background image

LN0369E2108E0   

         

Mass Flow Controller 

Instruction Manual 

   

MC-3000L-TC Series

 

  

Safety Precautions 

 

 
Incorrect handling may cause death or injury

 

 

(1)

 

Before connecting the fittings, check that no damage or defects are found on 
the  fittings.  Make  connections  properly  and  make  sure  that  a  leak  test  is 
conducted  before  actual  operation  to  prevent  fluid  from  leaking  into  the 

atmosphere (Hereafter, the measured fluid is called “gas” or “fluid”).

 

(2)

 

DO  NOT

  apply  any  fluids  corrosive  to  materials  exposed  to  gas.  Corrosion 

may cause fluid to leak into the atmosphere. Check the gas type to be used in 
advance.

 

(3)

 

This device is not designed as an explosion proof structure. 

DO NOT

 use this 

device in a place where explosion-proof structures are required. Doing so may 
cause fire or explosion.

 

 

Incorrect handling may lead to medium or slight injury or may 
cause damage to, or loss of, facilities or equipment

 

 

(1)

 

Observe the precautions listed in the WARNING (above).

 

(2)

 

Strictly  observe  the  electrical  specifications.  Not  doing  so  may  cause  fire, 
damage to sensors or malfunction.

 

(3)

 

This  device  is  not  designed  to  be  waterproof. 

DO  NOT

  locate  this  device 

outdoors or in a place where it may be splashed with water. Doing so may cause 
fire, trouble, or malfunction of the device.

 

(4)

 

DO NOT

 modify this device. It may cause fire or other problems. 

(5)

 

While a power supply is applied to the device, +15VDC and -15VDC must be 

applied  simultaneously.  If  only  +15VDC  or  -15VDC  is  applied,  electronic 

circuits will become unstable and it may cause a malfunction.

 

(6)

 

This  device  is  not  designed  to  handle  hot  swap.  Please  avoid  attaching  and 
removing the power supply connector and interface connector with the power 
switched  on.  Attachment  and/or  removal  with  the  power  on  may  result  in 
failure of the device.

 

(7)

 

This device is a precious device, please handle it carefully. Dropping down or 
handling  it  carelessly  will  cause  damage.  Please  use  assist  instrument  while 
moving or setting the device.

 

(8)

 

Regular maintenance is recommended for steady use of this device 
(Recommended proofreading frequency is once a year). 

 

 

1. Introduction

 

 

 

This manual explains basic operation of the MC-3000L-TC series (Hereafter, it is 
called "MFC"). Please read through this manual and other separate volumes (Digital 
Interface  Manual,  Special  Function  Manual,  Command  Chart)  carefully  to 
familiarize yourself with the features of this device. 

 

2. Summary 

   

The MFC is high performance mass flow controller for gas using thermal flow 
sensor and high response solenoid actuator valve. 

 

3. Features 

The MFC is the following features. 

(1)

 

LINTEC’s proprietary ambient temperature compensation type flow sensor is carried.

 

 

  ·

 

Low sensor temperature, long-time steady monitor, and few problems such 

       

           

as degradation of gas to be monitored. 

·

 

The effect of ambient temperature is less because the sensor temperature is 

 

   

kept at the ambient temperature.   

     

         

·

 

Steady temperature distribution of sensor and high-speed response. 

(2)

 

Digital computing system with microprocessor and high-resolution A/D, D/A 
converter is carried. 

 

· High functionality. 
·

   

By setting of device number (address), multiple devices can be controlled   

 

   

through a single interface. 

· Many additional are provided as standard, e.g., totalizer function, zero adjust, 
 

 

ramping function. 

(3)  Universal digital Interface (RS-485). 
(4)

 

   

Small dead space structure using a diaphragm valve. 

(5)

 

Superior  corrosion  resistant  sealing materials  make maintenance easy.  Metal 
seals (Au). 

(6)

 

Particle-free structure.

   

(7)

 

By using a metal case and various types of filters, steady operation can achieve 
even in an environment of high-frequency noise and stationary magnetic field.

 

(8)

 

Based on RoHS. 

