Saving Changed Parameters for Future Use
When you change parameters, you can save them for future use through the Custom Settings
Menu option.
To save changed parameters for future use, follow these steps.
1. With the ventilator turned on, make the desired parameter changes using the parameter
buttons, Dial and Accept button.
2.
Press
Menu
button.
3.
Using the
Dial
, highlight Powerup Settings, and then press the
Accept
button.
4.
Use
Dial
to highlight Custom Settings, then press the
Accept
button to highlight Save, and
then press the
Accept
button again to save the changed parameters.
To use or confirm that the changed parameters are saved as Custom settings, follow these steps:
1. Turn on the ventilator.
2.
When the Start up Menu displays, turn the
Dial
to highlight Custom Settings, and then press
the
Accept
button.
3.
The ventilator displays the custom parameter settings in the parameter windows.
Warning!
The Custom and Last Setting options save the current configuration of the ventilator.
Always ensure that the both the ventilation settings and alarm limits are appropriate
for the patient.
Using the Last Settings Enabled on the Ventilator
The ventilator preserves the last setting used on the ventilator (even if they were not saved as
Custom Setting). To use the last settings, do the following:
1. Turn ON the ventilator, wait for the Self-Check to complete, and the Start Menu to display.
2.
Turn the
Dial
to highlight the Last Settings choice.
3.
Press the
Accept
button to enable your choice.
The ventilator begins operation with the last settings used.
Warning!
Last Settings uses all of the parameters including alarms. You must ensure that both the