Installation Guide & Owner’s Manual

Direct-to-Drain Thermal Reduction  

System for Autoclave Wastewater

Model 

NSVC-2                 

For 

2

 

Autoclaves

VistaCool can be used with:

Midmark/Ritter M3 

SciCan STAT

IM

 2000

Midmark/Ritter M9* 

SciCan STAT

IM

 5000 

Midmark/Ritter M11*

* Requires Model V7502 and S7560 adapter kit

Содержание Vistacool NSVC-1

Страница 1: ...al Direct to Drain Thermal Reduction System for Autoclave Wastewater Model NSVC 2 For 2 Autoclaves VistaCool can be used with Midmark Ritter M3 SciCan STATIM 2000 Midmark Ritter M9 SciCan STATIM 5000 Midmark Ritter M11 Requires Model V7502 and S7560 adapter kit ...

Страница 2: ...ion V7501 6 Typical Installation V7502 7 Plumbing 8 Cabinetry 9 Connecting the System 10 Tank Positioning 12 Starting Using the System 12 Autoclave Specific Instructions 13 Maintenance 14 Troubleshooting 14 VistaCool Limited Warranty 15 VistaCool System Information Specifications 16 39 Western Hwy West Nyack NY 10994 www DuralineSystems com 877 561 0500 ...

Страница 3: ...ture reference and ensure that anyone responsible for operation and maintenance of the system is familiar with all details contained in this manual Please return the Warranty Registration form or register online immediately upon installation IMPORTANT INFORMATION For System Owners Please read this entire manual before proceeding with installation and operation and always follow local plumbing code...

Страница 4: ...ssembly VC250 A VistaCheck Dual Check Backflow Preventer S7506 Drain Line Assembly S7539 Cabinet Grommet x4 TU751 Cold Water Feed Line 1 4 OD x 1 8 ID Blue Transparent Polyurethane 6 TU753 Condensate Drain Line 1 4 OD x 0 170 ID Black Opaque LLDPE 6 TU755 Coolant Overflow Drain Line 3 4 OD x 1 2 ID Clear PVC 6 Tubing TU748 High Temperature Feed Line 1 4 OD x 0 185 ID Teflon 9 For 1 Autoclave S6130...

Страница 5: ...ck Dual Check Backflow Preventer S7506 Drain Line Assembly x2 x2 Tubing 18 TU751 Cold Water Feed Line 1 4 OD x 1 8 ID Blue Transparent Polyurethane 6 TU753 Condensate Drain Line 1 4 OD x 0 170 ID Black Opaque LLDPE 6 TU755 Coolant Overflow Drain Line 3 4 OD x 1 2 ID Clear PVC 6 TU749 High Temperature Feed Line 1 4 OD x 0 185 ID Teflon x2 For 2 Autoclaves S7539 Cabinet Grommet x8 S6130 Plug S6135 A...

Страница 6: ...10 12 11 4 Drain Line Sink Drain S7506 Drain Line Assembly S6119 or S6135 Cold Water Line Fitting 1 2 3 4 S6155 Water Supply Valve 5 VC250 A VistaCheck Backflow Preventer 6 TU751 Cold Water Feed Line 7 S7508 Inline Thermal Sensor Assembly 8 TU753 Condensate Drain Line 9 TU755 Coolant Overflow Drain Line 10 Manifold Coil Assembly 11 Thermal Regulator Assembly 12 4 ...

Страница 7: ...Drain Line Sink Drain S7506 Drain Line Assembly S6119 or S6135 Cold Water Line Fitting 1 2 3 4 S6155 Water Supply Valve 5 VC250 A VistaCheck Backflow Preventer 6 TU751 Cold Water Feed Line 7 S7508 Inline Thermal Sensor Assembly 8 TU753 Condensate Drain Line 9 TU755 Coolant Overflow Drain Line 10 Manifold Coil Assembly 11 Thermal Regulator Assembly 12 4 ...

Страница 8: ...ct the installed fitting from step 1 to the inlet of the inline water valve provided Be certain the valve is in the closed position blue lever 90 degrees to the valve body Cut another 3 to 4 piece of blue poly tubing and connect one end to the outlet end of the inline valve and the other end to the inlet of the supplied CSA certified VistaCheck dual check backflow preventer 1 2 3 To Coolant Overfl...

Страница 9: ...to keep the tubing from being damaged Simply drill a 7 8 pilot hole through the metal or wood partition then push the smooth plastic grommet into the holes Make sure to drill the holes so that the tubing makes a progressively descending angle from the autoclave to the VistaCool The hole entering the cabinet where the VistaCool is installed should be above the top of the tank manifold so the tubing...

Страница 10: ...stall a stainless hose clamp to hold the tubing firmly in place Do NOT use Vaseline or other lubricants other than clean water on the tubing Moisten the tubing ends with water before inserting tubing into push to connect fittings if lubrication is needed Cut the tubing to a sufficient length so that no tubing line is tight or stressed and so the VistaCool tank assembly can be moved within the cabi...

Страница 11: ...old Make sure the tubing does not kink when moving the autoclave or pulling the tubing through the cabinet partition holes 4 Connect the black 1 4 OD LLDPE condensate line tubing to the elbow fitting on the top of the manifold then connect the other end to one of the male adapter fittings on the drain adapter 5 Cut the black condensate line from step 4 as close to the drain adapter assembly as pos...

Страница 12: ...is primed when water can be seen leaving the 1 2 overflow fitting on the tank and running to the drain adapter 4 Set the water tank in an elevated position above the drain connection as described above to ensure proper drain flow The adjustable tank stand accessory S7545 for use with single unit V7501 S7547 for use with double unit V7502 may be needed to securely support the tank at the proper hei...

Страница 13: ...units with green displays can be connected to VistaCool First generation models with red displays 0XX are not compatible The single unit VistaCool V7501 should not be used with Midmark Ritter M9 or M11 units The double unit VistaCool V7502 is capable of handling any two compatible autoclaves Complete installation instructions for the M9 M11 units are included with the S7560 adapter kit IMPORTANT I...

Страница 14: ...eck for kinks in tubing from the autoclave to the VistaCool manifold Check that water is running from the tank coolant overflow line to the drain adapter Make certain water supply valve is on If there is no water to cool the coil s in the tank the inline thermal sensor located in the condensate line will respond by closing the pathway to alert you of the problem and protect the plumbing from steam...

Страница 15: ...ject to the conditions hereinafter set forth Vista Research Group LLC VRG will repair or replace to the original user any portion of a VistaCool which proves defective due to defective materials or workmanship of VRG Contact your nearest authorized VRG distributor dealer for warranty service At all times VRG shall have and possess the sole right and option to determine whether to repair or replace...

Страница 16: ... If you d prefer you may complete and return the included Warranty Registration sheet Make a copy of the form for your records then mail the original to us V750X IGOM 171211 V7501 V7502 Tank Material Polyethylene Natural White Tank Dimensions with fittings 7 6 W x 7 6 L x 13 75 H 19 3 cm x 19 3 cm x 34 9 cm 4 0 W x 13 0 L x 17 0 H 10 2 cm x 33 1 cm x 43 2 cm Tank Volume 1 50 Gallons 5 70 Liters 3 ...