 

 

4. Specification / Dimensions 

                   

(1) Specification 

Name 

Mass Flow Controller 

Model 

MC-3102L-TC 

MC-3202L-TC 

Standard flow rate in nitrogen 

10 SCCM to 5 SLM 

  ~25 SLM 

Operating temperature 
(High temperature part) 

-L 

80 to 100 ºC·0 to 80%RH (Condensation should be avoided) 

-H 

100 to 120 ºC·0 to 80%RH (Condensation should be avoided) 

Operating temperature 
(Control part) 

5 to 50 ºC·0 to 80%RH   

(Condensation should be avoided) 

Baking temperature 

Below 150 ºC   

Valve operation mode 

Normally closed valve 

Internal surface treatment 

-TC: No treatment      -SC: Precision polishing 

Minimum controllable flow rate 

2%F.S. 

Analog flow rate output signal 

Proportional to flow rate 0 to 5VDC 

Analog flow rate setting signal 

Accuracy 

±1.0%F.S. 

Repeatability 

±0.2%F.S. 

Response time 

1 second (Typical) 

3 seconds (Typical) 

Operating differential pressure 

50 to 300kPa 

150 to 300kPa 

Maximum operating pressure 

300kPa(G) 

Withstanding pressure 

1MPa(G) 

Storage temperature & Humidity 

5 to 50 ºC·0 to 80%RH (Condensation should be 

avoided) 

Leak integrity 

Less than 1×10

-11

Pa · m

3

/sec (He) 

Mounting direction 

Free 

Wetted materials 

Stainless steel 316L, PTFE, Au 

Seal Materials 

Au 

Actuator 

Solenoid actuator 

Fittings 

6.35VCR (124)      6.35SWL (127) 

Power supply 

+15VDC±3%: 100mA      -15VD±3%C: 200mA 

Digital interface 

RS-485 (2-wire) 

Option 

B:

 

Applied to liquefied gas 

Weight 

Approx. 1.2kg 

(Controlling part 0.2kg, High-temperature part 1kg) 

                 

·Connect the MFC to the frame ground.   

 

(2) Dimensions 

 

5. Ordering information 

 

M C - 3 1 0 2 L   –   T C   -   4 J R 2   B L 0 A 0   –   d d   –   e e   -   f f  

     

   

 

  [1]

 

            [2]

 

    [3]

 

   

   

[4]         

   

[5]   

     

[6]   

 

[7] 

 

[1] Series model MC: Mass Flow Controller Series 

 

  MC-3102L: Small flow rate

 

MC-3202L: Large flow rate   

[2] Inner surface treatment 

     

-TC: No treatment

   

-SC: Precision polishing

   

 

[3] Fitting 
      4JR1: 6.35UJR106, 4JR2:6.35UJR124 
      4VR1:6.35VCR106, 4VR2:6.35VCR124, 4SWL :6.35SWL 
[4] Optional 

     

   

B: Applied to liquefied gas,

 

J:

     

Valve is closed at below 2% setting   

     

   

L: Temperature range 80 to 100℃

 

H: Humidity range 100 to 120℃ 

     

    

※Default setting is labeled “AA0A0”. Please consult for more information.   

[5] Gas type 
[6] Full scale flow rate 
[7] Flow rate unit 

     

 

SCCM (0 ºC standard), SLM (0 ºC standard) 

 

6. Connection 

  (1) Analog interface connector

 

                                           

Mounted connector    : D-Sub 9 pin(male) 

                                           

Pair connector            : D-Sub 9 pin(female)    (Fixed size of screw : M3) 

Pin No. 

Single name 

Function 

Valve on-off input (Note1) 

+15VDC: OPEN,

 

  -15VDC: CLOSE 

Flow rate output signal 0 to 5VDC 

Flow rate output voltage plus side 0 to 5VDC 

Power supply input +15VDC±3% 

Power supply (plus) 100mA 

Power supply COMMON (Note2) 

COMMON ±15VDC 

Power supply input -15VDC±3% 

Power supply (minus) 200mA 

Flow rate setting signal 0 to 5VDC 
(Note1) 

Flow rate setting voltage plus side 0 to 5VDC 

Flow rate output COMMON (Note2) 

Flow rate output voltage COMMON 

Flow rate setting COMMON (Note2) 

Flow rate setting voltage COMMON 

N.C. 

N.C. 

 

                     

Note1) Input impedance of flow rate setting signal is 100kΩ. Valve override impedance is 100kΩ. 

 

     

           

   

Note2) Pin No.4,7,8 are connected internally. 

 
     

   

Wirings should be done as shown below in order to remove the effect of potential difference 

     

   

among the COMMON.   

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

   

Do not connect.

 

Flow rate setting signal COMMON[8],Flow rate output signal COMMON[7], 

   

and Power supply COMMON[4] in 

 

the power supply unit. 