Instruction Manual (Operation)  

LAQUAtwin CA-11 / K-11 / NA-11 / NO3-11 

 
Part Names

 

1.  Sensor 
2.  Sensor cover 
3.  Liquid Junction 
4.  Membrane 
5.  Lithium batteries 
6.  MEAS key 
7.  ON/OFF key 
8.  CAL key 
9.  Waterproof gasket 
10.  Strap eyelet 

 

Attach the sensor 

1.  Power OFF the meter 
2.  Confirm that the 

waterproofing gasket is 
clean and undamaged. 

3.  Slide the sensor onto the 

meter, catch "A" on the meter should fit into hole 
"a" on the sensor. 

 

Be careful not to twist the waterproof gasket. 

 

Detach the sensor 

1.  Power OFF the meter. 
2.  Lift the sensor tongue tip and 

slide the sensor a little away from 
the meter. 

3.  Pull out the sensor all the way 

from the meter. 

 

Insert batteries 

1.  Power OFF the meter. 
2.  Detach the sensor. 
3.  Slide both batteries into the battery 

case as shown. 

4.  Attach the sensor. 
Use 2x CR2032 batteries, place the 
plus sides (+) upwards. 

 

Remove batteries 

1.  Power OFF the meter. 
2.  Detach the sensor. 
3.  Use a ball-point pen or other tool 

to pry the batteries out from the 
clips. 

4.  Attach the sensor. 
 

Electrode conditioning 

 

Perform electrode conditioning before using the 
sensor for the first time or after several days of 

disuse. 

 

Perform calibration after conditioning. 

 

1.  Power OFF the meter. 
2.  Place some drops of 2000 ppm standard solution 

on the sensor. 

3.  Wait a few hours. 
4.  Dispose of the standard solution. 
5.  Rinse the sensor with DI or tap water. 
 

Power on / off 

1.  Press 

ON/OFF

 key to switch the 

meter on or off. 

  

 

Keys must be pressed for 1 second to 
function. 

Settings menu 

Customize the meter to the specific needs. 
 

Operating keys functions 

 

MEAS

:

 

Select options in the menu.

 

 

CAL 

: Confirm settings, go to next option.

 

 

ON/OFF

: Escape without saving.

 

 
Open settings menu 

1.  Power OFF the meter. 
2.  Simultaneously press and hold 

MEAS 

and 

ON/OFF

 keys (approx. 3 seconds). 

 

 

Set units of measurement 

Unit options vary with different meter models. 
3.  Press 

MEAS

 to scroll 

4.  Press 

CAL

 to confirm and go to the next setting.  

 

 
Set Low calibration point 
Set the value to match the standard solution. 
Value can be set from 5 to 990ppm. 

5.  Press 

MEAS

 to scroll. Keep pressed to increase 

scrolling speed  

6.  Press 

CAL

 to confirm and go to the next setting 

 

Set High calibration point 

Set the value to match the standard solution. 
Value can be set from 160 to 9990ppm. 
7.  Press 

MEAS

 to scroll. Keep pressed to increase 

scrolling speed  

8.  Press 

CAL

 to confirm and go to the next setting 

 

 

 
Stability mode

 

Choose A.S. (Auto Stable) or A.H. (Auto Hold) 
A.S. mode:   appears at stable value. If the value 
changes   disappears until new stability is reached 
(continuous measurement) 
A.H. mode:   appears at stable value and holds it in 
the display. Press MEAS to do another measurement. 

9.  Press 

MEAS

 to scroll.  

10.  Press 

CAL

 to confirm and go to the next setting. 

 

 
Multiplying compensation

 

Apply compensation coefficient (0.01 to 9.90) to to the 
measured value. The compensated result is displayed 
as the measured value. 
11.  Press 

MEAS

 to scroll. Keep pressed to increase 

scrolling speed 

12.  Press 

CAL

 to confirm and go to the next setting. 

 

 
Backlight on / off 

13.  Press 

MEAS

 to scroll. 

14.  Press 

CAL

 to confirm and go to 

END

. Meter will 

shut down automatically. 

 

 

 
 
 
 

Calibration 

 

Calibration is required before measurement. 

 

Use standard solutions within the measurement 
range in the specifications. 

 

Calibration values are saved even if the meter is 
switched OFF. 

 

Calibration value is rewritten if calibration is 
repeated using the same standard solution. 

 

Calibrate with solutions close to the expected value. 

 
1.  Open the sensor cover. 

 

2.  Rinse the  sensor with DI – or tap 

water. 
 

3.  Dry the sensor with a tissue.  
 
 

 

Rinse again with some of the next standard solution 
for increased accuracy 

4.  Cover the complete sensor with 

standard solution. 
 

5.  Close the sensor cover. 

 

6.  Switch on the meter. 

 

7.  Press 

CAL

 key. Set calibration 

value blinks and 

 is shown. 

       
 

8.  Press 

CAL

 to calibrate or 

MEAS

 to 

select the other calibration point. 

 

and   blink.  
After calibration complete, 

 and 

  stop blinking and the measured 

value is displayed. 

The 1

st

 point has now been calibrated. 

9.  Repeat step 1 to 8 for the 2

nd

 calibration point. 

 

Clear Calibration data 
Clear calibration data in the following cases: 

 

If the number of previous calibration points is 
uncertain 

 

After the sensor is replaced 

 

When Er4 persists 
 

1.  Power OFF the meter. 
2.  Simultaneously press and hold 

MEAS 

and 

CAL

 

keys (approx. 3 seconds). 

3.  Press 

CAL

 to confirm or 

ON/OFF

 to cancel 

 

Temperature adjustment 
Only available in A.S. Stability mode, see Setting 
Menu. Normally this is not necessary. Adjust the 
temperature only when the value is not correct. 
 

1.  Prepare a reference thermometer. 
2.  Allow the meter and thermometer to reach the 

same temperature. 

3.  Make sure you are in A.S. mode 
4.  Press 

MEAS

 to view temperature 

5.  Press 

CAL

, value blinks 

6.  Press 

MEAS

 to adjust the value to match the 

thermometer. 

Press 

CAL

 to confirm. 

 and 

ºC

 will blink until 

adjustment is done. 
 
 
 

Note 

IMPORTANT 

ppm (default) 

mg/L 

kg/10a 

 

(default)

 

 

 

(default) 

(default) 

Tip 

Waterproof 

gasket 

 

 

Backlight ON 

(default) 

(default) 