How to Set your IRwin Methane Leak Detector for 1 Point Calibration 

2.2% (2.5%) - Universal Mode 

(for a factory set unit)

QUICK START GUIDE

1 of 5

1

Make sure your IRwin is set in Universal Mode (see the 

How to set your IRwin for Universal 

Mode

 quick start guide)

Use        to navigate to 

General 

and press

2

3

4

5

Press 

 followed by

 

Use       to navigate to 

AbG Cal Setup

 and press

mimc66en1-02 (1905)

Use        to navigate to 

Custom 

and press

Summary of Contents for IRwin Series

Page 1: ...et unit QUICK START GUIDE 1 of 5 1 Make sure your IRwin is set in Universal Mode see the How to set your IRwin for Universal Mode quick start guide Use to navigate to General and press 2 3 4 5 Press followed by Use to navigate to AbG Cal Setup and press mimc66en1 02 1905 Use to navigate to Custom and press ...

Page 2: ...ollowed by 8 9 2 of 5 mimc66en1 02 1905 Use and to navigate to Calibration and press Use to navigate to Settings and press 10 If prompted enter the min Intermediate password code If not prompted go ahead to next step 6 If prompted enter the Supervisor Advanced password code If not prompted go ahead to next step ...

Page 3: ...nt Calibration 2 2 2 5 3 of 5 mimc66en1 02 1905 Press to get to the drop down menu 12 Press on the first choice of the drop down menu 13 Navigate to far right drop down menu and press 14 Use to select Disabled and press 15 Press again to view drop down menu ...

Page 4: ...Disabled and press 19 4 of 5 mimc66en1 02 1905 20 In Universal Low Range use to select the second choice in the drop down menu press Navigate to far right drop down menu and press Press again to view the drop down menu In Universal Low Range use to select the third choice in the drop down menu press ...

Page 5: ...r respective owners 2019 INFICON mimc66en1 02 1905 How to Set your IRwin Methane Leak Detector for 1 Point Calibration 2 2 2 5 21 22 23 Select Set by using and and press Navigate to far right drop down menu and press Use to navigate to drop down menus Set the actual value of the calibration gas you have nominally 2 2 or 2 5 by entering 2 and press 2 5 and press 24 Press followed by and ...

Reviews: