background image

MagCore®

 Plus II 

Operation  Manual

CONTENTS

Safety Precautions                                           

1

Intended use                                                   

2

IVD Symbols Reference                                   

4

Specification                                                     

5

━ Operating Parameters   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

5

━ Operating Environment   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

5

━ Applications  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

5

Accessories                                                      

6

Installation                                                       

6

━ Before Installation   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

6

━ System Overview   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

7

━ Install Reagent Cartridge  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

8

━ Install Tube, Tip  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

8

━ Installation  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

9

━ Installation Settings And Test  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

9

━ Installation  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   . 

10

Touch Panel and User Interface                       

12

Operation Before Starting the Instrument  .   .   .   .   .   .   .   .   .   .   . 

13

━ Emergency Stop  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   . 

14

━ Start Programs  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   . 

15

━ Saved Settings  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   . 

16

━ Access to Saved Settings  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   . 

16

━ Set and Select Users  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   . 

17

━ Set User  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   . 

17

━ Select User Name  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   . 

17

━ Scan Sample / Elution Tube  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   . 

18

━ Scan Kit Lot  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   . 

18

━ Pause and Abort  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   . 

19

━ Set Up the System Time   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

19

━ Set the Instrument Nickname  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   . 

20

━ Network Connection Settings  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   . 

20

━ LIMS Data Collection   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

21

1 . Get Data from USB Flash Drive   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

21

2 . Access to information by Network Connection  .   .   .   .   .   . 

22

━ Remote monitoring APP  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   . 

23

1 . Download the APP   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

23

2 . Registration  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

23

3 . Log in the APP and Pair the Instrument  .   .   .   .   .   .   .   .   .   .

23

4 . APP interfaces  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

24

Easy Maintenance                                        

27

━ Replace O-Ring   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

27

━ Piston Test  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

27

━ Clean Piercing Needles  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

30

━ Eject Tip  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

30

━ Heater Test   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

31

━ UV Sterilization          .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

31

━ Alarm History  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

32

━ System Status  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

32

━ Program Update   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

33

━ Clean the Waste Box   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

33

━ Replace Fuse  .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .

34

Document Revision History                         

36

15F, No.15, Qiaohe Rd,Zhonghe Dist, New Taipei City 235029, Taiwan

September 2022

EN-MagCore® Plus II  

Operation Manual-1.0.0

Summary of Contents for Plus II

Page 1: ...ser 17 SelectUserName 17 ScanSample ElutionTube 18 ScanKitLot 18 PauseandAbort 19 SetUptheSystemTime 19 SettheInstrumentNickname 20 NetworkConnectionSettings 20 LIMSDataCollection 21 1 GetDatafromUSBF...

Page 2: ...0 Power Sources This product should be operated only from the type of power source indicated on the label If you are not sure of the type of power supply at your lab consultthesellerorlocalpowercompan...

Page 3: ...Cbeforeuse Ifyouusethereagentsattemperaturesoutsidethe recommendedrange thekitmaynotworkwell DocumentthelotnumberofthekitincaseofcomplaintsorquestionsforRBCBioscienceTechnicalService regardinganycompo...

Page 4: ...oesnotwarrantthatsamplestreatedwithLysisBufferarecompletelyinactivatedand noninfectious Aftersampleprocessingiscompleted removeandautoclavealldisposableplastics Donoteat drinkorsmokeinthelaboratorywor...

Page 5: ...ng electricity Instructionmanualmustbereadbeforeoperating Consultinstructionsforuse Alternatecurrentinput Use byDate Nouseperiod Temperaturestorageconditions15 C 30 C Cataloguenumber productcode Lotnu...

Page 6: ...NA 200 l 400 l 1 200 l ViralNucleicAcid DNA RNA200 l 400 l 1 200 l FFPEGenomicDNA FFPETotalRNA ForensicDNA PlasmaandCirculatingCell FreeDNA 1 200 l 4 000 l TissueGenomicDNA PlantGenomic DNA BacterialD...

Page 7: ...tyCartridgeX10 8 SampleTube ElutionTubeX20 9 PipetteTipX20 10 WasteBoxX1 11 BarcodeScannerX1 Please refer to the machine size shown in the above figure to prepare an appropriate position for the machi...

Page 8: ...7 www rbcbioscience com System Overview TouchPanel LogoIndicator FrontDoor BackServiceLid OperationArea WasteBox ScannerPort USBPort MainPowerSwitch EthernetPort PowerOutlet...

Page 9: ...utT Rackoncartridgerack theninstalland pushcartridgeracktotheinnermostgently PleaseinstalltheTipandTubeaccordingto theinstructionsofextractionkitusermanual Pullcartridgerackoutward towarduser takeitou...

Page 10: ...mentNickname GotoPage17 Step3 NetworkConnectionSettings GotoPage17 Step4 TestPiston PleaserefertoPistonTestchapterforinformation abouttheoperation Step5 TestHeater PleaserefertoHeaterTestchapterforinf...

Page 11: ...OpentheFrontDoor ForRestrictedPers Installation A B AccessoryBox Cushioning Material Removecushioningmaterial andtakeoutaccessorybox R p 1 1 2 2 AccessoryBox Romovecushioningmaterial Takeoutaccessoryb...

Page 12: ...letiesthen pulloutthetouchpanel Removethetwobrackets 1 1 Closethefrontdoor 1 Switchonthepower themachinewillinitializeautomatically 2 Movetherackmoduleoutwardmanually thencuto andremovethetwocableties...

Page 13: ...edProtocols Maintenance EjectTip ExecuteExtraction Executequickrunwith selectionofpre saved protocols SystemMaintenance Functions Overviewofthelatest runswithbarcodes SystemSettings StatusColumnIcons...

Page 14: ...rument Afterloadingthereagents cartridgesandtips seepage5andsettherequiredparameterswithout openingthedoor Afterclosingthedoor pressSTARTtostartoperation PlungedUpwardfrom theLowerEdgeofthe FrontDoor...

Page 15: ...rectly stop the machine whenanaccidenttakesplace Whenanemergencysituationoccursandthemachineis operating pleasepressthestopbuttonoropenthefront doortostopthemachineimmediately Program Code 201 Pause C...

Page 16: ...tothemachine PressStarttogotothenextstep ToSelectUserandScanBarcodes SeePage12 PressNexttochoosethecodenumber Pleaseconfirmthereagentsetcodenumbertobe usedintheextraction Selectthenumberofthecartridge...

Page 17: ...ocolspage Selecttheparametersthatyouwanttosaveshown inthebluearea SavedProtocoloptionsavesupto sixrecords PressSaveProtocoltosavethecustomized parameters PressBacktoconfirmyoursettings Afterloadingthe...

Page 18: ...ckasshowninthe figure PressEnterwhenfinishediting Pressthenameblockyouwanttoselect Afterdisplayingoftheselectedname pressDone tofinish Setand SelectUsers SetUser SelectUserName Select User Select Scan...

Page 19: ...lution Tube Scan Kit LOT Back Next Jemes 01 31 201913 15 47 Select User Select Scan Sample Scan Elution Tube Scan Kit LOT Back Next Jemes 01 31 201913 15 47 SelectScanSampleorScanElutionTube Thesystem...

Page 20: ...ort Youcanpauseorabortthecurrentextraction whiletheprogramisinprogress PressthePause buttontopauseatanytime andthemachinewill immediatelysuspendoperation Thisbuttonis usedasanemergencystopbutton After...

Page 21: ...ortheLIMSdatatransmission pleasedirectly followthebelowstepstosetupthenetwork connection 1 Ifyouhavesetupalocalareanetworkinyour workarea pleasedirectlyinsertthenetworkroute intothenetworkportoftheins...

Page 22: ...ingLIMSdatatransfer youneedtofirst establisha21 in connectionwiththeIT stafftodeterminethefirewallofyourPC When transmittingasaLIMSdata weoperateonthe internalnetwork PleaseconfirmthedisplayedTCP IPlo...

Page 23: ...sofdata transmission YoucanuseanyFTPclientsoftware GettheinformationFTPlocationinformationisas follows Ftp instrumentIPlocation eg192 168 3 155 OpentheInternetbrowserandentertheFTP locationandloginwit...

Page 24: ...P 2 Registration ToregistertheAPP pleasecontactyourdistributor toobtaincustomerverificationQRcode verificationkey andinstrumentQRcode then followthebelowstepstocompleteregistration andpairingtheinstru...

Page 25: ...24 MagCore PlusII 4 APPinterfaces A Clicktheinstrumenttocheckthecurrentstatus Instrumenttype Serialnumber Start Endtime Selectedparameters Extractionstep pause abort alarmmessages...

Page 26: ...ddAdministratorUser AllowotheruserstousetheAPPwithoutre registeringtheiraccount LoginaccountmustbeAdmininordertochangeormanageusers Click 1 Click tomoveandsortthepositionoftheinstrument 2 Click toun p...

Page 27: ...26 MagCore PlusII SelectUserManagement Complete Entertheaccount e mailaccountissuggested Enterthepassword Enterthee mailaccount WhethertheaccountistheAdmin Whethertheaccountissuspended...

Page 28: ...secheckups periodicallytoensurethesystemstatusisalways good PleaseSelectReplaceO Ringbutton 1 PleasedisassemblethedamagedO Ring 2 ReplaceanewO Ring 3 GreasingtheO Ring andperformthePiston Test ThePist...

Page 29: ...pleaseresettheTipHolderandtryitagain 5 Pleasecheckwhethertheliquidlevelsofalltips areconsistentwitheachother Iftheliquidlevels ofalltipsarenotconsistentwitheachother pleasechecktheO Ring Pleaserefert...

Page 30: ...ofT Rack 2 Pleaseadd1mlwatertotheSampleTubeand putitinW1oftheT Rack 3 PleaseselecttheTestW4Tip 4 Pleasecheckwhethertheliquidlevelsinalltips areconsistentwitheachother Iftheliquidlevels arenotconsisten...

Page 31: ...rayedwith70 alcohol 2 PleasecleanthePiercingNeedleswiththecloth Warning ThetipsofthePiercingNeedles areextremelysharp Pleasebe carefulwhenyoucleanthem Ifthesystemisstoppedfrompowershortageor otherabno...

Page 32: ...mperatureshownon thethermometerisconsistentwithwhat sshown onthescreen Theacceptabletemperaturerangeis65 5 Cor 90 5 C PleaseSelecttheSterilizationButton PleaseselectthesterilizationtimeandpressStart b...

Page 33: ...ns aftertheiconofUSBFlashDrivedisappears PleasesavethedataintheUSBFlashDriveand transfertocomputer orthedataintheUSBFlash Drivemaybeoverwrittenlater FilePath USBFlashDrive AlarmLog datalog CSV FormatU...

Page 34: ...buydisposableWasteBoxpaperbags BinBags YoucanusetheUSBtoupdatethesystem Pleasecopythedownloadedprogramtotheroot directoryoftheUSBFlashDrive Notprovided TheninserttheUSBFlashDriveintotheUSBPort ofthema...

Page 35: ...34 MagCore PlusII ReplaceFuse 250V10A Please use the specified type of fuse tominimizetheriskofburning...

Page 36: ...35 www rbcbioscience com AutomatedNucleicAcidExtractor Model PlusII InputRaitng 100 240V 50 60Hz Max 4 5AMax 450W RBCBioscienceCorp 15F No 15 QiaoheRd ZhongheDist NewTaipeiCity235029 Taiwan...

Page 37: ...onHistory Date Changes September 2022 EN MagCore Plus II Operation Manual 1 0 0 Updated the following sections General Considerations Document Revision History Update to comply with the requirements o...

Page 38: ...15F No 15 QiaoheRd ZhongheDist NewTaipeiCity235029 Taiwan TEL 886289121200FAX 86289121300...

Reviews: