The thermometer is a non-contact infrared thermometer. Simply aim the thermometer at the target and press the ‘Scan 
key’ to display the surface temperature.   

 

Laser Information: (for models with 
Laser only) 
By pressing the ‘Scan key’, the device 
will begin to measure the temperature 
of the target and the laser will operate. 

 CAUTION! 

1.

 

WHEN DEVICE IS IN USE, DO NOT 
LOOK DIRECTLY INTO THE LASER 
BEAM–PERMANENT EYE DAMAGE 
MAY RESULT. 

2.

 

USE EXTREME CAUTION WHEN 
OPERATING THE LASER. 

3.

 

NEVER POINT THE DEVICE 
TOWARDS ANYONE’S EYES. 

4.

 

KEEP OUT OF REACH OF ALL CHILDREN. 

 
Always with the real time Max data 

1.Power on: Press the “Scan. key”. 
2.Temperature taking: When it’s power on, press the “Scan. key”, will get the   

real time Max data immediately and the reading of measurement will be continuous updated.   

 
HACCP CHECK   

The “HACCP CHECK” feature is incorporated in our thermometer 
temperature to graphically indicate critical temperature zone. The icons and 
LED indicators located above the display indicate a food product stays in a 
safe or unsafe HACCP “ Danger Zone” temperature. The green and red 
LED light will always be lit before power off.   
A Green LED appears with icon ”   ” indicates a safe cool or frozen 
condition below 4

o

C(40

o

F) or appears with icon ”   ” indicates a safe 

holding temperature above 60

o

C(140

o

F). 

When temperature is between 4

o

C and 60

o

C, the red LED with icon ” 

 ” 

will appear and indicate that the temperature is fallen within the HACCP “Danger Zone” from 4

o

C to 60

o

C (40~140

o

F)

 

 

°C OR °F MODE 

To change the thermometer from ‘°C’ to ‘°F’ or from ‘°F’ to ‘°C’, firstly turn the instrument on by pressing the ‘Scan key’, 
then press the ‘Mode key’ one time, the ‘°C’ or ‘°F’ symbol will flash, press the ‘Scan key’ to change to scale. 

EMISSIVITY RANGE 

The infrared thermometer is supplied with a default emissivity of 0.95. The emissivity of the thermometer can be 
changed from 0.05 (5E) to 1 (100E). Changes should only be carried out by experienced personnel. To change the 
emissivity, firstly turn the instrument on by pressing the ‘Scan’ button, then press the ‘Mode’ button two times for 
emissivity function. The 95E will flash on the LCD screen, then press the ‘Meas.’ button to adjust the emissivity value, 
press the ‘Mode’ button again to exit the set up screen. For information relating to the emissivity of specific materials, 
please contact the nearest retailer. Note: non-contact infrared thermometers are not recommended for use in measuring 
the temperature of shiny or polished metals. 
 

BAR DISPLAY:

 The fixed setting range for bar display is 20~220°C and Center=120°C. 

The bar display will be updated comply with the measurement reading 
changed.  
 

LCD ERROR MESSAGES 

The thermometer incorporates visual diagnostic messages as follows: 

 

 

‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the measurement 
range. 

 

 

‘Er2’ is displayed when the thermometer is exposed to rapid changes in the ambient temperature. 
‘Er3’ is displayed when the ambient temperature of the thermometer EXCEEDS 0°C (32°F) OR 
+50°C (122°F). In both cases you should allow plenty of time (minimum 30 minutes) for the 
thermometer to stabilize to the working/room temperature. 

 

For all other error messages it is necessary to reset the thermometer. To reset the thermometer, 
turn the instrument off, remove the battery and wait for a minimum of one minute, reinsert the 
battery and turn on. If the error message remains please contact the Service Department for 
further assistance. 

 

BATTERIES 

The thermometer incorporates visual low battery indication as follows:

 

 

 

 

‘Battery OK’: 

measurements 

 are possible 

‘Battery Low’: battery needs to 

be replaced, measurements 

are possible 

‘Battery Exhausted’: 

measurements are   

not possible 

When the ‘Low Battery’ icon indicates the battery is low, please use small, pointed screwdriver with the “X” 
shaped to release the screw on the battery cover, the batteries should be replaced immediately with AAA, 
1.5V * 2pcs. 

  It is important to turn the instrument off before replacing the batteries,

 

otherwise the thermometer may   

   malfunction. 

*

 

IP54  

Splash Proof 