background image

Operating instruction for pressure transmitter PASCAL CV4

4...20 mA, Type series CV3... with switching module

LABOM Mess- und Regeltechnik GmbH  Im Gewerbepark 13  27798 Hude  Germany

Hotline: +45 4408 804-444  Fax: +49 4408 804-100  e-mail: [email protected]  www.labom.com

Operating instruction BA_029_2019-08_10.00

Page 1/6

Page 1

Further information required? H49 (0) 4408 804-444 

BTA-No. 029 Rev. 1

F6

Operating Instructions 

for pressure transmitter

PASCAL CV, 4...20 mA, Type Series CV 3

 . 

. . with switching module

Basic modules

PROFIBUS PA

Display module

HART

®

 module

switching module

4...20 mA

Function modules

Various modules can easily be added to PASCAL CV  (see table page 5).

www.pascal-cv.com

LABOM Mess- und Regeltechnik GmbH  · P.O. Box 12 62 · 27795 Hude · Germany · Im Gewerbepark 13 · 27798 Hude ·  Germany  

Tel. +49 4408 804-0 · Fax +49 4408 804-100 · e-mail: [email protected] · www.labom.com

 

Features

  Modular pressure transmitter 

  Output signal: 

  4..20 mA, optional with HART

®

 protocol

  Function modules 

 

 

 

Multifunctional display with 5-segment digital display  

  and bar graph

 

  

Switching module with 2 floating channels, maximum  

  0.5 A switching current, electrically isolated at all   

    sides, without additional auxiliary power

  Function module replacement on site without recali-

  bration "plug and measure"

  Watchdog for electronics modules and measuring cell

  Classfication per SIL 2

  Accuracy: 

 0,15 %

  Turndown 5:1

  Degree of protection IP 66

  Piezoresistive measuring cell directly aerated, fully 

  welded, without inside gasket

Reviews: