Industrial X-Ray 

 

Cooler Manual XRC-3001-WA 

 

3000 Watt water to air cooler 

 

Summary of Contents for XRC-3001-WA

Page 1: ...COMET AG Herrengasse 10 CH 3175 Flamatt Tel 41 31 744 9000 www comet xray com XRC 3001 WA Doc No 50005550 Page 1 of 13 Industrial X Ray Cooler Manual XRC 3001 WA 3000 Watt water to air cooler...

Page 2: ...ent history Version Date Author Amendment s Status 12 15 2008 St Haferl released 1 0 07 22 2010 M Schmid Layout Structure 07 22 2010 St Haferl released 2 0 05 18 2011 M Schmid Technical Details 3 0 10...

Page 3: ...iption 5 3 Technical data 6 3 1 Physical dimensions 6 3 2 Performance data 6 3 3 Environmental specifications 6 3 4 Settings 6 3 5 Performance data and overview 7 4 Installation 8 4 1 Antifreeze conce...

Page 4: ...ing unit always with correct coolant level otherwise cooling capacity is reduced Clean filter regularly according to maintenance instructions other wise damage of pump may occur Never operate cooling...

Page 5: ...ed by a bypass valve Cooling hoses are connected to the cooling unit via Walther couplings Coolant inlet and outlet are marked with the labels Inlet O Outlet O The water flow is monitored by an adjust...

Page 6: ...sumption 2 6 A Noise level 55 dB A 50 Hz distance 1m in any direction 59 dB A 60 Hz distance 1 m in any direction Airflow at 50Hz 60Hz 2200 m3 h 2600 m3 h Safety class IP 33 3 3 Environmental specific...

Page 7: ...41 31 744 9000 www comet xray com XRC 3001 WA Doc No 50005550 Page 7 of 13 3 5 Performance data and overview Temperature difference forward ambient Cooling capacity in Watt Water Drain Water Outlet E...

Page 8: ...g to the concentration table below Open the lid and fill the cooling unit with the cooling mixture about 4 0 l Notice The liquid level must be always above the fins otherwise the cooling capacity will...

Page 9: ...the filter of the pump for clean condition if necessary more often For this proceed as following Note If the filter is not in clean condition damage of pump and motor may occur Proceed as following D...

Page 10: ...COMET AG Herrengasse 10 CH 3175 Flamatt Tel 41 31 744 9000 www comet xray com XRC 3001 WA Doc No 50005550 Page 10 of 13 6 Outline Drawing...

Page 11: ...41 31 744 9000 www comet xray com XRC 3001 WA Doc No 50005550 Page 11 of 13 7 Circuit Diagrams 7 1 Electrical diagram Attention The cooler has to be fused when used as a stand alone unit with 4A slow...

Page 12: ...COMET AG Herrengasse 10 CH 3175 Flamatt Tel 41 31 744 9000 www comet xray com XRC 3001 WA Doc No 50005550 Page 12 of 13 7 2 Block diagram...

Page 13: ...COMET AG Herrengasse 10 CH 3175 Flamatt Tel 41 31 744 9000 www comet xray com XRC 3001 WA Doc No 50005550 Page 13 of 13 9 Declaration of conformity...

Reviews: