Using the Z Vent Ventilator
4-14
www.zoll.com
9650-002360-01 Rev. A
BPM Parameter Window Reference
The following table provides the options and ranges for the BPM parameters:
BPM Parameter Window
Options/Range
Availability/Notes
Primary Parameter
BPM
Breaths per minute
1 to 80
Secondary Parameters
Ti (sec)
or
I:E
Ti 0.3 to 3.0
or
I:E 1:1 to 1:99
For the Basic Model, see
the Control Parameter in
the BPM context menu:
Ti 0.1 to 5.0
or
I:E 4:1 to 1:99
Inverse I:E is available as
an advanced feature of the
Z Vent
Alarm Limits
High breath rate
20 to 99, OFF
Low breath rate
2 to 99
Measured Value
Minute Volume (I)
0 to 99
BPM Context Menu
Control
Parameter
Default I:E
I:E or Ti (sec)
The control value is shown
in the parameter window,
the dependent value is
shown in the Auxiliary Box.
Rise Time -
Pediatric Default 5
Mask CPAP Default 3
1 to 10
Auxiliary Box
Cycle Off %
(% Cycle )
Default 25%
10 to 70%
Auxiliary Box
Spont Ti Limit
Default
Adult = 3.00
Infant = 2.00
Mask CPAP = 3.00
0.30 to 4.00