manualshive.com logo in svg
background image

Agilent Technologies 

N9010/20/30AK-REA

Installation Guide

Refer to Agilent X Series Signal Analyzers

Option SSD Solid State Disk Drive 

Installation Note (N9030-90009)

Manufacturing Part Number: N9030-90019

Printed in USA: July 2011

Supersede: March 2010

© Copyright Agilent Technologies, Inc. 2010, 2011

Содержание N9010AK-REA

Страница 1: ...ation Guide Refer to Agilent X Series Signal Analyzers Option SSD Solid State Disk Drive Installation Note N9030 90009 Manufacturing Part Number N9030 90019 Printed in USA July 2011 Supersede March 2010 Copyright Agilent Technologies Inc 2010 2011 ...

Страница 2: ...THIS DOCUMENT THAT CONFLICT WITH THESE TERMS THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL DFARS Restricted Rights Notice If software is for use in the performance of a U S Government prime contract or subcontract Software is delivered and licensed as Commercial computer software as defined in DFAR 252 227 7014 June 1995 or as a commercial item as defined in FAR 2 101 a or as Restricte...

Страница 3: ... are fully understood and met Definitions Specifications describe the performance of parameters covered by the product warranty temperature 0 to 55 C unless otherwise noted Typical describes additional product performance information that is not covered by the product warranty It is performance beyond specification that 80 of the units exhibit with a 95 confidence level over the temperature range ...

Страница 4: ...iv Installation Guide ...

Страница 5: ...eturn for Service 5 Safety and Regulatory Information 6 Introduction 6 Safety Earth Ground 6 Declaration of Conformity 6 Statement of Compliance 6 Before Applying Power 7 Servicing 8 Connector Care and Cleaning Precautions 8 Electrostatic Discharge Protection 9 Regulatory Information 10 Instrument Markings 10 Battery Collection 11 Compliance with German Noise Requirements 11 EMC Information 11 Agi...

Страница 6: ...Contents 2 Contents ...

Страница 7: ...N9030 90019 1 N9010 20 30AK REA ...

Страница 8: ...U in the N9010A and N9020A instrument must be equipped with a removable disk drive Early N9010A or N9020A instruments without removable disk drives will require installation of Option AK PC2 CPU with removable drive All N9030A instruments have removable disk drives For additional information please go to http www agilent com find security IMPORTANT For further information refer to Agilent X Series...

Страница 9: ...vices and Assistance on page 12 Keep the damaged shipping materials if any for inspection by the carrier and an Agilent Technologies representative Customer Furnished Equipment CFE eSATA Cable 2 5 inch SATA HDD enclosure Tools Required Torx Driver T 10 USB Mouse USB Keyboard Utilities Adjustments and Performance Verification Tests Utilities Required None Adjustments Required None Performance Testi...

Страница 10: ...res ESD can damage or destroy electronic components All work on electronic assemblies should be performed at a static safe workstation Refer to the documentation that pertains to your instrument for information about static safe workstations and ordering static safe accessories Removing the HDD SSD 1 Turn the instrument off and remove the AC power cord 2 Using a T 10 torx driver loosen the rear pa...

Страница 11: ...A enclosure in the desired location Follow the instructions that are included with the enclosure 6 Install the removable tray assembly eSATA W1312 60070 in the instrument Torque to 9 in lb 7 Connect the eSATA cable CFE from the removable tray assembly eSATA to the enclosure 8 Provide power per the enclosure instructions 9 Install the instruments AC power cord and turn on the instrument Return for ...

Страница 12: ...und incorporated in the power cord The mains plug shall be only be inserted in a socket outlet provided with a protective earth contact Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous Intentional interruption is prohibited CAUTION Always use the three prong AC power cord supplied with this product Failure to ensure adequate eart...

Страница 13: ...orrect components may cause product damage and serious injury CAUTION This product is designed for use in Installation Category II and Pollution Degree 2 per IEC 61010 Second Edition CAUTION Before switching on this instrument make sure the supply voltage is in the specified range CAUTION Verify that the premise electrical voltage supply is within the range specified on the instrument CAUTION Vent...

Страница 14: ...ning WARNING No operator serviceable parts inside Refer servicing to qualified personnel To prevent electrical shock do not remove covers WARNING The detachable power cord is the instrument disconnecting device It disconnects the mains circuits from the mains supply before other parts of the instrument The front panel switch is only a standby switch and is not a LINE switch disconnecting device Co...

Страница 15: ... components and assemblies or when making connections always wear a heel strap 9300 1126 when working in an area with a conductive floor If you are uncertain about the conductivity of your floor wear a heel strap always ground yourself before you clean inspect or make a connection to a static sensitive device or test port You can for example grasp the grounded outer shell of the test port or cable...

Страница 16: ...osition This symbol is used to identify a terminal which is internally connected to the product frame or chassis The CE mark is a registered trademark of the European Community If accompanied by a year it is when the design was proven The CSA mark is a registered trademark of the CSA International This instrument complies with Canada CSA 22 2 No 61010 1 04 This is a symbol of an Industrial Scienti...

Страница 17: ...eclare that this instrument is in conformance with the German Regulation on Noise Declaration for Machines Laermangabe nach der Maschinenlaermrerordnung 3 GSGV Deutschland EMC Information Complies with European EMC Directive 2004 108 EC IEC EN 61326 1 CISPR Pub 11 Group 1 class A AS NZS CISPR 11 This ISM device complies with Canadian ICES 001 Cet appareil ISM est conforme a la norme NMB du Canada ...

Страница 18: ... number and full serial number With this information the Agilent representative can determine the warranty status of your unit Shipping Your Product to Agilent for Service or Repair IMPORTANT Agilent Technologies reserves the right to reformat or replace the internal hard disk drive in your analyzer as part of its repair This will erase all user information stored on the hard disk It is imperative...

Отзывы: