background image

1

Staffan RA

To secure that the walking-aid will function safely for the user, once per 
year review of the walking-aid should be done. Therefore is the following 

interventions recommended. 

 

Wing nut

Control function. 

 

CE-marking

Make sure that the CE-marking 
is still there and readable.

 

Wheels

Make sure the wheels are not 
worn-out and are spinning easily 
without any noice.

 

Front forks

Make sure that the forks swings 
easily.  

 

Lubrication

The front forks shall function 
without noice, lubricate when 
needed.

 

Stability

Control that all wheels are steadily on the ground and that there are no chassis 
distortion. 

 

Test driving

Make a final control by test driving.

 

Cleaning

When cleaning the walker it is recommended machine washing, 60 degrees for
10 minutes. When using a pressure washer, do not wash directly to the bearings.
In addition to machine washing, regular cleaning with a mild cleanser is 
recommended.

 Note! Do NOT use thinner, petrol or other solvents.

 

Post treatment

After washing, all connections, tight spaces and drainage holes should be blown 
dry and sprayed with thin oil. If degreaser is used, should the moving parts be 
lubricated. 

 

The Paint

Control damages to the paint. To prevent corrosion, improve scratches. Touch-up 
paint can be ordered from Human Care HC AB.

 

Pipe frame

Control weldings and screw attaches. If the corrosion is extensive shall the 
rollator be replaced.

 

Handle

Must be fixed. Make sure the handle pipes can be pushed 

down to the chassis. Change the grip if needed.

 

Brakes

Control that both brakes and parkingbrakes are working. 
Wires should be intact and stapped to the chassis. 

RECONDITIONING MANUAL

Reviews: