background image

product manual

06.09

 H-1322B

Marshall and TSR  

Compression Machine

Summary of Contents for Marshall and TSR H-1322B

Page 1: ...product manual 06 09 H 1322B Marshall and TSR Compression Machine ...

Page 2: ......

Page 3: ...on 3 Operation 3 Control Panel 3 Preset Speed of Platen 3 Direction of Travel 4 Maximum Travel Limit Switches 4 Chart Scales 4 Maximum Load 4 Flow Measurement 4 Running a Test 4 Calibration 5 Maintenance 6 General Warnings 7 Saftey Warnings 7 Electrical Warnings 7 Important Notice 7 Updated Products 7 Fitness for Application 7 Parts Drawing 8 Notes 9 1 ...

Page 4: ...nd TSR Compression Machine has a total frame capacity of 10 000 lb 45kN Stress is measured with a supplied S Type load cell The apparatus features an AC single speed motor with an operating rate of 2 in min or 50 8mm sec and a built in yt chart recorder for monitoring load vs rate of compression The 16 bit Analog to Digital converter can be connected directly to a computer or serial printer for li...

Page 5: ...shall Test Machine uses a slo blow fuse that is fitted into the rear of the machine The replacement part a 5 x 20 mm 10A 250V ceramic or glass Initial Installation Remove the packing list and verify that all parts have been received Normally you will receive one crate comprised of compression and recorder frame with calibrated load cell HM 2300 100 10 000lbf capacity 44kN fastened to it 100 charts...

Page 6: ...ights will be illuminated The maximum travel of the platen is 3 50 88mm Chart Scales A sample chart is included with these instructions The vertical axis major divisions each represent depending on the range chosen 500 or 1000 pounds force and the horizontal major divisions each represent 05 inches of flow or platen travel Maximum Load The Marshall TSR Compression Machine is rated a 10 000 lbf 44k...

Page 7: ...ine proceeds until a 5 reduction in load occurs at which time it stops waits 4 seconds then reverses and goes to the lower limit of down travel 6 Load the next specimen and repeat Step 4 7 Upon completion of tests move the recorder lever to the Load position and remove the chart paper Important If a test is incomplete for some reason reset the processor by turning off the machine power with the ro...

Page 8: ...se or decrease the gain as required Calibration of the Flow or X axis is accomplished by measuring the platen travel with a rule while keeping time with a stop watch usually 1 2 in 15 seconds is the measured distance and time check while observing the chart reading If the travel is not within 50 04 on the chart adjust the gain using pot R131 Maintenance The machine and transducers should be kept c...

Page 9: ...it board while the machine is under power will result in permanent damage to the circuit board IMPORTANT NOTICE The information contained herein is supplied without representation or warranty of any kind Humboldt MFG Co therefore assumes no responsibility and shall have no liability consequential or otherwise of any kind arising from the use of the described equipment contained in this manual Upda...

Page 10: ...8 ...

Page 11: ...9 Notes ...

Page 12: ...r parts of such product for the warranty of this product please refer to Humboldt Mfg Co catalog on Terms and Conditions of Sale The purchaser is responsible for the transportation charges Humboldt Mfg Co shall not be responsible under this warranty if the goods have been improperly maintained installed operated or the goods have been altered or modified so as to adversely affect the operation use...