background image

Instructions for use

QC02 LED REF 055501

Summary of Contents for QC02 LED

Page 1: ...Instructions for use QC02 LED REF 055501...

Page 2: ...Distributed by Faro S p A Via Faro 15 20876 Ornago MB Italy Phone 39 039 6878 1 Email info faro it www faro it Legal manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Page 3: ...4 5 4 Pulling the handpiece off 15 6 Checking for malfunctions and troubleshooting 16 6 1 Checking for malfunctions before initial start up 16 6 2 Troubleshooting 16 6 2 1 Replacing the LED lamp 16 6...

Page 4: ...hnical Service If you have any technical questions or complaints please contact the KaVo Technical Service 49 0 7351 56 1000 service instrumente kavo com Target group The instructions for use are inte...

Page 5: ...symbol Transportation and storage conditions temperature range Transportation and storage conditions air pressure Transportation and storage conditions Humidity Protect from moisture Keep dry Protect...

Page 6: ...User instructions 6 28 WARNING In cases which if not prevented can lead to death or severe injury CAUTION In cases which if not prevented can lead to minor or moderate injury CAUTION In cases which if...

Page 7: ...cessories appropriately before disposal 2 2 Electrical shock Connecting a non KaVo system to the medical device can lead to electrical shock and injury to patient user and third parties 4 During insta...

Page 8: ...7 Service and repair Repairs servicing and safety checks may only be performed by trained service personnel The following persons are authorised to do this Service technicians of KaVo branches after t...

Page 9: ...8 Disposal 9 28 2 8 Disposal This product is subject to the EC directive governing waste electrical and elec tronic devices and must be submitted to special disposal in Europe 4 For further informatio...

Page 10: ...oduct are not permitted and can be hazardous The medical device is designed for coupling of the standard supply hose DIN EN ISO 9168 to MULTIflex compatible dental handpieces turbine handpieces air mo...

Page 11: ...he Instructions for Use of the product to be connected to the QC02 LED coupling Check the suitability of the product for operation with the QC02 LED coupling NOTE Operation on KaVo units 4 1042 on the...

Page 12: ...ns CAUTION Startup after refrigerated storage Malfunction 4 Prior to startup strongly refrigerated products must be allowed to warm up to a temperature of 20 C to 25 C 68 F to 77 F Temperature 29 C to...

Page 13: ...ct and ac cessories WARNING Dispose of the product in appropriate manner Infection hazard 4 Reprocess the product and accessories before disposal Also refer to 7 Reprocessing steps in accordance with...

Page 14: ...of the hose 4 Re tighten with the enclosed wrench 4 Apply a small amount of KaVo Spray to the O rings on the supply hose The coupling remains screwed onto the hose 5 2 Removing the coupling from the h...

Page 15: ...01 5 Operation 5 4 Pulling the handpiece off 15 28 4 Check the secure fit of the LUX instrument on the coupling by pulling on it 5 4 Pulling the handpiece off 4 Grasp the coupling and pull the instrum...

Page 16: ...y 180 and insert it Also refer to 6 2 1 Replacing the LED lamp Page 16 6 2 Troubleshooting WARNING Repair WITHOUT using KaVo original spare parts Parts such as the cover can come loose and cause injur...

Page 17: ...Case 1 KaVo MULTI LED lamp is on Case 2 KaVo MULTI LED lamp is faint 4 Increase the cold light intensity on the unit until the desired illuminance is reached Case 3 KaVo MULTI LED lamp is red or off 4...

Page 18: ...EF 055501 6 Checking for malfunctions and troubleshooting 6 2 Troubleshooting 18 28 4 Push the O ring to the front and remove it 4 Spray a lint free cloth with KaVo Spray apply it to the new O rings a...

Page 19: ...aVo recommends the following products based on the compatibility of the ma terials The microbiological efficacy must be ensured by the disinfectant manu facturer and proven by an expert opinion Approv...

Page 20: ...n a washer disinfector only NOTE Adapters are needed for automated cleaning Order adapter separately Also refer to 8 Components and miscellaneous Page 23 7 3 1 pre cleaning CAUTION Never reprocess thi...

Page 21: ...fector and the VARIO TD programme and a mild alkaline cleaner made by Dr Weigert In addition KaVo recommends the use of a rinsing agent 4 For programme settings as well as cleansers and disinfectants...

Page 22: ...iliser after the steril isation cycle The medical device has a max temperature resistance of up to 138 o C 280 4 o F Sterilisation parameters Select a suitable process from the following sterilisation...

Page 23: ...ealers Components Description KaVo material number KaVo MULTI LED lamp 1 007 5372 Miscellaneous Description KaVo material number Key 0 411 1563 Spare washer 0 553 5262 O ring 6 65x0 8 1 004 2776 O rin...

Page 24: ...le for defects and consequences thereof that have arisen or may arise from natural wear improper handling cleaning servicing or main tenance non compliance with operating maintenance or connection ins...

Page 25: ......

Page 26: ......

Page 27: ......

Page 28: ...1 015 3097 bd 20230726 02 en...

Reviews: