Ultrasonic Endo Activation Device 

 

 

User Manual 

 

 

 

 

 

 

 

 

 

 

 

 

Changzhou Bomedent Medical Technology Co., Ltd. 

 

 

 

Summary of Contents for Actor I pro

Page 1: ...Ultrasonic Endo Activation Device User Manual Changzhou Bomedent Medical Technology Co Ltd ...

Page 2: ...ss II Classification of electric shock protection level Class B Classification of liquid entry prevention level IPX0 Sterilization or disinfection method please refer to the cleaning and disinfection section Classified by the safety level in the case of using the product with flammable anesthetic gas mixed with air or flammable anesthetic gas mixed with oxygen or nitrous oxide not to be used in th...

Page 3: ...duct Installation 6 4 1 Installation of Silicone Case 6 4 2 Installation of Working Tip 6 4 3 Removal of Working Tip 6 5 Product Use 7 5 1 Operation 7 5 2 Charging 9 5 3 Wireless Foot Control 10 6 Troubleshooting 12 7 Cleaning Disinfection and Sterilization 13 7 1 Cleaning and Sterilization of Working Tip Wrench and Silicone Case 13 8 Storage Maintenance and Transportation 15 9 Technical Parameter...

Page 4: ...f the working tip can be adjusted so as not to harm the root canal and its frequency can be dynamically adjusted to ensure cleaning efficiency c The working tip silicone case and wrench can be sterilized by high temperature and high pressure d This device can be illuminated to make it clearer 1 2 Model Actor Ⅰ pro 1 3 Scope of Application Intended use for cleaning of the root canal area of teeth U...

Page 5: ... manufacturer Please beware of collision especially falling When the working tip and silicone case need to be removed please turn off the power first So as not to accidentally touch the switch on the handle thus leading to its accidental starting and causing injury to human body When the battery on the screen displays flashing please timely charge it When charging please use the original power ada...

Page 6: ... Product Configuration 2 1 External Structure of Main Frame of Product 2 2 Main Accessories of Product Working tip Wrench Silicone case Power adapter Power cord Charging base Wireless foot switch optional ...

Page 7: ...harging indicator light on charging base Switch button Power on when the device is off press and hold the switch button for 1 second and the device will enter the standby state with the ring tone Power off in standby state press and hold the switch button for 1 second and the device will be turned off with the ring tone Operation and stop in standby state press the switch button for 0 1 seconds an...

Page 8: ...ally Bluetooth button Turn on bluetooth when both the wireless foot switch and the device are in standby state press any button of the wireless foot switch first and then press the bluetooth button to open the bluetooth Turn off bluetooth when the bluetooth is on press the bluetooth button to turn off the bluetooth Bluetooth indicator light The indicator light is always on in blue it means that th...

Page 9: ...10 is normal Warning When installing the working tip please ensure that the main frame is powered off When the working tip is installed on the main frame it will not become loose or fall off to ensure that the working tip is firmly installed Before inserting the working tip check the assembly interface between the working tip and the main frame Please do not use the damaged working tip or device I...

Page 10: ...he standby state Run the device in standby state press the switch button to run the device the power indicator light is on the illuminating light opens and the working tip begins to vibrate Warning Once the device is started please do not touch the working tip to avoid scalding Please do not operate the device without flushing fluid in the root canal Make sure there is enough flushing fluid in the...

Page 11: ... mouth to make sure the device is functional The working tip may be damaged suddenly when it enters the root canal which is too curved or not in good shape If the user feels uncomfortable please stop using the device immediately Even in normal use the working tip may be separated from the main frame due to metal fatigue and wear Please replace it in time Please do not use the working tip more than...

Page 12: ...the charging is completed When the charging base is powered on the charging indicator light on the charging base is always on in green and when the main frame is placed on the charging base the charging indicator light on the charging base turns to be on in blue It is expected that the time for the device to be fully charged is about 4 hours which can guarantee the operation of the device in the s...

