![ZOLL X Series Operator'S Manual Download Page 82](http://html1.mh-extra.com/html/zoll/x-series/x-series_operators-manual_4055648082.webp)
Chapter 5 Alarms
5-2
www.zoll.com
9650-001355-01 Rev. M
Visual Alarm Indicators
In addition to status messages that appear on the display, the X Series unit lights the red or
yellow LED on the front panel to indicate the priority level of the highest-priority active alarm.
The X Series LEDs indicate the priority level of the highest-priority active alarm as shown in
the following table.
Audible Alarm Indicators
The X Series unit sounds an audible alarm to indicate the priority level of highest-priority
active alarm. The X Series indicates the priority level of the highest-priority active alarm by
sounding the audible alarm tones described in the following table.
Alarm Indicator Self-Test
The X Series unit performs a self-test of the audio and visual alarm indicators upon power-up.
To ensure that the alarms and alerts are functioning properly, verify that two alarm tones are
heard and the green, yellow, and red LEDs are illuminated upon power up.
Active Alarm/Alert Priority
Visual Alarm/Alert Indicator
High Priority -- Patient Alarm
Flashing Red LED
Medium Priority -- Technical Alert
Flashing Yellow LED
Information Signal; -- Equipment Status
Message
Visual message only. No LED
indicator
Active Alarm/Alert
Priority
Audible Alarm/Alert Indicator
Alarm/Alert Volume (measured in
accordance with ISO 3744)
High Priority --
Patient Alarm
Two sets of five short beep tones,
repeated at 15-second intervals
Adjustable range:
Maximum Level > 69 dB(A)
Minimum Level > 61 dB(A)
Medium Priority --
Technical Alert
One set of three longer beep tones,
repeated at 30-second intervals
Adjustable range:
Maximum Level > 66 dB(A)
Minimum Level > 59 dB(A) Volume
is 3 to 12 dBA below high priority
alarm
Information Signal--
Equipment Status
Message
A single short beep tone, not
repeated
Volume is at least 6 dBA below the
medium priority alert
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...