Specifications
A-14
www.zoll.com
9650-001355-01 Rev. M
Monitor/Display
Input
: 3-lead, 5-lead, or 12-lead patient cable, paddles, Multifunction or OneStep electrodes.
Type
: Color LCD, 640 x 480 pixels, 800 MCD
Sweep Speed
: 25 mm / sec or 50 mm / sec (User Selectable)
Lead Selections
: Paddles (Pads), I, II, III, AVR, AVL, AVF, V1-6.
Frequency Response
:
Pads/Paddles:
0.67 to 20Hz Limited response
3/5/12 Lead Continuous Monitoring (user selectable):
0.67 to 20Hz Limited response
0.67 to 40Hz Monitor response
Acquired 12-lead snapshots (supervisor selectable):
0.525 to 40Hz Filtered Diagnostic response
0.525 to 150Hz Diagnostic response
Per methods a, b, c of EC11 3.2.7.2
(Automatically sets chart recorder response)
Common Mode Rejection
:
Complies with AAMI EC13-2002 section 4.2.9.10.
Tall T-Wave Rejection
:
Meets AAMI EC13-2002, section 4.1.2.1c for 0.9 mV
T-wave (0.8 mV with diagnostic response) and 1mV QRS.
Diagnostic Signals Applied to Patient Connections
:
Leads off / active noise suppression sensing circuit is < 0.1 microamps DC. The impedance
/respiration detector signal frequency is 72 ± 7 kHz at 77 microamps RMS
pseudo-sinewave into 100 ohms.
Heart Rate Range
: 30 to 300 BPM.
Heart Rate Accuracy
: +/- 3% or +/- 3BPM, whichever is greater.
Displayed Heart Rate:
Average of last 5 beat-to-beat intervals.
Heart Rate Alarms
: User-selectable.
Size
: 0.125, 0.25, 0.5, 1, 2, 4 cm/mv and auto-ranging.
Heart Rate Meter Response Time
:
Responds to a 40 BPM step increase in heart rate within 4.5 seconds per
AAMI EC-13-2002, section 4.1.2.1.f. Responds to a 40 BPM step decrease within
3.9 seconds per AAMI EC-13-2002, section 4.1.2.1.f. Response times include a
1.0-second display update interval.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...