Page 13: ...3 Wireless Foot Control We always design the product from the perspective of the doctor s use to make its operation easier Actor I pro is a safe accurate compact and convenient product When a doctor performs root canal treatment on the molar teeth they need to go deep into the mouth so it is inconvenient to press the buttons on the device In order to provide doctors with more convenient operation ...

Page 14: ...or one to one connection through bluetooth that is one module can only be connected with one product When multiple devices are searching and pairing at the same time the target device may not be connected so please pair it with one to one separately When turning on the bluetooth function please stay away from other bluetooth WiFi and other 2 4G wireless products After successful bluetooth connecti...

Page 15: ...ame is not inserted into the charging base properly Please check whether there is any foreign body in the main frame 5 2 The working tip fails to vibrate 1 The working tip is not tightened 2 The working tip is damaged 3 The main frame is damaged 1 Tighten the working tip with a wrench 2 Replace the working tip with a new one 3 Please contact the distributor 4 2 The indicator light on the main fram...

Page 16: ... are used for each patient High temperature and high pressure steam sterilization method is recommended The recommended high temperature and high pressure steam sterilization parameters are as follows a Perform steam sterilization at 134 for 5 minutes placed in a sterilization bag The sterilization temperature should not exceed 135 The steam sterilizer shall be used in accordance with Standard EN ...

Page 17: ...th cleaner 4 Soak Soak the working tip wrench and silicone case in the cleaner for 5 minutes 5 Rinse Rinse the working tip wrench and silicone case with purified water for 2 minutes to remove residual cleaner on its surface 6 Dry Use a dry absorbent and soft cloth to wipe the residual water on the surface of the working tip wrench and silicone case 2 Sterilization Steps Parameters Sterilize Put th...

Page 18: ...y specialized maintenance personnel or special repair shops b Please keep this product dry Rainwater moisture and liquids may contain minerals that can corrode the electronic circuit of this product c Do not throw knock or vibrate this product Rough handling of this product may damage its internal circuit board and wires d Do not apply paint to this product as it will leave debris in the removable...

Page 19: ... 0 15A max Output DC5V 1A Liquid permeation protection IPX0 Classification of electric shock protection type Class II Internal power unit Classification of electric shock protection level Class B Main vibration offset of working tip 100μm Vibration frequency of working tip 45kHz 5kHz Effective connection range of bluetooth 5m Power consumption 3W Use environment Temperature 10 40 Humidity 10 70 At...

Page 20: ...anufacture For indoor use Upward Damp proof Fragile Please handle it with care 11 Environmental Protection This product does not contain harmful ingredients and can be disposed and destroyed according to the relevant local regulations 12 After sales Service Since the date of sale if this device cannot work normally due to quality problems it can be repaired by our company or the designated agent b...

Page 21: ... that it can work properly in the configuration in which it is used Guidance and manufacturer s declaration electromagnetic emissions The device is intended for use in the electromagnetic environment specified below The customer or the user of the device should assure that is used in such an environment Emission test Conformity Emission test Conformity Electromagnetic Environment guidance RF Emiss...

Page 22: ...ifferential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycles 40 UT 60 dip in UT for 5 cycles 5 UT 70 UT 30 dip in UT for 25 cycles 5 UT 5 UT 95 dip in UT for 5 s 5 UT 95 dip in UT for 0 5 cycles 40 UT 60 dip in U...

Page 23: ...itters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from s...

Page 24: ...smitter in meters Meters m 150 kHz 80 MHz 1 2 d P 80 MHz 800 MHz 1 2 d P 800 MHz 2 5 GHz 2 3 d P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum ...

Page 25: ......

Page 26: ......

Page 27: ......

Page 28: ...1980 ChangZhou BoMedent Medical Technology Co Ltd NO 9 Changyang road West Taihu Science Technology Industrial Park Changzhou City jiangsu China Caretechion GmbH Niederrheinstr 71 40474 Duesseldorf Germany Document Number RD RCW 030 Version number A0 Date of compilation November 30 2020 ...