Using the Z Vent Ventilator
4-22
www.zoll.com
9650-002360-01 Rev. A
Secondary Parameters and Alarm Thresholds/Limits
Pressing the SpO
2
parameter button highlights the Low SpO
2
Alarm Limit alarm and enables
its value to be changed.
The Low SpO
2
Alarm Limit alarm is the only secondary parameter for the SpO
2
parameter
window. The default Low SpO
2
value at start up is 94%. The Low SpO
2
Alarm Limit is
adjustable from 86% to 99% (or OFF).
To change an alarm threshold setting, follow these steps:
1. Press the SpO
2
parameter button until the Low SpO
2
Alarm Limit highlights.
2.
Turn the
Dial
to adjust the value of the parameter (%).
3.
Press the
Accept
button.
Context Menu
The Masimo context menu is the same for both the SpO
2
and HR parameter windows.The
menu provides for the following pulse oximeter controls:
•
Standby
•
Pulse Ox
•
Fast SAT
•
Sensitivity
•
APOD
•
Averaging
•
Signal Strength
•
Signal IQ
The menu allows for the Pulse Oximeter monitoring to be placed in standby. It also provides
options to turn on the fast SAT, adjust Sensitivity, and other Masimo specific parameters, see
the table below for additional details.
To change a menu setting, follow these steps:
1. Press and hold the SpO
2
parameter button until the menu displays.
2.
Release the SpO
2
parameter button, turn the
Dial
until the Context menu item you want to
change highlights, then press the
Accept
button. The current setting for the context menu
parameter highlights.
3.
Turn the
Dial
to adjust the value of the parameter.
4.
Press the
Accept
button.