IVAC®598 Volumetric 

Pump

Directions For Use

en

Summary of Contents for IVAC 598

Page 1: ...s IVAC 598 Volumetric Pump Directions For Use en ...

Page 2: ...bol Definitions 6 Operating Precautions 7 Getting Started 9 Loading an Infusion Set 10 Starting the Infusion 11 Basic Features 12 Alarms and Display Messages 14 Flow Sensor Operation Optional 16 Specifications 17 Maintenance 19 CareFusion Dedicated Infusion Sets 21 Trumpet and Flow Rate Curves 22 Spare Parts 24 Service Contacts 25 Document History 26 ...

Page 3: ...ein Open KVO rate Integral ultrasonic air in line detection Detection of upstream downstream occlusions Low occlusion pressures nominal 6 to 8 psi Lightweight and portable with self contained rechargeable battery Automatic flow shutoff with activation of audible and visible alarms Audible and visible low battery alert about one hour before battery alarm Optional flow sensor Model 192 Optional remo...

Page 4: ... infusion rates settings and values are shown in the Specifications section Conventions used in this manual BOLD Used for Display names software commands controls and indicators referenced in this manual for example AC Power Indicator HOLD ON OFF switch Single quotes Used to indicate cross references made to another section of this manual Italics Used to refer to other documents or manuals and als...

Page 5: ... READ VOL CLEAR VOL VOL ml RATE KVO VTBI ON OFF RUN HOLD ml hr Model 598 V o l u m e t r i c P u m p Door CAM Follower Assembly Latch Air In Line Sensor Keyboard Display Indicators AC power Indicator CAM Follower Assembly Pole Clamp PE Connector AC Power Connector Fuse Holders Flow Sensor Connector Test Connector Cover ...

Page 6: ...ator When illuminated the pump is connected to an AC power supply and the battery is being charged VTBI The value displayed is the Volume To Be Infused KVO The pump is infusing at the Keep Vein Open rate of 5ml h or current rate whichever is less VOL The value displayed is the volume infused RATE The value displayed is the current infusion rate ml Millilitres The value displayed is the Volume To B...

Page 7: ... PE Connector k Type CF applied part Degree of protection against electrical shock IPX1 Protected against vertically falling drops of water r Alternating Current s Device complies with the requirements of Council Directive 93 42 EEC as amended by 2007 47 EC T Date of Manufacture t Manufacturer U Not for Municipal Waste W Fuse Rating EC REP Authorised Representative in the European Community ...

Page 8: ... for semi rigid containers should take place after the spiking of the container and priming of the drip chamber Steps for Semi rigid containers 4 5 3 2 Spike the container Fill drip chamber to fill line Prime the set by opening closing the roller clamp 1 Close the roller clamp Open the air vent to allow pressure equalisation ready for infusion Steps for the Collapsible bags Follow steps 1 to 3 as ...

Page 9: ...representative Magnetic Resonance Imaging MRI The Pump contains ferromagnetic materials which are susceptible to interference with magnetic field generated by the MRI devices Therefore the Pump is not considered an MRI compatible pump If use of the Pump within an MRI environment is unavoidable then CareFusion highly recommends securing the Pump at a safe distance from the magnetic field outside th...

Page 10: ... verify that the AC Mains 3 indicator is lit A The Pump will automatically operate from its internal battery if the Pump is switched on without being connected to the AC power supply A Should the pump fail to perform correctly replace in its original protective packaging where possible and contact a qualified service engineer for investigation Pole Clamp Installation A pole clamp is fitted to the ...

Page 11: ...pillage onto the pump Ensure that the tubing is inserted completely into the top set retainer through to the tubing guide avoiding any slack Prime the IV infusion set slowly and close the roller clamp 1 Air In Line Sensor Latch shown open Top set retainer Primed IV Infusion Set Lower Tubing Segment Open the door 2 Load the primed IV infusion set 3 Open the latch Load the upper tubing segment into ...

Page 12: ...d the prompt VTBI will be displayed When operating the pump with a flow sensor and a VTBI is not desired press the L switch until OFF is displayed Use the chevron switches 7 I J L and K to set VTBI Connect the infusion set to the patient access device 8 Press 9 RUN HOLD to start infusion The run indicator will appear when the pump mechanism begins to operate Changing the Infusion Rate Press the 1 ...

Page 13: ...hospital protocol 4 Prepare the new infusion set load infusion set into pump and close the door see Loading the Infusion Set 5 Restart infusion see Starting the Infusion 6 A When changing the infusion set or the fluid container use aseptic technique according to hospital protocol It is recommended that infusion sets are changed in accordance with the Directions For Use Carefully read the Direction...

Page 14: ...op of Needle Free Valve port with 70 Isopropyl alcohol 1 2 1 seconds and allow to dry approximately 30 seconds NOTE Dry time is dependent on temperature humidity ventilation of the area Prime valve port If applicable attach syringe to Needle Free Valve port and aspirate 2 minuscule air bubbles When used with administration sets always refer to individual set directions for use as 3 change interval...

Page 15: ...logged filter Repeated alarms may indicate pump mechanism requires cleaning Refer to the Cleaning and Storage section of this DFU rELOAd Set adapter improperly loaded into air detector Reload adapter into air detector Press firmly until flush rELOAd Air Air detector senses air in tubing or Set adapter improperly loaded into air detector Ensure air detector is clean Reload set adapter into air dete...

Page 16: ...m condition Press RUN HOLD switch once to restart infusion Rate alternating with rate value Pump is in rate setting mode This is not an alarm condition Press RUN HOLD or VOL TO BE INF switch to advance to VTBI setting mode VtBI alternating with VTBI value Pump is in VTBI setting mode This is not an alarm condition Press RUN HOLD switch once to start infusion Prompts Display Cause Action set vtbi V...

Page 17: ...herever possible IVAC Flow Sensor Model 192 Handles Drip chamber Flanges Pull back handles Plug the flow sensor into the flow sensor interface located on the top rear part of the pump 1 Attach the IVAC Flow Sensor Model 192 to the drip chamber of the infusion set by pulling back the handles Refer to the illustration 2 above Proceed with load priming and set up instructions as described in section ...

Page 18: ...acy 5 under specified conditions Note Accuracy specified under the following conditions Head height 30 76 2cm Test solution distilled water Environmental ambient temperature Back pressure 20 gauge needle Infusion Set Model 59073 59073 has been superseded by G59073 which is identical to the previous infusion set A Use of high viscosity liquids or non CareFusion branded sets may adversely affect acc...

Page 19: ...1ml h is nominally 35 minutes Time to alarm at 100ml h is nominally 0 31 minutes Air in Line Accuracy 100 µl 50 µl Maximum Occlusion Alarm Pressure 775 mmHg Nominal 310 mmHg to 413 mmHg Post Occlusion Bolus Bolus volume generated at 1 ml h when the occlusion alarm threshold is reached 0 5 ml Bolus volume generated at 100 ml h when the occlusion alarm threshold is reached 0 5 ml Critical Volume 25 ...

Page 20: ...ith reference to the TSM A Please refer to Technical Service Manual for calibration procedures The units of measurement used in the calibration procedure are standard SI units The International System of Units Battery Operation The internal rechargeable battery allows continued operation when the AC power is unavailable for example during patient transfer or AC power failure A fully charged batter...

Page 21: ...eaning agents as these may damage the exterior surface of the Pump Do not steam autoclave ethylene oxide sterilise or immerse this Pump in any fluid If the Pump has visible cracks or damage to the case do not clean and immediately take out of service for examination by a qualified service engineer The infusion set is a disposable single use item and should be discarded after use according to hospi...

Page 22: ...260cm G59073 No Filter Length 260cm G59093 15 Micron Filter Length 260cm G59103E 2 SmartSite Needle Free Valve Ports 1 Burette 150ml Length 270cm G59173E 1 SmartSite Needle Free Valve Port No Filter Length 260cm G59273E 2 SmartSite Needle Free Valve Ports No Filter Length 260cm G59293E 2 SmartSite Needle Free Valve Ports No Filter Length 260cm G59593 15 Micron Filter Length 270cm G59693E 1 SmartSi...

Page 23: ...cular integration the clinical effect cannot be determined from the trumpet curves alone The start up curves represent continuous flow versus operating time for two hours from the start of the infusion They exhibit the delay in onset of delivery due to mechanical compliance and provide a visual representation of uniformity Trumpet curves are derived from the second hour of this data Tests performe...

Page 24: ...tart up Graph at 999ml h 24 hour Flow Rate l h 0 0 2 0 4 0 6 0 8 1 0 20 40 60 80 100 120 Flow Rate ll h 0 0 2 0 4 0 6 0 8 1 0 20 40 60 80 100 120 Time minutes Time minutes Trumpet Graph at 999ml h Initial Period Trumpet Graph at 999ml h 24 Hour Flow Rate Error D D D D D G G G G G A A A A A 10 8 6 4 2 0 2 4 6 8 10 2 6 10 14 18 22 26 30 Flow Rate Error D D D D D G G G G G A A A A A 10 8 6 4 2 0 2 4 ...

Page 25: ...p is included within the Technical Service Manual The Technical Service Manual 1000SM00016 is available in electronic format on the World Wide Web at http www carefusion com alaris intl A username and password are required to access our manuals Please contact a local customer services representative to obtain login details ...

Page 26: ...87 Fax 46 8 544 43 225 BE DK IT US CareFusion Leuvensesteenweg 248 D 1800 Vilvoorde Belgium CareFusion Firskovvej 25 B 2800 Lyngby Danmark CareFusion Via Ticino 4 50019 Sesto Fiorentino Firenze Italia CareFusion 10020 Pacific Mesa Blvd San Diego CA 92121 USA Tel 32 2 267 38 99 Tlf 45 70 20 30 74 Tél 39 055 30 33 93 00 Tel 1 800 854 7128 Fax 32 2 267 99 21 Fax 45 70 20 30 98 Fax 39 055 34 00 24 Fax...

Page 27: ...1000DF00645 Issue 2 26 26 IVAC 598 Volumetric Pump Document History Document History Revision CO Number Date 1 10649 May 2010 2 10809 July 2010 ...

Page 28: ...ary information of CareFusion Corporation or one of its subsidiaries and its receipt or possession does not convey any rights to reproduce its contents or to manufacture or sell any product described Reproduction disclosure or use other than for the intended purpose without specific written authorization of CareFusion Corporation or one of its subsidiaries is strictly forbidden t CareFusion Switze...