background image

- 1 -

- 2 -

- 3 -

- 4 -

AUTOMOTIVE

DIGITAL MULTIMETER

KT-2021

To prevent an electrical shock hazard to the operator and/or damage to the instruments, 
read this instruction manual carefully before using the instrument. 

WARNINGS

 with the 

symbol       on the instrument and this instruction manual are highly important.

Important Symbols

The symbol in this manual advises the user of an electrical shock 
hazard that could cause injury or material damages.

  

1. DISPLAY

LCD

    

a. Numerical Display

 

4 digit 6000 count, with Bar Graph

    

b. Units and Symbols

 

 2. OPERATING PRINCIPLE

 Sigma-delta Conversion

  3. MEASURING PRINCIPLE

 

Average Rectification

 

 4. SAMPLING RATE

 

5 times per second (Bar Graph

40 times per second max.)

 

 5. RANGE SELECTION

Auto / Manual (Frequency, Injection Pulse width and

Capacitance measurement are Auto only)

 

 6. BATTERY WARNING

:   

indication at approx. 2.3V or less

  

7. DISPLAY HOLD

Hold indicating values by pressing HOLD Key

  8. AUTO POWER OFF

Power turns off automatically after approx. 34 minutes  

  9. OVERLOAD PROTECTION

10. OPERATING TEMPERATURE & HUMIDITY

0 to 30

less than

 

80

RH 

31 to 40

less than

 

50

RH in non-condensing.

11. STORAGE TEMPERATURE & HUMIDITY

       -20 to 60

less than

 

80%RH in non-condensing.

12. TEMPERATURE COEFFICIENT

Accuracy at 23

±

5

×

0.15/

13. SAFETY LEVEL

CE marking approved.

14. POWER SUPPLY

1.5V R03(AAA) Battery 

×

2

15. FUSE

Fast-acting 0.63A/500V

 (

φ

6.0

×

32mm

) ×

1,

Fast-acting 6.3A/500V

 (

φ

6.0

×

32mm

) ×

1

16. POWER CONSUMPTION

approx.

 

4.3mA

(in Auto Power Off 

approx. 10

μ

A, in backlighting

approx. 27mA)

17. CONTINUOUS OPERATING TIME

approx. 180 hours (manganese cell),

approx. 360 hours (alkaline cell)

18. DIMENSIONS & WEIGHT

161(H)

×

80(W)

×

50(D)mm, approx. 340g

19. ACCESSORIES

20. OPTIONAL ACCESSORIES

a) 

μ

A & mA

0.63A/500V fast-acting fuse

b)  A

6.3A/500V fast-acting fuse

c)  V

1050Vrms or 1450Vpeak

d)  mV, 

Ω

 and others

600Vrms

Thank you for purchasing 

"KT-2021 AUTOMOTIVE DIGITAL MULTIMETER"

. To obtain the 

maximum  performance  of  this  instrument,  read  this  Instruction  Manual  carefully,  and  take 
safe measurement.

Confirm if the following items are contained in the package in good condition.
If there are any damage or missing items, ask your local dealer for replacement.

1. Digital Multimeter (with holster)
2. Test Lead (100-66)
3. RPM Sensor (653 / for Direct Ignition Engine)
4. RPM Sensor (654 / for High Tension Code)
5. Temperature Probe (818-02)
6. Alligator Clip (943)
7. Carrying Case (1024)
8. Spare Fuse (F20, 0.63A/250V and F32, 6.3A/500V)
9. Instruction Manual

2. SPECIFICATIONS

Do  not  measure  High  Power  Line  (High  Energy  Circuits)  that  might  exceed  6kVA.  High 
Power  Line  is  very  dangerous  and  sometimes  includes  High  Surge  Voltage  that  could 
cause explosive short in the instrument and could result in serious injury to the operator. For 
dangerous voltage measurement, always keep the instrument away from your body without 
holding it in your hands.  Do not touch the instrument, test leads, and any part of the circuit.

WARNING

FOR SAFETY MEASUREMENTS!!

INSTRUCTION MANUAL

The  symbol  listed  in  IEC  61010-1  and  ISO  3864  means  "Caution  (refer  to 
instruction manual)".

