GasAlertClip Extreme English Instruction Sheet 

D2139/5 Page 1  

Trimmed and folded dimensions: 4.125 (w) x 5.875 (h)  

 

BW TECHNOLOGIES

 

GasAlertClip Extreme 

2 or 3 Year Gas Detector

 

Instruction Sheet 

Introduction 

The GasAlertClip Extreme gas detector (the detector) is a personal safety 
device that warns when hazardous gas exceeds factory set alarm setpoints. 
The detector stores and transmits gas alarm event data. It is your 
responsibility to respond properly to the alarms. 

Gas Detected 

Unit of Measure 

Oxygen (O

2

Percent by volume (%) 

Carbon monoxide (CO) 

Parts per million (ppm) 

Hydrogen sulfide (H

2

S) 

Parts per million (ppm) 

Sulfur dioxide (SO

2

Parts per million (ppm) 

Safety Information - Read First 

Warning: Substitution of components may impair Intrinsic Safety. 

Warning: To prevent ignition of flammable or combustible 
atmospheres, disconnect power before servicing. 

⇒ 

Do not activate the detector after the date on the package. 

⇒ 

This product is a gas detector, not a measurement device. 

⇒ 

Perform a self-test each day prior to use. 

⇒ 

Ensure the sensor grill is free of dirt and debris. 

⇒ 

Ensure the sensor grill is not obstructed. 

⇒ 

Periodically test the response of the sensor to gas by exposing the 
detector to a target gas concentration that exceeds the low alarm 
setpoint. Manually verify that the audible and visual alarms are 
activated. 

⇒ 

Periodically calibrate the GasAlertClip Extreme O

2

 
D2139/5 (English) 
iERP: 119566 
© 2005 BW Technologies LP. All rights reserved. Printed in Canada 

ec

 Note: 

This instrument contains a lithium battery. Do not mix with the solid 
waste stream. Spent batteries should be disposed of by a qualified 
recycler or hazardous materials handler. 

Parts of the GasAlertClip Extreme 

  

 

 

Audible alarm 

Visual alarm 

Infrared download port 

Liquid crystal display (LCD) 

Alligator clip 

Start/test button 

Sensor and sensor grill 

Display Elements 

 

 

Maximum exposure in alarm 

Data transmission 

High and low alarm setpoints 

Self-test status 

Gas type 

3 / 6 

Alarm condition 

7 / 8 

Detector life-remaining indicators 

1 / 8 

Months/hours/days since last maximum exposure 

 

Pushbutton 

Pushbutton 

 

Description

C

• 

To activate the detector, press and hold 

C

 for 5 

seconds. 

• 

Press 

C

 within 24 hours of receiving a gas alarm 

to view the maximum gas exposure. 

• 

When 

Test

 displays, press and hold 

C

 for 

approximately 1 second to activate the self-test. 

• 

To calibrate the O

2

 detector,

 

press and hold 

C

 

for approximately 3 seconds. 

• 

To display the gas alarm setpoints, press 

C

• 

To transmit the data, press 

C

 when 

Prn

 and 

 

display. 