background image

ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37 

 

 

 

CC-04 

 

Calibrateur Volt/mA   

Notice d’utilisation 

 

 

 

 

 

         

Summary of Contents for CC-04

Page 1: ...ADEL Instrumentation 28 Rue de Stalingrad 38300 BOURGOIN JALLIEU 1 04 74 93 06 37 CC 04 Calibrateur Volt mA Notice d utilisation ...

Page 2: ...1 Pushbutton Functions 3 2 2 Display 5 3 Using 7 3 1 Input 7 3 1 1 Measure Voltage dc MEASURE V 7 3 1 2 Measure Millivolt dc MEASURE mV 8 3 1 3 Measure Current dc MEASURE mA 9 3 1 4 Measure mA with Loop Power 10 3 2 Output 11 3 2 1 SOURCE mA Mode Sourcing mA 11 3 2 2 Simulate a Transmitter SIMULATE mA 13 3 2 3 Output Voltage dc SOURCE V 14 3 2 4 Output Millivolt dc SOURCE mV 15 3 2 5 Auto Ramp Fun...

Page 3: ... 3 4 Charging Function 19 3 5 BL Backlignt 19 4 Technical Index 20 4 1 MEASURE V dc 20 4 2 MEASURE mV dc 20 4 3 MEASURE mA dc 20 4 4 SOURCE SIMULATE mA dc 21 4 5 SOURCE V dc 21 4 6 SOURCE mV dc 21 4 7 LOOP Power 21 4 8 General Specifications 22 5 Maintenance and Calibrati 23 5 1 Repair 23 5 2 Cleaning 23 5 3 Replace the Battery and Fuse 23 6 Accessories Included 25 Ⅱ ...

Page 4: ...ound explosive gas vapor or dust Never apply more than 30 V between any two terminals or between any terminal and earth ground Please select the appropriate port and scale span before using it Never use this Calibrator when the case cover open Once the symbol low battery appears replace the battery to avoid false readings that can lead to electric shock Remove test lead from the Calibrator before ...

Page 5: ...d Low battery Conforms to European Union requirements 二 Description Volt mA Calibrator hereafter to be referred as Calibrator is an accurate tool for current or voltage supplying and measuring The Calibrator can souring output 0 24mA or simulates a two wire transmitter output 0 24mA it can measures 0 20mA or 4 20mA current loop as well as measure voltage dc 0 100mV 0 10V It is widely used to calib...

Page 6: ...se Backlight ON OFF pushbutton Press to step through modes SOURCE mA mode display SOURCE mA SIMULATE mA mode display SIMULATE mA Output voltage dc display SOURCE V Output millivolt dc display SOURCE mV MEASURE voltage dc display MEASURE V MEASURE milivolt dc display MEASURE mV MEASURE current dc display MEASURE mA Measure current dc with Loop Power display MEASURE mA LOOP It would storage the curr...

Page 7: ...RCE mV mode the voltage output could be adjusted to the resolution of 0 01mV 0 1mV 1mV 0mV default value is 0 01mV By rotating the knob the current or voltage signal output could be increased or decreased in current setting resolution If turn the knob into auto ramp function or 25 stepping mode the direction and value could be changed for output signal Press to increase the current or voltage with...

Page 8: ...could cycle through select a ramp output type stepped ramp display slow ramp display fast ramp display Press to select SpanCheck function SpanCheck is displayed From 0 and press it again to 100 For example The 0 20mA SpanChecks starting from 0 is 4mA and press it again to 100 is 20mA Press and simultaneously to select the SpanCheck Span for example 4 mA 20 mA 0 100 default value 0 mA 20 mA 0 100 o...

Page 9: ...brator Description 6 Symbol Explanation MEASURE Measure current or voltage SIMULATE Simulate mA current SOURCE Sourcing output current SpanCheck Check of the full scale 0 100 Loop Loop power mAV The combination of unit symbol mA mV V Auto ramp type Stepping direction indicate symbol Low battery ...

Page 10: ...s to step to MEASURE mode The display shows MEASURE V 2 Touch test lead probes to the circuit across the load or power source as shown in right figure 3 If the voltage over 10V the LCD will show OL The lower display area shows the voltage value percentage 7 Caution To prevent damage to the unit under test ensure that the Calibrator is in the correct mode before connecting the test leads ...

Page 11: ... to MEASURE mode The display shows MEASURE mV 2 Touch test lead probes to the circuit across the load or power source as shown in right figure 3 If the voltage over 100mV the LCD will shows OL The lower display area shows the current value percentage 8 Caution To prevent damage to the unit under test ensure that the Calibrator is in the correct mode before connecting the test leads ...

Page 12: ...EASURE mode The display shows MEASURE mA 2 Touch test lead probes to the circuit across the load or power source as shown in right figure 3 If the current over 24mA the LCD will shows OL The lower display area shows the current value percentage 9 Caution To prevent damage to the unit under test ensure that the Calibrator is in the correct mode before connecting the test leads ...

Page 13: ...hown below To Measure mA dc current with Loop Power 1 Press to step to MEASURE mA LOOP mode The display shows MEASURE with unit mA measure mA and Loop symbol 2 Touch test lead probes to the circuit across the load or power source as shown below 3 If the current over 24mA the display will show OL The lower display area shows the current value percentage 10 Caution To prevent damage to the unit unde...

