Instruction Manual 

VIB-MESH NEBULIZER 

■ Intended Use: 

This device utilizes state-of-the-art 

electrospray technology that sprays liquid 

medication in aerosol form and delivers it 

directly to the patient for breathing. 

 

■ Intended User:

 

Adult and pediatric patients suffer from 

asthma, Chronic Obstructive Pulmonary 

Disease (COPD) such as emphysema and 

chronic bronchitis, or other respiratory 

diseases that are characterized by 

obstruction to air flow. 

 

CAUTION: 

Federal law restricts this device 

to sale by or on the order of a physician. 

 

  Thank you for purchasing this product. To ensure safe and correct use of 

this product, please read the instruction manual carefully before using. 

  Please keep the instruction manual at a proper place for future reference. 

  This is a single patient device. Do not allow multiple patients to use the 

same device. 

 

Model HL100

 

 

Nebulisers.net.au

 

Summary of Contents for Vib-Mesh HL100

Page 1: ...emphysema and chronic bronchitis or other respiratory diseases that are characterized by obstruction to air flow CAUTION Federal law restricts this device to sale by or on the order of a physician Thank you for purchasing this product To ensure safe and correct use of this product please read the instruction manual carefully before using Please keep the instruction manual at a proper place for fut...

Page 2: ...epletion or mechanical failure We recommend that you have spare batteries and a backup device available to you Before Using Safety Precautions 3 Product Features 4 Components 4 Component Names and Functions 5 Usage How to assemble the nebulizer 6 How to connect to the power supply 7 How to fill the medication 9 How to operate the nebulizer 11 Cleaning How to clean after using 14 How to replace the...

Page 3: ...ands Caution If device does not shut off automatically when medication is depleted and gives off a high frequency sound press the START STOP button to turn the power off immediately to avoid the mesh breaking Please go to page 18 for troubleshooting Please clean the nebulizer parts carefully after each use Otherwise it may not function Water is not applicable for use If you fill water in the medic...

Page 4: ... 10 seconds Time varies depending on specific medication types The package contains the following components If you find any components missing please immediately contact the retailer from which you purchased the product 1 Main unit 2 Main unit cover 3 Mask adapter 4 Medication chamber 5 Alkaline batteries AA 1 5Vx2 Optional 6 Mouthpiece 7 Carrying pouch 8 AC adapter Optional 9 Inhalation mask S w...

Page 5: ...5 Component Names and Functions ...

Page 6: ...erwise it may result in a bad connection and the nebulizer may not function properly Please keep the electrodes of main unit and medication chamber clean otherwise the nebulizer may not function properly 2 Attach the mask adapter Please securely attach the mask adapter to the main unit 3 Attach the inhalation mask or mouthpiece Attach the connecting tube to the inhalation mask Inhalation mask for ...

Page 7: ...attery cover Battery life and replacement Brand new alkaline batteries can last about 4 days if used daily for 20 minutes When the low power indicator starts blinking orange color it means the batteries need to be replaced Please immediately replace with new alkaline batteries Under normal circumstances the nebulizer can be used for roughly another 10 minutes with alkaline batteries If the low pow...

Page 8: ...Plug the AC adapter s DC connector into the main unit s power supply inlet 2 Plug the AC adapter into an electric outlet Please purchase the AC adapter specifically designed for this product Do not purchase and use AC adapters of other brands Please unplug the AC adapter after using Do not leave it connected to the power supply for a long time General Recommendations Nebulisers net au ...

Page 9: ...the PUSH button is pressed down before pushing the medication chamber forward To avoid rupturing the mesh please do not poke it with your finger or other objects 2 Fill the medication Fill the medication as shown in the figure Recommended fill volume Approx 8 ml maximum 0 5 ml minimum Please close the cover of the medication chamber To prevent the medication leaking from the chamber ensure the cov...

Page 10: ...ation mask or mouthpiece Attach the connecting tube to the inhalation mask Inhalation mask for children is S size Inhalation mask for adults is L size Attach the mouthpiece Make sure the medication chamber is attached correctly or the bad connection may result in malfunction To ensure the nebulizer will function properly keep the electrodes of the main unit and the medication chamber clean Attenti...

Page 11: ...turned on the nebulizer s power indicator green will blink for 1 second and shut off automatically Water is not applicable for use If you fill the water in the medication chamber the nebulizer s can be not turned on It is normal for the nebulizer spraying 1 second then having 0 5 second pause after turning power on However it will spray continuously after that 0 5 second pause Press and hold on th...

Page 12: ...medication stays in contact with the mesh otherwise the nebulizer will shut off automatically after approximately 10 seconds When the medication is about to be depleted it is recommended that you tilt the nebulizer the buttons side slightly toward you This allows the remaining medication to contact the mesh to nebulization Do not shake the nebulizer strongly in the usage Otherwise the nebulizer ma...

Page 13: ...ment press the START STOP button to turn the power off The power indicator light will go out If the AC adapter is being used please unplug from the wall outlet after turning off the power The unit will give off a high frequency sound and turn off the power automatically when the medication is depleted Press Nebulisers net au ...