WARNING

CAUTION

The symbol in this manual advises the user of an electrical shock 
hazard that could result in serious injury or even death.

INTRODUCTION

1. UNPACKING AND INSPECTIONS

1 pce.
1 set
1 pce.
1 pce.
1 pce.
1 pce.
1 pce.
1 pce. each
1 pce.

AUTO,      ,     ,     ,     , RPM     , RPM     ,     ,     , 

Ω

, k

Ω

, M

Ω

,

Hz, kHz, ms, V, 

μ

F, 

μ

A, mA, A, 

, Trig and decimal point

2-1. GENERAL SPECIFICATIONS

2-2. MEASUREMENT SPECIFICATIONS

 

(23

±

5

, <80

RH in non-condensing)

1. DC Voltage  (     mV / V )

6.000 V
60.00 V
600.0 V

1000 V

±

0.4

rdg

±

3dgt 

±

0.7

rdg

±

3dgt 

Range

60.00 mV
600.0 mV

10M

Ω

,

50pF

600Vrms

Accuracy

±

0.4

rdg

±

3dgt 

±

0.3

rdg

±

3dgt 

Resolution

0.01 mV

0.1 mV

Input

Impedance

Overload

Protection

10M

Ω

,

50pF

1050Vrms

1450Vpeak

1 mV

10 mV

100 mV

1 V

2. AC Voltage  ( 

mV / V )    50Hz to 500Hz

Average Rectification

3. DC Current  (    

μ

A / mA / A )

6A Continuous measurement is possible.
10A is measurable within 30 seconds but 5 minutes cooling down interval is necessary.

4. AC Current  ( 

μ

A / mA / A )   50Hz to 500Hz

6A Continuous measurement is possible.
10A is measurable within 30 seconds but 5 minutes cooling down interval is necessary.

Average Rectification

5. Resistance  ( 

Ω

 )

Range

600.0 

Ω

6.000 k

Ω

60.00 k

Ω

600.0 k

Ω

6.000 M

Ω

60.00 M

Ω

0.45V DC

Accuracy

±

0.5

rdg

±

6dgt 

±

0.5

rdg

±

3dgt 

±

0.8

rdg

±

4dgt 

±

1.0

rdg

±

5dgt 

±

1.5

rdg

±

5dgt 

Resolution

0.1 

Ω

Ω

10 

Ω

100 

Ω

1 k

Ω

10 k

Ω

Open Circuit Voltage

6. Frequency  ( Hz )

Function

6 V

60 V

600 V

1000 V

±

0.1

rdg

±

3dgt

Range (Auto)

10 Hz to 10 kHz

10 Hz to 50 kHz

45 Hz to 1 kHz

Sensitivity

0.5 V rms

5 V rms

50 V rms

500 V rms

Accuracy

NOTE :

 Measurement cannot be made depending on the vehicles.

RPM4
RPM2

RPM2-M

Range

240 to 20000RPM
120 to 10000RPM

60 to 5000RPM

7. IP-RPM ( Inductive Pickup type )

±

0.2

rdg

±

20dgt

Accuracy

LVD

V/

Ω

 Terminal

CAT

1000V AC/DC, 

μ

A/

mA/A Terminal

CAT

450V AC/DC

EMC

EN61326-1 (2006)

NOTE :

 Measurement cannot be made depending on the vehicles.

RPM4
RPM2

RPM2-M

Range

60 to 20000RPM
30 to 10000RPM

15 to 5000RPM

8. IG-RPM ( Contact Signal type )

±

0.2

rdg

±

20dgt

Accuracy

Selectable Number of Cylinders

1,2,3,4,5,6,8,10,12

(Select by 

CYLINDER Key

)

      WARNING 1. Checks of Body and Test Lead

      WARNING 3. Warning for High Voltage Measurements

      WARNING 4. Dangerous Voltage Measurement Procedure

      WARNING 2. High Power Line Measurements is Prohibited  

Before  measurement,  confirm  the  body  of  this  instrument  and  handle  insulators  of  the  Test 
Lead have no cracks or any other damages. Dust, grease and moisture must be removed.

Do not hold multimeter in your hands. Keep safety distance from power source or circuit 
to be measured not to touch the dangerous voltage.
Attach black and red alligator clips to test lead pins.
Turn off the power of the circuit to be measured when connecting test leads.
After measurement, before detaching alligator clips (test leads), turn the circuit power off 
and discharge the all capacitors.



For dangerous voltage measurement, strictly observe the warnings below.

3-1. WARNINGS

Correct  knowledge  of  electric  measurements  is  essential  to  avoid  unexpected  danger  such 
as  operator's  injury  or  damage  to  the  instrument.  Read  carefully  and  observe  the  following 
precautions for safety measurements.

3. SAFETY PRECAUTIONS

Even  for  Low  Energy  Circuits  of  electric/electronic  appliances,  such  as  heating  elements, 
small motors, line cords and plugs, High Voltage Measurements are very dangerous. Do not 
touch any part of the circuit. 

Do not measure High Power Line (High Energy Circuits) such as Distribution Transformers, 
Bus  Bars  and  Large  Motors.  High  Power  Line  sometimes  includes  High  Surge  Voltage  that 
could  cause  explosive  short  in  the  instrument  and  could  result  in  shock  hazard.  Generally, 
shock hazard could occur when the current between the circuit, that involves more than 33V 
rms or 46.7V DC or peak, and ground goes up to 0.5mA or more.

      WARNING 1. Correct Selection of Function Switch

      WARNING 2. Maximum Input Observance

      WARNING 3. Test Lead Detachment

Do not measure anything that might exceed the specified maximum input values.

Always confirm that 

FUNCTION

 Switch is set to the correct position. Do not measure voltage 

except at Voltage measurement function.

Detach  test  leads  from  measuring  circuit  before  changing  measurement  functions  or 
removing rear cover for battery or fuse replacement.

In case of live-line measurement, strictly observe the warnings below 

Do not hold multimeter in your hands. Keep safety distance from power source or circuit 
to be measured not to touch the dangerous voltage.
Black test lead : Attach black alligator clip and connect to 

 (earth) side of the circuit.

Red test lead : Connect to 

 (positive) side of the circuit.


3-3. GENERAL WARNINGS AND CAUTIONS

Children  and  the  persons  who  do  not  have  enough  knowledge  about 
electric measurements must not use this instrument.
Do  not  measure  the  electricity  naked  or  barefooted  to  protect  yourself 
from electrical shock hazard.
Be careful not to get hurt with the sharp test lead pins.
Do  not  polish  the  case  or  attempt  to  clean  it  with  any  cleaning  fluid  like 
gasoline or benzine. If necessary, use silicon oil or antistatic fluid.
Away the instrument from hot and humid conditions like in the car. Do not 
apply hard mechanical shock or vibration.
Remove the battery when the instrument is out of use for a long time. The 
exhausted battery might leak electrolyte and corrode the inside.

WARNING 1.

 

WARNING 2.

 

WARNING 3. 
CAUTION  1.

CAUTION  2. 

CAUTION  3.
CAUTION  4.              

 

Power turns off automatically after approx. 34 minutes of last operation.

To cancel it : 

Turn the power on holding down 

Select 

Key. Auto power off is disabled.

NOTE :

 Small current consumption remains even in the auto power off condition.

Be sure to set 

FUNCTION

 Switch to 

"OFF" 

after finishing the measurement.

Caution (refer to instruction manual)

Alternating Current

Direct Current

Earth (Ground)

Fuse

Double Insulation

DC / AC

Do not apply around, or remove from HAZARDOUS LIVE conductors.

LCD  displays  "

OL

"  when  measurement  value  exceeds  the  maximum  value  of  each 

measurement range.

The  following  symbol  marks  shown  on  the  instrument  and  instruction  manual  are  listed  in 
IEC 61010-1 and ISO 3864.

0.63A/500V and 6.3A/500V fuses are installed to protect current measurement functions. If 
the fuse is blown out, replace it referring to "

7-1. BATTERY AND FUSE REPLACEMENT

".

