INTRODUCTION

Thank you for purchasing the Symmetry IQ product.  This product is intended for use by dental  
professionals only.  This device generates ultrasonic vibrations intended for use in dental applications  
such as scaling, periodontal therapy, root canal treatment and cavity preparation. Read this operation  
manual carefully before use and retain for reference.

PRODUCT FEATURES AND BENEFITS

• Compact, lightweight device.

• Three color-coded selectable and adjustable pre-defined power modes for a variety of dental procedures.

• Easy to read digital display for trouble-free operation.

• Convenient memory feature to store frequently used operational settings.

• Fiber-optic handpiece to enhance visibility. 

• Lightweight, large diameter handpiece encourages a light grasp to reduce operator fatigue. 

• Soft, textured silicone finger grip for clinical comfort and control. 

• Autoclavable handpiece for infection control. 

• Durable, lightweight handpiece cord.

• Two methods of irrigation are available (dental unit/city water and irrigation bottle). The selection of no  
   irrigation is also an option.

• Irrigation tube is isolated within the control unit to prevent solution from coming in contact with the  
   electronics of the unit.

• Two irrigation bottles for use with the IQ 4000 unit.

• Easy to clean design to minimize the risk of cross-contamination.

COMPONENT PARTS

  Control Unit
  Symmetry IQ Fiber-optic Handpiece
  Replacement Bulb
  Water Tube
  Symmetry IQ Fiber-optic Handpiece Cord
  Spanner Wrench (5X8) – 2 pieces
  Foot Control
  AC Power Cord
  Irrigation Connector (AC 120V Only)
  Irrigation Bottles – 2 pieces
  O-rings for irrigation bottles (not pictured)
  Tip Wear Guide (not pictured)
  Power Range Guide (not pictured)
  Three Symmetry S-Series Tips (not pictured)

1.
2.
3.
4.
5.
6.
7.
8.
9.

10.
11.
12.
13.
14.

Summary of Contents for Symmetry IQ 4000 Series

Page 1: ...operator fatigue Soft textured silicone finger grip for clinical comfort and control Autoclavable handpiece for infection control Durable lightweight handpiece cord Two methods of irrigation are available dental unit city water and irrigation bottle The selection of no irrigation is also an option Irrigation tube is isolated within the control unit to prevent solution from coming in contact with t...

Page 2: ...tions are not followed NOTICE General information needed to operate the device safely WARNING To avoid electrical shock do not plug or unplug the power cord with wet hands To minimize risk of short circuit and electric shock prevent water from contacting the control unit Subjecting handpiece to a strong impact or drop onto a hard surface could result in electric shock Touching the back end of the ...

Page 3: ...tip carrier See figure 11 Attach the handpiece before turning on the power Do not drop or otherwise expose the control unit to excess trauma Use Hu Friedy Symmetry S Series tips to ensure proper performance Do not autoclave any parts unless indicated so in the User Manual Sterilization by ultraviolet light may cause discoloration of the handpiece If water is spilled on the irrigating pump cover wi...

Page 4: ... commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC61000 4 11 5 Ut 95 dip in Ut for 0 5 cycle 5 Ut 95 dip in Ut for 0 5 cycle 40 Ut 60 dip in Ut for 5 cycles 40 Ut 60 dip in Ut for 5 cycles 70 Ut 30 dip in Ut for 25 cycles 70 Ut 30 dip in Ut for 25 cycles 5 Ut 95 dip in Ut for 5 sec 5 Ut 95 dip in Ut for 0 5 cycle 5 Ut 95 dip ...

Page 5: ...ansmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80MHz and 800MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from...

Page 6: ...Group 1 The Symmetry IQ 4000 Series uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR11 Class B The Symmetry IQ 4000 Series is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply n...

Page 7: ... power of transmitter W Separation distance according to frequency of transmitter m 150kHz to 80MHz d 1 2 P 80MHz to 800MHz d 1 2 P 800MHz to 2 5GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance in meters m can be estimated using the equation applicable t...

Page 8: ...AUTION Be sure to insert the water tube deep into the connector on the control unit If water has not been used recently at the water outlet of the dental unit discolored water may come out Wait until the water coming out of the dental unit is clean before connecting to the control unit Use of Irrigation Bottle Remove the dust cover from the base unit Fig 4 Remove the cap of the irrigation bottle a...

Page 9: ...ndpiece holder on the control unit To remove handpiece from the handpiece cord first remove tip then firmly grip the front section of the handpiece in one hand and the rear section of the handpiece cord plug with the other hand and pull to separate Do not twist to separate WARNING Do not touch the back end of the handpiece where electrical connections are made to the cord It may result in electric...

Page 10: ...e Guardian wrench until you hear clicking When using the Guardian wrench without a tip previously inserted into the tip carrier take the following step place the Guardian wrench over the tip by slipping tip through the small hole in the bottom closed end of the tool This will position the tip within the Guardian wrench for safety and convenience 3 Tip Removal Hold the handpiece vertically Insert t...

Page 11: ...r would be detrimental to the dental procedure being performed and should only be used for a few seconds at a time 7 Power Range Setting WARNING Do not exceed recommended power range indicated on the Symmetry Power Range Guide Higher than recommended power may damage tooth structure and or tip Set the recommended power level with the power key on the front panel The light in the power display pane...

Page 12: ...e the foot control The safety feature is now reset Depressing the foot control repeats the same operation If the power switch to the control unit is turned on while the foot control is depressed accidentally the control unit will beep to alert user PRESET MEMORY FUNCTION The control unit can be preset for water mode power setting and operating mode To set the desired preset conditions into memory ...

Page 13: ... containing phenols Do not spray the cleaning solution directly on the handpiece 2 Wipe debris off the end of the optic fibers at the handpiece with an alcohol soaked cotton swab 3 Place the handpiece in a cassette or sterilization pouch and seal 4 Autoclave the handpiece for 15 min at 273 F 134 C Procedures for Symmetry Piezo Tips 1 Clean disinfect inspect and steam sterilize before each use 2 Ti...

Page 14: ...er adjustment knob at less than maximum during auto cleaning will result in incomplete cleaning in the 30 second cycle 8 When the auto cleaning sequence completes the unit will revert to the settings prior to cleaning 9 Remove the irrigation bottle by pulling straight up 10 After use please thoroughly clean and dry all bottles prior to next use Manual Cleaning Method Cleaning of Irrigation Tube Us...

Page 15: ...rs can be slippery and cause improper operation To prevent damage to the pump rollers use a gentle pushing motion to insert the replacement irrigation pump onto the drive shaft Do not twist Run the replaced irrigation pump about 10 seconds on highest water volume setting before operation to adopt irrigation tube to new pump Changing Water Filter Close the water valve of the dental unit to which th...

Page 16: ...iece sleeve with the same mark on the handpiece cord Push the sleeve to the cord until it clicks Fig 28 CAUTION Do not touch glass part of new lamp and exercise care when mounting a new lamp to avoid dislodging the o ring from the groove WARNING Do not use the device if it appears to be defec tive Note If a failure occurs contact the supplier of your Sym metry IQ product Repair by unauthorized per...

Page 17: ...ctice for the Symmetry IQ 4000 a system or device must be installed to ensure that no contaminated water can flow or be sucked back into the drinking water supply When installing such a system please observe that the requirements of the EN1717 Regulations are adhered to HELPFUL INFORMATION The use of higher than recommended power setting may cause excessive wear or tip breakage Always remember to ...

Page 18: ...The tip is not tightened properly Tighten the tip with the Guardian wrench until you hear a click Noise from the handpiece Output has not been properly adjusted for the tip Adjust the power setting to the recommended power range indicated on the Guardian wrench or package insert The tip is not tightened firmly Tighten the tip with the Guardian wrench until you hear a click Vibration failure in the...

Page 19: ...set too low Adjust the liquid flow water knob Connections are not secure Check the connections between the handpiece and the handpiece hose Check the connection to the unit Check the connection to the foot control Unit in Dry Mode Change water selection Irrigation Bottle is not properly engaged Adjust bottle to be properly seated Water filter is blocked Check the water filter for blockage Fiber op...

Page 20: ...lations in force uses other than those specified in this manual non compliance with instructions contained in this manual service to the Unit being provided by non authorized persons components exhibiting normal wear from usage damages defects being caused by force majeure or any other condition that is beyond the control of Hu Friedy Repair service or replacement provided under the guarantee neit...

Page 21: ...Fax Email Address City State Province Zip Postal Code Country Number of Practitioners In Practice Dentists Hygienists Assistants Dental Specialty Dealer Name U S and Canada Europe Symmetry IQ 4000 Registration Symmetry IQ 4000 Registration Hu Friedy Mfg Co Inc Hu Friedy Mfg Co Inc 3232 North Rockwell Street Zweigniederlassung Deutschland Chicago IL 60618 Rudolf Diesel Str 8 USA D 69181 Leimen ...