Page 14: ...ted If the device give off the high frequency sound and the distilled water is depleted please release the START STOP button to turn the power off Otherwise the mesh of the medication chamber may be broken Please clean the remaining medication after each use Otherwise the mesh of the medication chamber may become blocked 2 Dismantle the nebulizer Remove the medication chamber mask adapter connecti...

Page 15: ...the medication chamber 5 Wipe off the main unit with new gauze Dab a piece of gauze with water and lightly wipe off the stains from the main unit Then use new gauze to dry Please clean the electrodes on the main unit and medication chamber This ensures a normal electrical conduction and hence a normal nebulization Please do not clean the nebulizer with a volatile liquid such as benzene or thinner ...

Page 16: ...roduct or any nearby retailers 1 Remove the medication chamber from the nebulizer Press the PUSH button on the rear side of the main unit and push the medication chamber toward the front side of the main unit 2 Re attach the medication chamber to the nebulizer Attach the medication chamber correctly as shown in the figure To avoid damaging the nebulizer please ensure the PUSH button is pressed dow...

Page 17: ... mask adapter mouthpiece or connecting tube and inhalation mask as shown in the figure 2 Put on the main unit cover Please put on the main unit cover as shown in the figure This will protect the nebulizer from possible damage during carrying 3 AC adapter For easy carrying please bind the AC adapter and its electric wire together with a ribbon band as shown in the figure Nebulisers net au ...

Page 18: ...rying pouch for carrying Please do not carry a nebulizer that still contains medication or water The medication may leak out and damage or stain the nebulizer Do not store the nebulizer in an area with high temperature or humidity or in direct sunlight Nebulisers net au ...

Page 19: ... of clogged medication or water and restart the power See page 14 Electrodes on nebulizer and medication chamber are stained Remove the stains and restart the power See page 14 After turning power on power indicator lights for one second and then immediately goes out Medication chamber is not completely attached Re attach the medication chamber correctly and restart the power See page 6 No medicat...

Page 20: ...14 Electrodes on nebulizer and medication chamber are stained Remove the stains and restart the power See page 14 Mesh of medication chamber is severely clogged If it still cannot be used after cleaning please replace with a new medication chamber See page 13 15 Nebulizer shuts off in usage Medication chamber is loosened and not completely attached Re attach the medication chamber correctly and re...

Page 21: ...of clogged medication or water and restart the power See page 14 Electrodes on nebulizer and medication chamber are stained Remove the stains and restart the power See page 14 Medication chamber is broken Purchase and replace with a new medication chamber See page 15 Overflow of medication from medication chamber Rupture of medication chamber or ageing of silicone ring Replace with a new medicatio...

Page 22: ...ays if used daily for 20 minutes Use 2 AA alkaline batteries Warranty 2 years Excluded medication chamber and tubes Operating Conditions 10 40 C 50 104 F 30 85 R H Storage Conditions 20 70 C 4 158 F 85 R H Accessories Main unit cover mouthpiece mask adapter alkaline batteries carrying pouch instruction manual inhalation mask S with connecting tube optional inhalation mask L with connecting tube op...

Page 23: ... ml min 2ml 1 NaF Sound Noise level at 1 m distance 50 dB Note Performance may vary with different drugs such as suspensions or high viscosity See drugs supplier s data sheet for further details MMAD Mass Median Aerodynamic Diameter Particle size distribution compliant with EN 13544 1 Test result of cascade impactor measurements for particle size Nebulisers net au ...

Page 24: ...m which you purchased the nebulizer HL100 MS Inhalation mask S with connecting tube For single use For Children HL100 ML Inhalation mask L with connecting tube For single use For Adult HL100 MB Medication chamber HL100 MP Mouthpiece HL100 MC Main unit cover HL100 MA Mask adapter HL100 CP Carrying pouch HL100 AC AC adapter Nebulisers net au ...

Page 25: ... abnormal operation caused by electromagnetic interference between electrical and electronic equipments do not use the device near a cell phone or microwave oven Discard the used product to the recycling collection point according to local regulations Distributor Chipsharbour Electronics Pty Ltd 4 17 Evans Avenue Eastlakes NSW 2018 Australia Web Nebulisers net au Phone 02 8011 4618 IMPORTANT NOTE ...

Page 26: ...experienced radio TV technician for help Changes or modifications not expressly approved by the party responsible for compliance could void the user s authority to operate the equipment Guidance and Manufacturer s declaration electromagnetic emissions The device is intended for use in the electromagnetic environment specified below The customer or the user of the device should assure that it is us...

Page 27: ... Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 line s kV line s to earth line s earth Mains power quality should be that of a typical commercial or hospital environment interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycl...

Page 28: ...field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment NOTE UT is the AC mains voltage prior to application of the test level ...

Page 29: ...ms 3 V m Portable and mobile RF communications equipment should be used no closer to any part of the device including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 d 1 2 80 MHz to 800 MHz d 2 3 800 MHz to 2 5 GHz where P is the maximum output power rating of the transmitter in watts ...

Page 30: ...roadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above the device should be observed to verify normal operation If abnormal performanc...

Page 31: ...maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 80 MHz to 800 MHz d 1 2 800 MHz to 2 5 GHz d 2 3 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equat...