4. NAME ILLUSTRATION

fig-1

LCD

Range Key
Level Key

FUNCTION Switch

KT-2021

Automotive DMM

Hold Key

±

TRIGGER Key

Key

CYLINDER Key

Select Key

     Key

Input
Terminals

(


(

(

fig-2

4-1. LCD

: 

Auto-ranging

: 

Low battery warning

: 

Dwell angle measurement

: 

Continuity test

: 

Lights up

      in display hold function

: 

RPM measurement

: 

Direct current (DC)

RPM

: 

Alternating current (AC)

: 

Resistance measurement

: 

Frequency measurement

: 

Injection pulse width measurement

: 

Voltage measurement

: 

Capacitance measurement

: 

Current measurement

: 

Duty cycle

Ω

, k

Ω

, M

Ω

Hz, kHz

ms

mV, V

μ

F

μ

A, mA, A

4-11. Input Terminals

4-10. Level Key

Use  this  key  to  select  the  sensitivity  level  in  IP-RPM,  IG-RPM,  Dwell  Angle  and  Injection 
Pulse Width measurements.

Connect black test lead to COM terminal and red test lead to RPM/Temp+/

Ω

/V/     Terminal 

for all measurements except 

μ

A, mA, A measurements.

NOTE

Connect red test lead to mA/

μ

A terminal for mA/

μ

A measurement and connect it to

A terminal for 6A (10A for 30sec) measurement.

5. MEASUREMENT PROCEDURES

5-1. PREPARATION FOR USE

Read  INSTRUCTION  MANUAL  carefully  to  understand  the  specification  and  functions 
correctly. "

3. SAFETY PRECAUTIONS

" is very important for safety measurement.

1. INSTRUCTION MANUAL

2. BATTERY

Two 1.5V R03 batteries are installed in this instrument. When "      " lights up on LCD, replace 
them into the new ones referring to "

7-1. BATTERY AND FUSE REPLACEMENT

".

3. FUSE

4. OVERLOAD INDICATION

5. AUTO POWER OFF

6. SYMBOL MARK

Specified ranges depend on engine rpm and number of cylinders (cyl).

Range

0.0

°

 to 360.0

°

0.0

 to 100.0

Accuracy

±

1.2

°

/krpm

±

1dgt 

±

0.04

/krpm/cyl

±

2dgt 

9. Dwell Angle (      ), Duty Cycle ( 

 )

Selectable Number of Cylinders

1,2,3,4,5,6,8,10,12

(Select by 

CYLINDER Key

)

Specified ranges depend on trigger slopes, engine rpm and number of cylinders (cyl).

Accuracy

±

0.05ms

±

1dgt

±

0.04%/krpm

±

2dgt

±

0.05ms

±

1dgt

±

0.04%/krpm/cyl

±

2dgt

Range

0.05ms to 250.0ms

0.0% to 100.0%

0.05ms to 250.0ms

0.0% to 100.0%

10. Injection Pulse Width ( ms ),  Duty Cycle ( 

 

)

Multi Point Injection

Single Point Injection

WARNING

Do not measure any voltage that might exceed maximum input value.
(V

1000V AC/DC, mV

600mV AC/DC). 

Read  "

3.  SAFETY  PRECAUTIONS

"  carefully  to  avoid  electric  shock  hazard  and 

serious damage to the instrument.
Do not touch any part of power line or the circuit to be measured.

1

2

3

4

5

6

Connect  the  instrument 

IN 

PARALLEL

 to the circuit.

Use  alligator  clips  for  dangerous 
voltage measurement.

NOTE

Display  shows  "      V

→〜

V

Hz  " 

each time the 

Select

 Key is pressed.

Connect  black  test  lead  to 

  (earth)  side 

of  the  circuit  being  measured  and  connect 
red test lead to 

 (positive) side.

NOTE :

 

NOTE :

Read the measurement value on LCD.
After  finishing  the  measurement,  set 

FUNCTION

 Switch to 

"OFF"

.

5-2. VOLTAGE AND FREQUENCY MEASUREMENT (    V / 

V / Hz ) (    mV / 

mV)

KT-2021

Automotive DMM

fig-3

Battery voltage
measurement

Black(

)

Red(

)

Battery

(    V / 

V / Hz )

Insert black test lead to 

COM

 terminal, and insert red test lead to 

RPM/Temp

/

Ω

/V/

terminal.
Set 

FUNCTION

 Switch to "       ".

Press 

Select

 Key to select DC(    ) or AC(    ).

V

Hz

100-66  Test  Lead,  653  RPM  Sensor  (for  Direct  Ignition),  654  RPM  Sensor  (for  High 
Tension  Code),  818-02  Temperature  Probe,  943  Alligator  Clip,  1024  Carrying  Case, 
Holster,  1.5V  R03(AAA)  Battery  (Installed),  F20  Spare  Fuse  (0.63A/500V),  F32  Spare 
Fuse (6.3A/500V), Instruction Manual

11. Diode Test  (        )

Range

1.000 V

Test Current

0.50mA typ.

Open Circuit Voltage

1.6V DC typ.

Accuracy

±

1.0

rdg

±

3dgt 

K-type thermocouple / sensor accuracy is not included.

Range

50

°

C to 1000

°

C

58

°

F to 1832

°

F

Accuracy

±

0.5

rdg

±

3dgt 

±

0.5

rdg

±

6dgt 

12. Temperature  ( 

°

C

 

 

°

F )

Range

6.000

μ

F

60.00

μ

F

600.0

μ

F

2000

μ

F

Accuracy

±

2.0

rdg

±

5dgt 

±

3.5

rdg

±

5dgt 

±

4.0

rdg

±

5dgt 

14. Capacitance  (        )

660 AC/DC Clamp Adapter, 100-41 Test Lead Kit, 100-62 Test Lead Set, 944 Test Pin, 
946 Battery Clip, 793 Coil Contact Pin

When using Clamp Adapter, read its instruction manual carefully.

Threshold Level

between 10

Ω

 and 200

Ω

Response Time

32ms

13. Continuity  (       )

Turns the power on and selects measurement functions. After measurement, turn it to 

"OFF"

.

4-3. Select Key

Use this Key to select sub-measurement functions.

4-2. FUNCTION Switch

Press this key for 1 second or more to turn on the LCD backlight. Press for 1 second or more 
again, backlight is turned off.

4-4. 

  

Key

4-6. CYLINDER Key

4-8. 

±

TRIGGER Key

Change 

/

 Trigger slopes in Duty Cycle and Injection Pulse Width measurements. (Refer to 6-2).

4-5. 

   

Key

4-7. Hold Key

Display Hold

Use  this  key  to  select  the  engine  strokes  in  IP-RPM  and  IG-RPM  measurements:  Four-
stroke / Two-stroke / Two-stroke, 2-cylinder simultaneous ignition.

Use this Key to set number of cylinders in Dwell Angle and IG-RPM measurements.

Holds displayed value on LCD by pressing this Key. ("

  

" lights up).

To release it

Press 

Hold 

Key again.

FREQUENCY MEASUREMENT  ( Hz )

(1)

(2)
(3)

Press 

Select

 Key during voltage measurement. "Hz" lights up on LCD and Frequency 

of the measuring voltage is displayed.
Press 

Select

 Key again to return to voltage measurement.

In Frequency measurement, input sensitivity can be changed in the following procedures.

a.

b.
c.

Press 

Level

 Key during frequency measurement. Input sensitivity level is changed 

as Level 1 (6V)

2 (60V)

3 (600V)

4 (1000V). (default setting is Level 1 (6V)).

Sensitivity level is shown with the number of "    " under the bar graph.
Change of input sensitivity is effective in the following cases.
When measured value is unstable, reduce the current input sensitivity level.
When measured value is "0", increase the current input sensitivity level.

(    mV / 

mV )

1

2

3

4

5

6

Insert black test lead to 

COM

 terminal, and insert red test lead to 

RPM/Temp

/

Ω

/V/

terminal.
Set 

FUNCTION

 Switch to "           ".

Press 

Select

 Key to select DC (     ) or AC (     )

NOTE

Display shows "     mV

→〜

mV

→°

C

→°

F " each time the 

Select

 Key is pressed.

Connect black test lead to 

 (earth) side of the circuit being measured and connect red 

test lead to 

 (positive) side.

NOTE : 

Connect the instrument 

IN PARALLEL

 to the circuit.

NOTE : 

Use alligator clips for dangerous voltage measurement.

Read the measurement value on LCD.
After finishing the measurement, set 

FUNCTION

 Switch to 

"OFF"

.

Do not measure voltage in "                           " position. This will cause electrical shock 
hazard to the operator and/or damage to the instrument.
Be careful not to drop off the instrument and test leads into the engine room.
Ignition  system  generates  high  voltage  that  could  cause  an  electrical  shock  hazard. 
Stop the engine when connecting or removing the test leads.
Read "

3. SAFETY PRECAUTIONS

" carefully before measurement.


WARNING

5-3. RPM MEASUREMENT  (IG-RPM)  

measuring RPM with test lead

-ms

IG-RPM

Stop the engine.
Insert black test lead to 

COM

 terminal, and insert red test lead to 

RPM/Temp

/

Ω

/V/

terminal.
Set 

FUNCTION

 Switch to "                       ", and press 

Select

 Key until "RPM" appears 

on LCD.

-ms

IG-RPM

1

2

3

4-9. Range Key

Use this Key to change Auto or Manual-ranging (Refer to 6-1).

3-2. PREVENTION OF FAILURE

Maximum Input

600mV AC/DC

1000V AC/DC

6.000 V
60.00 V
600.0 V

1000 V

±

2.0

rdg

±

5dgt

±

2.2

rdg

±

5dgt

Range

60.00 mV
600.0 mV

10M

Ω

,

50pF

600Vrms

Resolution

0.01 mV

0.1 mV

Input

Impedance

Overload

Protection

10M

Ω

,

50pF

1050Vrms

1450Vpeak

Maximum Input

600mV AC/DC

1000V AC/DC

Accuracy

±

2.0

rdg

±

5dgt 

1 mV

10 mV

100 mV

1 V

60.00 mA
600.0 mA
6.000 mA
10.00 mA

Range

600.0 

μ

A

6000 

μ

A

0.25mV

μ

A

0.63A

500V

Fuse

Resolution

0.1 

μ

A

μ

A

Burden Voltage

Overload

Protection

Maximum Input

6000

μ

A AC/DC

0.01 mA

0.1 mA

1 mA

10 mA

Accuracy

±

0.7

rdg

±

3dgt 

±

0.5

rdg

±

3dgt 

±

0.7

rdg

±

3dgt 

±

0.5

rdg

±

3dgt 

±

0.7

rdg

±

3dgt 

±

0.5

rdg

±

3dgt 

2.5mV

mA

0.63A

500V

Fuse

600mA AC/DC

0.03V

A

6.3A

500V

Fuse

6A AC/DC

60.00 mA
600.0 mA
6.000 mA
10.00 mA

Range

600.0 

μ

A

6000 

μ

A

0.25mV

μ

A

0.63A

500V

Fuse

Resolution

0.1 

μ

A

μ

A

Burden Voltage

Overload

Protection

Maximum Input

6000

μ

A AC/DC

0.01 mA

0.1 mA

1 mA

10 mA

Accuracy

±

2.2

rdg

±

5dgt 

±

2.0

rdg

±

5dgt 

±

2.2

rdg

±

5dgt 

±

2.0

rdg

±

5dgt 

±

2.2

rdg

±

5dgt 

±

1.2

rdg

±

5dgt 

2.5mV

mA

0.63A

500V

Fuse

600mA AC/DC

0.03V

A

6.3A

500V

Fuse

6A AC/DC

600Vrms

Overload Protection

6553 count max.

Display

600Vrms

Overload

Protection

Resolution

1

°

C

1

°

F

600Vrms

Overload Protection

600Vrms

Overload Protection

Resolution

0.001

μ

F

0.01

μ

F

0.1

μ

F

F

600Vrms

Overload Protection

Reviews: