BA_EE360_e // V2.0 // Technical data subject to change // 194475

Operation manual

EE360

Moisture in Oil 

Transmitter

Summary of Contents for EE360

Page 1: ...BA_EE360_e V2 0 Technical data subject to change 194475 Operation manual EE360 Moisture in Oil Transmitter...

Page 2: ...and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the...

Page 3: ...arm module option AM2 14 4 2 Integrated power supply 100 240 V AC option AM3 16 4 3 RS485 Module Modbus RTU option J3 16 4 4 Ethernet Module Modbus TCP option J4 17 4 4 1 Available TCP and UDP ports 1...

Page 4: ...ng head The replacement of the filter cap shall be done with utmost care not to touch the sensors Installation electrical connection maintenance and commissioning shall be performed by qualified staff...

Page 5: ...n the Modbus RTU interface An optional relays module can be used for alarms and process control The state of the art TFT colour display can show all measurands simultaneously and offers extensive erro...

Page 6: ...activity aw is the relative measure for moisture in oil It represents the ratio between the actual amount of dissolved water and the maximum possible amount of dissolved water in the oil at a certain...

Page 7: ...2 mm 0 2 not included in the scope of supply 54 2 13 104 4 09 71 2 80 5 1 0 20 166 6 54 180 7 09 153 6 02 54 2 13 166 6 54 5 1 0 20 7 0 28 Fig 4 Dimensions and mounting template of polycarbonate enclo...

Page 8: ...to the mounting template below Mount the bottom part of the enclosure with 4 screws screw diameter 4 2 mm 0 2 not included in the scope of supply Stainless steel enclosure 75 2 95 110 4 33 73 2 87 5...

Page 9: ...nd scaling are freely selectable All settings can be performed via display and push buttons or using the EE PCS Product Configuration Software Error indication The analogue outputs feature an error in...

Page 10: ...g PE n c phase L1 phase L1 n c Power supply 100 240 V AC for polycarbonate enclosure Power supply 100 240 V AC for stainless steel enclosure 2 1 4 3 5 RS485 B D GND RS485 shielded RS485 A D n c Modbus...

Page 11: ...ss 3 3 1 Installation of the probe directly in the process see Fig 10 Installation of the EE360 probe directly into the process For direct probe installation into a duct stop valves should be provided...

Page 12: ...may deviate by max 40 C 72 F from the regular tempera ture during normal operation Installation of the probe see Fig 11 Installation of the probe by utilizing the ball valve set Install the probe into...

Page 13: ...ly expulsed when releasing the lock not May sure nut to bend and damage the probe cable Loosen slowly the lock nut with a spanner spanner width 24 only till the probe is pushed out by the overpressure...

Page 14: ...C 2 NO NC C Standard Inverted REL 1 REL 2 RELAY 2 RELAY 1 PROTECTIVE CONDUCTOR TERMINAL Maximum wire cross section 1 5 mm Fig 12 Alarm module option AM2 The metal housing must be grounded during opera...

Page 15: ...Configuration Software or using the display and the push buttons Menu Alarm number Alarm output 1 Alarm 2 Alarm 1 Mode Switch hystersis Measurand Switching point Switching point Hysteresis Standard I...

Page 16: ...st excess current and short circuit must be designated to a wire cross section of 0 8 mm AWG 18 6A fuse The metal housing must be grounded during operation National regulations for installation must b...

Page 17: ...Ethernet module features power over Ethernet PoE and RJ45 connector with IP65 protection class The Ethernet connection must be disconnected before opening the housing Due to strong self heating effect...

Page 18: ...otocol format as payload data See Ethernet communication protocol which can be found on https www epluse com en ser vice support download center 4 4 2 IPv4 Settings Factory setting IP Address 192 168...

Page 19: ...0 22 or higher Firmware for Display V 1 0 9 or higher Order code Ethernet Module for retrofitting plastic enclosure HA010606 for remote probe type T5 T10 HA010607 for duct mounting type T2 4 5 Modbus...

Page 20: ...ic unit 1 2 Barcode for probe serial number Fig 19 Pluggable probe Probe replacement procedure Plug off the damaged probe Plug on the new probe Update the 7 parameters with EE PCS Product Configuratio...

Page 21: ...over and plug in the USB cable PCB LED 1 LED 2 LED 4 LED 3 Current output Error LED error LED on error category 1 error LED flashes error category 2 Voltage output power supply main board Fig 21 Statu...

Page 22: ...the user from 1 second to 12 hours The data logger setup can be performed via display and push buttons Menu Sampling rate Data logging OFF 1 s 10 s 1 min 5 min 15 min 30 min 1 h 5 h 12 h The data logg...

Page 23: ...g 23 Data logging Each point in the graph represents a logged value The points are connected by a linear interpo lation One point The data logging continues even when the data memory is full new data...

Page 24: ...scaling error indication relay configuration mode set points state adjustment 1 and 2 point humidity temperature adjustment reset to factory adjustment calibration status configuration of Modbus data...

Page 25: ...e 9 x 10 x not available with 0 1 V output Tab 2 Overview of error codes Error category 1 non critical error can be solved by the user The display blinks and the buzzer beeps every 10 seconds The red...

Page 26: ...n the lock of the filter cap Remove the filter cap from the tip of the probe with utmost care not to touch the sensors Solder off the damaged humidity sensing element Shorten the leads of the replacem...

Page 27: ...an be performed in the E E Elektronik calibration laboratory For infor mation on the E E capabilities in ISO or accredited calibration please see www eplusecal com Calibration and adjustment by the us...

Page 28: ...TMENT NEW ADJUSTMENT Fig 28 Offset adjustment measurement value 1 Measured value Displayed value reference value 1 ADJUSTMENT POINT LOW ADJUSTMENT POINT HIGH reference point LOW factory setting 30 RH...

Page 29: ...be material Stainless steel 1 4404 AISI 316L Enclosure material for plastic enclosure Polycarbonate UL94 V0 approved for metal enclosure Stainless steel 1 4404 AISI 316L Protection class IP65 Cable gl...

Page 30: ...for retrofitting plastic housing HA010606 for remote probe type T5 T10 HA010607 for duct mounting type T2 1 Only for devices with PC4 option 2 2 pieces necessary per device 3 For upgrade to Modbus RTU...

Page 31: ...settings Status Menu only available with connected modules Detailed information Menu Sampling rate Data logging Axis 1 Measurand Axis 2 Measurand Scale maximum value value value value Scale minimum OF...

Page 32: ...ature C Temperature F Temperature K Mode Current Voltage ON Output number Output 1 Output 2 OFF Measurand min max value OK value Cancel Analog range Scaling Error indication OFF Water activity aw Wate...

Page 33: ...asuring system Working pressure Parameters Humidity coefficient Resistance R0 Capacity C76 Capacity gain Capacity offset Temperature coefficient TK Resistance Offset Daylight saving time UTC settings...

Page 34: ...point Switching point Hysteresis Standard Inverted Normal state Switch window Measurand Switching point low Switching point Hysteresis Switching point high Switching point Hysteresis Standard Inverted...

Page 35: ...35 Operating instructions for EE360 Moisture in Oil...

Page 36: ...cn BEIJING Tel 86 10 84992361 info epluse cn SHANGHAI Tel 86 21 61176129 info epluse cn E E GERMANY www epluse de Tel 49 6172 13881 0 info epluse de E E FRANCE www epluse fr Tel 33 4 7472 35 82 info...