Chapter 3 Monitoring Overview
3-6
www.zoll.com
9650-001355-01 Rev. M
Configuring the Waveform Display
You can display up to four waveform traces on the Waveform Display window. The first
waveform trace always uses an ECG lead as its source (such as
Pads
or Leads
l
,
ll
, or
lll
, and so
on. The default is
Pads
). If
Pads
are not connected, the unit can be configured to
automatically default to another ECG lead for the first trace. As you insert the remaining three
traces, you can specify that the traces use an ECG lead as the waveform source, or that the trace
derive its waveform from other available monitoring functions (such as
Resp
,
CO2
,
SpO2
or
IBP channels
P1
,
P2
, or
P3
).
If configured, the unit can display four ECG traces on startup, when no other monitoring
devices are attached.
The X Series unit can also cascade a trace onto the adjoining trace area to double the duration
of the trace display.
On the Waveform Display window, to insert a new trace (
Insert
) or cascade (
Cascade
) a
displayed trace, highlight and select the trace label above the trace. In the following example,
the unit is configured to cascade the ECG Lead l trace:
Note:
The X Series unit automatically inserts a new waveform when you turn on a parameter
(CO
2
) or a new sensor signal is present (SPO
2
, IBP). The X Series unit automatically
removes a waveform when you turn off a parameter or remove a sensor and the unit
displays the resulting equipment alert.
06/06/2011 12:34:56
Adult
00:17:43
I, II,
III...
12
CO
2
R
SYNC
1 cm/mV
II
HR
bpm
80
121
79
(96)
NIBP
mmHg
38
12
BR
CO2
mmHg
SpO2
%
97
T1
ºF
98.6
Pads
III
aVL
aVF
V
Insert
Source
II
aVR
I
Cascade
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...