ZORN INSTRUMENTS ZFG 3000 ECO User Manual Download Page 1

 
 
 

 

User Manual  

for the  

Light Weight Deflectometer 

 

 

ZFG 3000 GPS 

ZFG 3000 ECO 

 

in accordance with the  

German technical test requirements  

for soil and rocks in road construction 

TP BF – StB Part B 8.3 

 

 

 
 

 

 

 

Benzstraße 1, 39576 Stendal (Germany) 

Phone: +49(0) 3931 / 25273 0, Fax: +49(0) 3931 / 25273 10 

www.zorn-instruments.de

         

[email protected]

 

 

Technical changes reserved 

Effective: 11/03/2013 

Summary of Contents for ZFG 3000 ECO

Page 1: ...e with the German technical test requirements for soil and rocks in road construction TP BF StB Part B 8 3 Benzstra e 1 39576 Stendal Germany Phone 49 0 3931 25273 0 Fax 49 0 3931 25273 10 www zorn in...

Page 2: ...2 ZFG 3000 GPS ZFG 3000 ECO...

Page 3: ...ter 10 6 Care and maintenance 10 6 1 Care of mechanical loading device 10 6 2 Battery 10 6 3 Protection against total discharge 10 6 4 Automatic switch off when not in use 10 7 Device view 11 8 Instru...

Page 4: ...for example during rail track construction trench backfilling for general backfill in boreholes or other locations with difficult access low space requirements and low test equipment weight low time...

Page 5: ...between 30 and 120 seconds to display the location coordinates on its display The location coordinates are now displayed in place of the name Zorn Instruments If no satellite signals can be received e...

Page 6: ...sure 300mm 10kg Execute 1st pulse 1 A single acoustic beep acknowledges the start of measurement The measuring mode currently set is now displayed e g loading plate size and mass of drop weight It may...

Page 7: ...you can print out the report by pressing the Print button The measurement is completed with OK and a new measurement using the same sequence can be started with OK again After the measurement is comp...

Page 8: ...Print button to print out the report afterwards 4 3 Delete card Select the Delete card menu command in the menu The number of saved individual measurements is displayed in the second line Contain dat...

Page 9: ...f button 4 7 Calibration self test Select the Calibration menu command in the menu see 4 a shows a value of 0 0 If the measuring cable connection between loading plate and settlement gauge is correct...

Page 10: ...drop weight tester must be stored carefully during transport and the fall height must be checked at regularly The correct fall height is stated on the calibration report and on the calibration namepl...

Page 11: ...imit stop with release device 3 Guide tube 4 Round handle 5 10 kg fall weight 6 Transport lock 7 Spring element 8 Anti tip device 9 Centring cone 10 Carrying handle 11 Loading plate 12 Socket for conn...

Page 12: ...ng the weight too harshly which can lead to faulty measurements 9 Technical data for mechanical loading device Weight of fall weight 10 kg 15 kg Total weight of loading device 15 kg 20 kg Maximum puls...

Page 13: ...omatic switch off when not in use 4 min when battery is empty 4 5V 12 Scope of delivery The package includes ZFG3000 GPS electronic measuring device including GPS module 10 kg loading device with tria...

Page 14: ...PR ZISIONS PR FTECHNIK Kalibrierstelle Benzstr 1 D 39576 Stendal Tel 49 0 3931 25273 0 Fax 49 0 3931 25273 10 eMail info zorn instruments de Please visit the BAST homepage for more calibration institu...

Page 15: ...he planum it is sufficient to achieve a deflection modulus on the planum of at least Ev2 100 MN m or alternatively Evd 50 MN m and for Construction Classes V and VI a deflection modulus of at least Ev...

Page 16: ...path with traffic load IV service road walking path V service road Wheelchair accessible residential street VI Section 14 3 5 Table 9 Guide value for correlation of static deflection modulus Ev2 or d...

Page 17: ...45 30 25 A 1 11 A 1 12 Strengthening 5 Maintenance P 230 M 230 Ballast top layer 80 40 kg1 2 1 00 30 40 50 45 30 25 A 1 13 A 1 13 Solid track 100 45 kg2 1 00 3 40 3 40 3 40 45 30 25 A 1 14 A 1 14 P 1...

Page 18: ...ased on guide values for allocation to Dpr ZTVE StB 94 Tab 8 1 Proposal for allocation of Evd to Ev2 in acc w FGSV AA Pr ftechnik Oct 96 edition Soils group Compaction ratio Dpr Deflection modulus Ev2...

Page 19: ...deficiency free of charge or replace damaged faulty parts with new ones No warranty claims can be made especially for damage caused by improper use normal wear handling which does not comply with the...

Page 20: ...20 16 EC conformity declaration...

Page 21: ...n CD 3 Follow the instructions Read in data 4 Start the program ZFG 3000 5 Click in menu File on Read in 6 Select i e the Folder ZFG on the SD card and click on ZXZ nrz 7 All files in folder will be r...

Page 22: ......

Page 23: ......

Page 24: ...Benzstra e 1 D 39576 Stendal Telephone 49 0 3931 25273 0 Fax 49 0 3931 25273 10 www zorn instruments de info zorn instruments de...