background image

Manual - Valiant 2 CPET /

XL

Cardio Pulmonary

Lode BV
Zernikepark 16
NL 9747 AN
Groningen
The Netherlands
t: +31 50 571 28 11
f: +31 50 571 67 46
@: [email protected]
www.lode.nl

M938900/938905_MANUAL VALIANT 2 CPET (XL) / VERSION: V1.1 - 2016-08-25

Summary of Contents for 938900

Page 1: ...nt 2 CPET XL Cardio Pulmonary Lode BV Zernikepark 16 NL 9747 AN Groningen The Netherlands t 31 50 571 28 11 f 31 50 571 67 46 ask lode nl www lode nl M938900 938905_MANUAL VALIANT 2 CPET XL VERSION V1 1 2016 08 25 ...

Page 2: ...Validity 6 Install your treadmill 7 Maintenance 7 1 Emergency stop installation 8 Troubleshooting 8 1 Belt alignment 9 Cleaning 9 1 Cleaning 10 Control with the Lode Ergometry Manager LEM 11 Control with an External Device 12 List of symbols used 13 Electromagnetic Compatibility EMC 14 Specification remarks 15 Accessories 16 Specifications ...

Page 3: ...supervision of well trained medical specialists in the field of use The ultimate judgment whether a test subject should undertake a stress test with the device must be made by the responsible medical specialist based on the limitations of each individual the medical history and all other applicable circumstances Neither the manufacturer nor its distributors assume any responsibility for the final ...

Page 4: ...g while using the device Service of this device is restricted to factory trained personnel only Do not jump onto the rotating belt Do not jump off the rotating belt not even to the front Do not stop moving on the running belt Do not turn around on the running belt Do not move sideways or backwards Do not make movement that could get you out of balance The test subject shall wear clothes when using...

Page 5: ... with protective earth to avoid the risk of electric shock The functional earth connection is for potential equalisation only The operator must keep away from the footrail while lowering The test subject should wear proper sports or running shoes no spikes If this equipment is modified appropriate inspection and testing must be conducted to ensure continued safe use of the equipment Cleaning and u...

Page 6: ... No modification of this equipment is allowed Not suitable for use in the presence of flammable anaesthetics The appliance is not to be used by persons including children with reduced physical sensory or mental capabilities or lack of experience and knowledge The test subject should wear proper sports clothing The operator should check possible entrapment of clothing fingers or feet before the sta...

Page 7: ... and or SpO2 blood saturation measurement These measurements may be inaccurate depending on use circumstances These measurements are indicative and cannot be used for diagnostic purposes 3 Contra Indication The device is to be operated by classified personnel only As stated in the intended use the device is intended to be used in a medical environment During the intended use the test subject will ...

Page 8: ...6 Install your treadmill 1 While unpacking the Lode treadmill ergometer leave the transport straps on both ends in place 2 Place the treadmill in the location where it will be used 3 Remove Both transport straps and remaining packaging materials 4 The power supply cord should be installed by authorised service personnel and instructions for correct connection and anchoring must be ensured securing...

Page 9: ...aintenance Maintenance and all repairs should only be carried out by an authorized agency The manufacturer will not be held responsible for the results of maintenance or repairs by unauthorized persons The check up and or technical maintenance must be carried out conform the procedure described in the service manual of the unit Opening of the equipment by unauthorized agencies is not allowed and w...

Page 10: ...ey on the left side Turning the key clockwise will make the belt move to the right side of the treadmill Turning the key counterclockwise will make the belt go left The red marked area shows where you can check the belt alignment The goal is to align the belt edge exactly in the center of the hole 10 21 M938900 938905_MANUAL VALIANT 2 CPET XL VERSION V1 1 2016 08 25 ...

Page 11: ... you can timely intervene is the belt goes to one side quickly To finalize the alignment procedure set the treadmill to maximum speed for 2 minutes and check the alignment of the belt Note NEVER use any lubricants on the walking belt The belt is self greasing using a wax layer that will dispose on the deck on high friction points and on higher speeds The belt does not need to be re waxed and runs ...

Page 12: ...e moist and not dripping wet If the cleaning agents and disinfectants used are caustic or contain alcohol they may damage and or discolor the upholstery If you are not certain about the discolouring effect of a cleaning agent you may try a little on a part of the product that is not visible during normal use Lode cannot advise a specific cleaning agent since local recipes may differ 10 Control wit...

Page 13: ...cable to the external device You can control the device with e g an exercise ECG device or Metabolic Card 2 Download the interfacing instructions from the internet at www lode nl after you have registered your device 3 If your device has a control unit select Terminal mode to allow external commands When the device does not have a control unit the terminal mode is automatically selected See for de...

Page 14: ...EC of 27 January 2003 of the European Parliament and of the Council on waste electrical an electronic equipment are necessary to reduce the waste management problems linked to the heavy metals concerned and the flame retardants concerned Trapping zone Warning do not jump onto the rotating belt Do not jump off the rotating belt not even to the front Do not stop moving on the running belt Do not tur...

Page 15: ...13 Electromagnetic Compatibility EMC 15 21 M938900 938905_MANUAL VALIANT 2 CPET XL VERSION V1 1 2016 08 25 ...

Page 16: ...16 21 M938900 938905_MANUAL VALIANT 2 CPET XL VERSION V1 1 2016 08 25 ...

Page 17: ...17 21 M938900 938905_MANUAL VALIANT 2 CPET XL VERSION V1 1 2016 08 25 ...

Page 18: ...18 21 M938900 938905_MANUAL VALIANT 2 CPET XL VERSION V1 1 2016 08 25 ...

Page 19: ...our treadmill may differ from the speed mentioned at the specifications 15 Accessories Colour display 3 5 single display Partnumber 945810 Clear feedback Colour display 3 5 single display Control Unit with 7 touch screen for treadmill Partnumber 945814 Multifunctionality Control Unit with Touchscreen Programmable Control Unit with 7 touch screen for Partnumber 945815 Programmable Programmable Cont...

Page 20: ...nt 2 Handrails Side Adjustable Pediatric for Valiant 2 Partnumber 938812 Making your treadmill suitable for children Adjustable pediatric side handrails Safety Belt and Fall stop for Valiant 2 Partnumber 938803 Extra safety Safety Belt and Fall stop Ambient sensor pack Partnumber 945826 Check environmental conditions during test Ambient sensor pack for treadmills Universal treadmill Arm Support Pa...

Page 21: ... Power cord NEMA Maximum rated power input 2500 VA 115 V AC 50 60 Hz 2 phases 230 V AC 50 60 Hz Maximum motor power 1 8 kW Standards Safety IEC 60601 1 2005 ISO 13485 2016 compliant ISO 9001 2015 compliant Standard emergency lanyard Certification CE class Im according to MDD93 42 EEC CE class of product with optional SpO2 IIa CE class of product with optional BPM IIa CB according to IECEE CB Envir...

Reviews: