48NDR

P. 1 / 8

EM-9443 Rev.5

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN

Phone: +81(6)6659-8201   Fax: +81(6)6659-8510   E-mail: [email protected]

BARGRAPH INDICATING ALARM

(with 4-digit digital meter; RTD input)

MODEL

48NDR

INSTRUCTION MANUAL

BEFORE USE ....

Thank you for choosing M-System.  Before use, please check 
contents of the package you received as outlined below.
If you have any problems or questions with the product, 
please contact M-System’s Sales Office or representatives.

 PACKAGE INCLUDES:

  Bargraph indicator (incl. scale plate) .......................... (1)
  Mounting bracket ......................................................... (2) 
 Bezel*

1

 ........................................................................... (2)

  Watertight packing ....................................................... (1)
*1. The size of included bezel differs depending on option code.

 MODEL NO.

Confirm that the model number described on the product is 
exactly what you ordered.

 INSTRUCTION MANUAL

This manual describes necessary points of caution when 
you use this product, including installation, connection and 
basic maintenance procedures.

POINTS OF CAUTION

 CONFORMITY WITH EC DIRECTIVES

• This equipment is suitable for Pollution Degree 2 and  In

-

stallation Category II.  Reinforced insulation (signal input 

to output to power input: 300V) is maintained.  Prior to 

installation, check that the insulation class of this unit 
satisfies the system requirements.

• The equipment must be mounted on a panel surface.  Once 

mounted on a panel, take appropriate precautions to pre-
vent operators to be exposed to the terminal block.

• Altitude up to 2000 meters
• Insert a noise filter for the power source connected to the 

unit.   TDK-Lambda  Noise  Filter  Model  MZS-1206-33  or 

equivalent is recommended.

• The  equipment  must  be  installed  such  that  appropriate 

clearance and creepage distances are maintained to con-

form  to  CE  requirements.    Failure  to  observe  these  re

-

quirements may invalidate the CE conformance.

 POWER INPUT RATING & OPERATIONAL RANGE

• Locate the power input rating marked on the product and 

confirm its operational range as indicated below:

  AC rating:   85 – 264V, 50/60 Hz, approx. 5.5 – 8VA
  DC rating:   24V ±15%, approx. 3.5W

 ENVIRONMENT

• Indoor use
• If the unit’s environmental protection is compromised (e.g. 

when multiple units are to be mounted side by side) when 
heavy dust or metal particles are present in the air, install 
them inside an enclosure with a proper ventilation.

• Do not install the unit where it is subjected to continuous 

vibration.  Do not subject the unit to physical impact.

• Environmental temperature must be within -10 to +55°C     

(14 to 131°F) with relative humidity within 30 to 90% RH 

in order to ensure adequate life span and operation.

 REQUIREMENTS TO ENSURE IP65

• Observe the designated panel cutout size (31.5 

×

 138 mm).

• Single mounting only.  IP65 is not ensured when the units 

are clustered side by side.

• The watertight packing included in the product package 

must be placed behind the front cover.

• The mounting bracket must be fastened tightly until they 

hit the panel.

• Confirm visually that the packing is not contorted or ex

-

cessively run off the edge after installation.

 WIRING

• Do not install cables (power supply, signal input and out

-

put) close to noise sources (high frequency line, etc.).

• Do  not  bind  these  cables  together  with  those  in  which 

noises are present.  Do not install them in the same duct.

 TERMINAL BLOCK

• The terminal block is separable in two pieces.  Loosen two 

screws on top and bottom of the terminal block to sepa-
rate.

• Be  sure  to  turn  off  the  power  supply,  input  signal  and 

power supply to the output relays before separating the 
terminal block.

Terminal Block

Separable

Terminal Block Screw
(top & bottom)

 AND ....

• The unit is designed to function as soon as power is sup

-

plied, however, a warm up for 10 minutes is required for 
satisfying complete performance described in the data 
sheet.

Summary of Contents for 48NDR

Page 1: ...invalidate the CE conformance POWER INPUT RATING OPERATIONAL RANGE Locate the power input rating marked on the product and confirm its operational range as indicated below AC rating 85 264V 50 60 Hz approx 5 5 8VA DC rating 24V 15 approx 3 5W ENVIRONMENT Indoor use If the unit s environmental protection is compromised e g when multiple units are to be mounted side by side when heavy dust or metal ...

Page 2: ...ction 138 1 0 L Panel thickness 1 6 8 0 mm L 31 5 36 N 1 N number of units 1 0 CLUSTERED MOUNTING no ingress protection 1 0 USING BEZELS Bezels are used when the unit needs to be adapted to an ex isting panel cutout It is not required when the panel cutout size is as specified in the unit s data sheet 31 5 138 mm Three types of bezels are selectable when ordering Standard type Fits into M System s...

Page 3: ...ze 31 5 138 mm Panel Watertight Packing 2 Push the mounting brackets into the grooves on top and bottom of the rear module until they hit the panel s rear side Leave at least 3 centimeters both above and below the unit for ventilation EXTERNAL DIMENSIONS unit mm inch 75 2 95 103 4 06 10 39 18 71 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 16 M3 SCREW TERMINALS 2 M3 SCREWS 144 5 67 137 5 5 41 4 1 16 36 ...

Page 4: ...xisting panel cutout of Fuji Electric PAJ PAK PBA 44 138 mm etc STANDARD BEZEL 2 10 4 41 7 5 30 3 9 15 136 5 5 37 144 5 67 11 4 45 7 5 30 4 9 19 136 5 5 37 144 5 67 8 5 33 10 4 41 7 5 30 136 5 5 37 144 5 67 OPTION F BEZEL 4 12 13 14 15 16 U V La Lb Lc 12 13 14 15 16 U V LLa 11 La LLb Lc LLc LOAD Inductive Load Coil Varistor or CR Circuit DC Powered LOAD Inductive Load Coil Diode Varistor or CR Cir...

Page 5: ...de 1 orange orange green orange Input L L Input H H Input green red H L Pattern 2 model suffix code 2 orange green green red red red H L 50 40 30 20 10 0 10 20 30 40 50 C C 100 SET H S P Z L 90 80 70 60 50 40 30 20 10 0 SET ENTER Button H Setpoint Adj Trip Indicator red Span Adj amber Decimal Point Position Adj amber Zero Adj amber L Setpoint Adj Trip Indicator green UP Button Mode Selector DOWN B...

Page 6: ... Delay for L Setpoint Increase Decrease Complete Digital Indicator SET L LL M Press at once ON Time Delay for LL Setpoint Increase Decrease Complete Digital Indicator SET LL Z M Press at once Zero Adjustment Digital Indicator Bargraph Z SET Z M Increase Decrease Zero Adjustment Digital Indicator Only Complete Z ON LED OFF SPAN ADJUSTMENT Apply 100 input signal before adjustment All alarm setpoints...

Page 7: ...ue and press SET DIGITAL INDICATOR ZERO ADJUSTMENT 1 Apply 0 input signal 2 Hold down M for 3 seconds and push M to choose Z 3 Press UP or DOWN buttons until the digital indicator comes to a desired value and press SET BARGRAPH DIGITAL INDICATOR SPAN ADJUSTMENT 1 Apply 100 input signal 2 Hold down M for 3 seconds and push M to choose S 3 Press UP and DOWN buttons at once Blinking S LED is switched...

Page 8: ...ing on the digital indicator 3 Press SET and DOWN buttons at once The M button remain activated for 1 minute after it has been used to choose a indicator LED programming mode During this period pressing the M button briefly is enough to choose another indicator After 1 minute it turns to de activated state CALIBRATION PROCEDURE This unit is calibrated at the factory to meet the ordered specificati...