Page 14: ...igure flashing when you set an output value 1 Press to step to SOURCE mA mode The display shows SOURCE with unit mA 2 The test leads connected as shown in the right figure the red test wire connected to LOOP and the black test wire connected to mV V socket 3 If the current over 24mA the display will show OL 4 Press in and rotary the knob to step up down in different increments 5 Press key stepping...

Page 15: ...ADEL Instrumentation 28 Rue de Stalingrad 38300 BOURGOIN JALLIEU 1 04 74 93 06 37 Volt mA Calibrator Using 12 ...

Page 16: ...leads as shown in figure 1 Press to step to SIMULATE mA mode The display shows SIMULATE with unit mA 2 The general connection is that red is positive and black is negative the red test wire connected to mV V and the black test wire connected to COM socket 3 If the current over 24mA the display will show OL 4 Press in and rotary the knob to step up down in different increments 5 Press key stepping ...

Page 17: ... in the right figure the general connection is that red is positive and black is negative the red test wire connected to mV V and the black test wire connected to COM socket 3 If the current over 10V the display will show OL 4 Press in and rotary the knob to step up down in different increments 5 Press key stepping up down 25 of full scale for output value 6 Press key to select auto ramp output ty...

Page 18: ...wn in the right figure the general connection is that red is positive and black is negative the red test wire connected to mV V and the black test wire connected to COM socket 3 If the current over 100mV the display will show OL 4 Press in and rotary the knob to step up down in different increments 5 Press key stepping up down 25 of full scale for output value 6 Press key to select auto ramp outpu...

Page 19: ...nsmitter s response Similary In SOURCE V SOURCE mV mode auto ramp function apply a varying voltage singal continuously from the Calibrator to load circuit Press to enter into Auto Ramp mode and select a ramp type The Calibrator applies or controls a continuously repeating current or voltage signal in one of three ramp types in the span range Step 25 stepping ramp with 5s of pause per step Slow ram...

Page 20: ... 0 point of full scale output press key again it can execute 100 point of full scale output cycle repeatedly SpanCheck function can be used to check the zero points or span spacing of a transmitter under SOURCE or SIMULATE mode Press to select SpanCheck Press any pushbutton except pushbutton or rotate the knob to exit this function The Calibrator has two mA output spans 4 mA 20 mA 0 100 default va...

Page 21: ...00 Ω MEASURE SIMULATE mode 12 hours The Calibrator turns off automatically after 15 minutes of inactivity on default To decrease this time or disable this feature 1 When the Calibrator off press simultaneously The display shows PS XX in which XX indicates the tun off time with the unit of minute OFF indicates that the power saver is disabled 2 Rotate the knob to decrease or increase the tun off ti...

Page 22: ... the indicator of the power adapter turn red indicate that the Calibrator is being charged 3 When the indicator where on the top of the Calibrator from red to green indicate that the charging has been completed and the charging time is about two to three hours Must be charged fully when charging otherwise it may affect the life of the Li Polymer battery pack If the running time of the Calibrator i...

Page 23: ...alibration with operation temperature 18 28 which shown as of reading counts 4 1 MEASURE V dc Range Resolution Accuracy Input Impedance 0 10V 1mV 0 02 of reading 4 Counts 1MΩ 4 2 MEASURE mV dc Range Resolution Accuracy Input Impedance 0 100mV 0 01mV 0 02 of reading 4 Counts 1MΩ 4 3 MEASURE mA dc Range Resolution Accuracy Sampling resistor 0 24 mA 1 uA 0 02 of reading 4 Counts 10Ω 20 ...

Page 24: ...0 24 mA 1 uA 0 02 of reading 4 Counts SOURCE mode with load ability Compliance To 1000 Ω at 20 mA External loop voltage specification under SIMULATE mode Typically 24V max 30V min 12V 4 5 SOURCE V dc Range Resolution Accuracy 0 10V 1mV 0 02 of reading 4 Counts 4 6 SOURCE mV dc Range Resolution Accuracy 0 100 mV 0 01 mV 0 02 of reading 4 Counts 4 7 Loop Power 24V 26V 21 ...

Page 25: ... to 60 Operation temperature 10 to 55 Operation elevation Max 2000m in height Temperature coefficient 0 005 range 10 to 18 or 28 to 55 Relative humidity 95 up to 30 75 up to 40 45 up to 50 35 up to 55 Vibration Random sampling 2 grams 5 500Hz Impact 1m free drop test Size 195 L 92 W 55 H mm Weight About 400g including the rubber cover Conformity to safety standards Meet the standard of IEC 61010 1...

Page 26: ...Cleaning Periodically wipe the case with a damp cloth and detergent Do not use abrasives or solvents 5 3 Replace the Battery or Fuse If the symbol low battery is displayed need to replace the battery immediately If replace the battery use the same specification alkaline 9V battery or 7 4V Li Polymer battery pack Steps are as follows 1 Press to turn the Calibrator off 2 Remove the test leads fron t...

Page 27: ...ion If replace the fuse the steps are as follows 1 Press to turn the Calibrator off 2 Remove the test leads fron the terminals 3 Lift off the battery cover on the back of the Calibrator as shown in fifure 4 Remove the damaged fuse 5 Insert the replacement fulse and restore the battery cover make sureit is securely in place 24 ...

Page 28: ... mA Calibrator Accessories Included 6 Accessories Included Item Description Quantity 1 A 7 4V Li Polymer battery 1 2 Test lead red Test lead black 1 1 3 Bending test plug wire clip red Bending test plug wire clip black 1 1 4 Adapter DC 10V 1A 1 5 Instruction sheet 1 6 Certification 1 25 ...

Reviews: