The Service Information provides the following information  

  

 Precautions and safety information 

•   Specifications     
•   Performance test procedure 
•   Calibration and calibration adjustment procedure 
•   Basic maintenance (replacing the battery and fuses) 

 
 
 

 

 

Form number:  61362 
Rev 1. date:  May 2001

 

 
 

Form number 61362 

rev 1 May 2002 

Summary of Contents for 61-362

Page 1: ...ovides the following information Precautions and safety information Specifications Performance test procedure Calibration and calibration adjustment procedure Basic maintenance replacing the battery and fuses Form number 61362 Rev 1 date May 2001 Form number 61362 rev 1 May 2002 ...

Page 2: ... 1 Specifications 2 General Specification 2 Voltage Specifications 2 Current Specifications 2 Resistance Specifications 2 Diode Continuity Specifications 2 Performance Verification 3 Table 1 Verification 61 362 3 Calibration 4 Replacing the Battery 4 Replacing Fuses 5 Form number 61362 rev 1 May 2002 ...

Page 3: ... manual before starting Not all Caution Warning or Danger precautions are listed in this manual CAUTION These statements identify conditions or practices that could result in damage to the equipment or other property WARNING These statements identify conditions or practices that could result in personal injury or loss of life Specific precautions Use proper Fuse To avoid fire hazard use only the f...

Page 4: ...mensions 5 75 H X 2 75 W X 1 5 D without holster Weight Approximately 18 0 oz including battery Safety UL1244 and Design to comply with IEC 1010 1 Cat III RANGES and ACCURACY SPECIFICATION 61 362 Function Setting Ranges Accuracy 3 200V 32 00V 320 0V 600V AC Voltage 50Hz to 500Hz 2 0 4 digits 320 0mV 32 00V 320 0V 600V 1 2 1digit DC Voltage 3 200V 0 8 1digit 320µA 3200 µA 32 00mA 320mA 2 5 4 digits...

Page 5: ...9 36 30 64 ACV 32 30 00 500Hz 29 36 30 64 ACV 320 300V AC 50Hz 293 6 306 4 ACV 320 300V AC 500Hz 293 6 306 4 ACV 600 500V AC 50Hz 486 514 ACV 600 500V AC 60Hz 486 514 DCV 320mV 300mV DC 296 3 303 7 DCV 3 2V 3 000V DC 2 975 3 025 DCV 32V 30 00V DC 29 63 30 37 DCV 320 300 0V DC 296 3 303 7 DCV 600 500V DC 493 507 A DC 320µA 300µA 293 9 306 1 A DC 3200 µA 3000µA 2939 3061 A DC 32mA 30mA 29 39 30 61 A...

Page 6: ...lowing calibration procedure on an annual basis 61 362 Calibration Procedure Calibration 1 Set the Function Range Switch to the 3 200V DC position 2 Set the output of the DC calibrator for 3 000V and connect it to the V Ohm and COM input terminals 3 Adjust VR1 VR 200 ohm until the display reads 3 000V 1 digit Note This is the only adjustments required for the 61 362 Calibration is complete Battery...

Page 7: ...r in low resistance ohms mode to check the two fuses mA input 0 5A 250V fast acting fuse Amp input 10A 600V fast acting fuse 4 Remove the defective fuse with the recommended fuse mA with 0 5A 250V fast acting fuse Type LA 3895 is recommended Amp with 10A 600V fast acting fuse Type LA 3897 is recommended 5 Install bottom cover and secure with screws Form number 61362 rev 1 May 2002 ...

Reviews: