Self-Monitoring Blood Glucose System 

 

 

Quick Reference Guide 

 
 

   

 

Содержание MDT2

Страница 1: ...MDT2 Self Monitoring Blood Glucose System Quick Reference Guide...

Страница 2: ...battery door on the back of the meter by pushing the tab in the direction of the arrow 2 Insert two batteries The meter will beep to confirm the batteries are inserted correctly 3 Put the battery door...

Страница 3: ...s the wrong unit please contact your local dealer for repair 3 Repeat step 2 to set the date and time The flashing field is the one you are currently setting 1 After inserting batteries the meter turn...

Страница 4: ...emove the control solution bottle cap and wipe the tip of the bottle with a tissue If you do not choose CtL the test results will be included in the memory of blood test 1 Put a test strip into the me...

Страница 5: ...ve the test strip and throw it away after you have compared the reading to the range printed on the test strip bottle 8 When the icon of shows the meter beeps you have enough control solution for the...

Страница 6: ...cet into the lancing device to get it ready for use 1 Unscrew the Cap 2 Insert the lancet into the lancing device firmly then twist off the protective cover 3 Recap the front cap Lancets are for use o...

Страница 7: ...not to be shared between users Sharing lancing devices and lancets may transmit blood borne pathogens such as viral hepatitis 6 Set the lancing device aside until later in the test 5 Pull on the slidi...

Страница 8: ...se and dry thoroughly 4 When the blood drop flashes on the display obtain a drop of blood from your fingertip 5 Gently squeeze your finger to assist the flow of blood 2 Insert a test strip into the me...

Страница 9: ...the strip Be sure to get enough blood on the strip s reaction zone Otherwise an inaccurate reading may result 8 When the icon of shows the meter beeps sufficient blood is applied to the strip s react...

Страница 10: ...d hold the device with a clear adjustable tip against the forearm 4 Press the trigger button to activate the lancing device Hold the device against forearm and increase pressure until the blood sample...

Страница 11: ...ld the device against palm and increase pressure until the blood sample size is sufficient 1 Massage the puncture area of palm for a few seconds Running A Blood Glucose Test With Blood From Your Palm...

Страница 12: ...the cap 3 Pointing the lancing device toward a container for sharp or biohazard material slide the ejection button up to release the covered lancet into the container 2 Without touching the used lance...

Страница 13: ...or 2 seconds 1 Press OK and then to review AC PC average values of 30 days under and The average value of 7 14 30 90 day 2 Press OK to review all results in order from records of 480 to 001 3 To delet...

Страница 14: ...er can be turned on or off 2 The meter starts to transfer data to the GlucoManagerTM software 3 When all data are uploaded from the meter to the PC the meter shows OK and beeps 1 Use the USB cable to...

Страница 15: ...e contact your local dealer Message What it means What to do Humidified Used strips The meter has detected a problem with the test strip Repeat the test with a new strip Refer to pages 8 9 for informa...

Страница 16: ...place the batteries first If ERROR 005 appears again please contact your local dealer System error There may be a problem with the meter Replace the batteries first Refer to pages 2 If this error mess...

Страница 17: ...gain obtain and follow instructions from your healthcare professional without delay The test result is lower than 20 mg dL 1 1 mmol L This may require immediate treatment according to your healthcare...

Страница 18: ...result may occur if the test continues Relocate the meter to a location with temperature between 10 C 40 C 50 F 104 F The Lt and thermometer icon appears Temperature is too low outside the required r...

Страница 19: ...y to the mode of control solution Re check your glucose level If the result appears again obtain and follow instructions from your healthcare professional without delay Volume detector error The volum...

Страница 20: ...ting button is on the back side of the meter Press the button for 3 seconds to get back to the initial status No result in memory The meter doesn t run any test yet You can still perform a blood gluco...

Страница 21: ...ence Park Hsinchu Taiwan 30077 TEL 886 3 6686868 FAX 886 3 6686866 E mail info epsbio com tw Website http www epsbio com Obelis s a Boulevard G n ral Wahis 53 B 1030 Brussels Belgium TEL 32 2 732 59 5...