Using the Z Vent Ventilator
4-10
www.zoll.com
9650-002360-01 Rev. A
To adjust or change any Apnea Backup settings,
1. From the menu, use the
Dial
to highlight the parameter to change, and then press the
Accept
button (The current highlighted parameter value highlights).
2.
Use the
Dial
to set the desired value
3.
Press
Accept
to enable the change.
Mode Parameter Window Reference
The following table provides a reference to the primary parameters, secondary parameters, and
context menu parameters and other options for the Mode parameter window.
Warning!
Apnea Backup settings are appropriate for most adult patients. For small adult and
pediatric patients, adjust the Apnea Backup settings so that they are appropriate for the
patient.
Mode Parameter Window
Options/Range
Availability/Notes
Primary Parameter
Mode
AC
SIMV
Z Vent Optional Mode
CPAP
BL
Secondary Parameters
Breath Target
(V) or (P)
LC
(Leak Compensation)
ON or OFF
(Default OFF)
AC (P) and SIMV(P)
modes
OFF
BL Mode
Default ON
CPAP Mode
Alarms
None
Measured Value
None
Mode Context Menu
Context menu is available
in CPAP and BL Modes
Apnea Backup
BPM
1 to 80
PIP
10 to 80
I:E, Ti
1:99, 0.1 to 3
Control selected in context
BPM context menu