background image

E2610 series gas detectors are compact and easy-to-use instruments.
The  devices  utilise  novel  fully  calibrated  and  temperature  compensated  gas
sensors with excellent repeatability, stability and long lifetime.
Two  relays  with  switch-over  contact  may  be  used  for  remote  signalling  or
ventilation  control.  Flashing  LED  and  an  internal  buzzer  give  alarms  at  two  set-
points.

Safety requirements

Always adhere to the safety provisions applicable in the country of use. Do not
perform any maintenance operation with the power on. Do not let water or foreign
objects inside the device.

Operating conditions

The device should be used in explosion-safe (non ATEX -rated) indoor areas at

the atmospheric pressure ±10% and 15...90 %RH without condensation

Avoid  exposure  to  highly  corrosive  gases  (H

2

S,  SO

2

,  HCl,  Cl

2

  etc),  or  high

concentrations of basic gases, such as ammonia

Avoid mechanical shock or strong vibrations.

Avoid sources of electromagnetic interference

See Specifications table for more details

Installation

There  are  no  precise rules or standards  to follow  when installing the gas detectors.
The following points must be taken into account:

application (air quality control or leakage detection),
properties of the space under investigation (room geometry, direction and velocity of
air flows etc),
detected  gas  (relative  density  to  air,  whether  the  gas  is  flammable,  or  toxiс,  or
oxygen displacing,  whether the gas is heated),
safety:  strong  vibrations,  mechanical  shock,  and  the  sources  of  strong
electromagnetic interference should be avoided,
the device should be accessible for maintenance and repair.

For  early  leakage  detection  install  the  sensor  as  close  as  possible  to  the  potential
leakage  sources  (flanges,  valves,  pressure  reducers,  pumps,  etc),  taking  into
consideration  other  points  listed  above.  Do  not  locate  the  detector  close  to
ventilation openings and strong air currents. Avoid the areas without air circulation
(corners,  niches)  as  well.  For  general  area  monitoring  without  definite  leakage
sources,  the  detectors  should  be  distributed  evenly  in  the  room.  For  personal
safety control the detectors are installed in the breathing zone (at the height of the
head  of  people  or  animals).  Recommended  sensor  position  is  vertical,  pointing
downwards. See Installation guidelines section for more information.
The gas detector should be mounted on a wall at a place located in proximity to
possible gas source. Since carbon monoxide has practically the same density as
air, the detector should be installed in the breathing zone  (1.5 m above the floor,
the  height  may  vary  depending  on  application)  or  near  the  potential  leakage
source.  Do  not  locate  the  detector  close  to  ventilation  openings  and  strong  air
currents.  Avoid  the  areas  without  air  circulation  (corners,  niches)  as  well.  The
device should be available for maintenance and repair.

Connections

1. Detach the base of the enclosure by gently pulling along four guiding pins.

2. Attach the base to the wall with two screws. (see drawings below).

85

37

8

5

63

3. Use rounded cutouts on the side of the base to let in the cables from the power
supply and of the external devices as shown below.

For easier connection, terminal blocks of E2610 series devices are removable. To
dismount the terminal block, pull it off from the PCB.
Connect the power terminals N and L to the 24 V AC/DC source if you are using
detector version -24 or to 230 VAC mains if you are using detector version -230
(see diagram below).
The terminals on the E2610 series devices are suitable for a wide range of wires
with cross-section 0,2...1,5 mm

2

. We recommend to strip the wire end by 5...6 mm

and tin it, or to use the wire end sleeves. To connect wire, loosen the screw, insert
the wire end into terminal hole and tighten the screw.

RE1

RE2

N

L

PSU

RE1

RE2

s

e

n

s

o

r*

x5

buzzer

*Sensors may have different shape and/or size

Connect  external  devices.  Relay  switch-over  outputs  may  be  used  to  control
directly  24  V  or  230  V  (for    versions  -24  and  -230  respectively)  powered  alarm
sirens, ventilation fans, shut-off valves or other actuators.

Attach terminal blocks to the board

4. Push enclosure to the base.

5. Turn on the power. It will take ca. 60 seconds after switching on for the sensor
to stabilize.

Operation

During  the  first  ca.60  seconds  after  powering  on  E2610 performs  a warming-up
and self-diagnostic routine, indicated by the flashing of each LED. The upper dual-
color LED remains continuously green in normal operation and blinks red in case
of  device  or  sensor  fault.  The  warm-up  time  depends  on  the  sensor  type,
unpowered period and atmosphere.
During the first 30 seconds after powering on you may select automatic or manual
mode of alarm release. By shortly (< 2 s) pressing the button on the device’s front
panel  you  enable  the  automatic  mode,  by  pressing  the  button  for  2...10 s  —
manual mode. The activation of the automatic mode is followed by a single LED
blinking and acoustic signal. If manual mode is activated, double acoustic and light
signal follows.
If gas concentration exceeds the LOW alarm setpoint, the bottom red LED starts
flashing at a rate of 1 Hz, and the relay RE1 switches over. The first alarm stops
automatically  if  the  gas  concentration  drops  below  80%  of  the  LOW  alarm
setpoint.
If gas level exceeds the HIGH alarm setpoint, the bottom red LED starts flashing
and the buzzer starts beeping at a rate of 2 Hz, and also the relay RE2 switches
over.  Depending  on  the  selected  release  mode,  the  HIGH  alarm  stops
automatically or can be stopped by pressing the button, on condition that the gas
level has dropped below 80% of the LOW alarm setpoint.
Beyond  the  warm-up  period,  holding  down  the  button  for  2...10  seconds  and
releasing causes E2610 to reset and perform the self-diagnostic routine for testing
purposes.
When holding the button down for over 10 seconds, E2610 imitates the reaching
of the HIGH setpoint with the respective light and sound indication and switching
over the relays.

Warranty

This product is warranted to be free from defects in material and workmanship for
a  period  of  one  year  from  the  date  of  original  sale.  During  this  warranty  period
Manufacturer will, at its option, either repair or replace product that proves to be
defective.  This  warranty  is  void  if  the  product  has  been  operated  in  conditions
outside  ranges  specified  by  Manufacturer  or  damaged  by  customer  error  or
negligence or if there has been an unauthorized modification.

Reviews: