

To ensure adequate water pressure all positions need to be equipped with jets.

– The mobile injector unit can hold up to 32 pieces of laboratory glassware.

– Different volume glassware can be processed, however, the maximum volume per piece is 1 litre.

– To process pipettes special sleeves are needed, MIBO E 336. They are available to order.

– Laboratory glassware, which because of its design cannot be flushed through properly, will not be

sufficiently cleaned and disinfected in this mobile unit.

– This mobile unit should only be used for processing laboratory glassware which has been declared as

suitable for machine processing by the glassware manufacturer.

The jet for flushing out the powder dispenser is designed to rinse out the powder dispenser in the machine
door after use (if applicable to the machine the unit is used in). This jet must therefore not be used for
cleaning glassware.