![ZOLL X Series Operator'S Manual Download Page 238](http://html1.mh-extra.com/html/zoll/x-series/x-series_operators-manual_4055648238.webp)
Chapter 19 Real CPR Help
19-4
www.zoll.com
9650-001355-01 Rev. M
Constant Metronome Disabled
AED Mode
•
If Enable Metronome is set to Always or AED Mode Only, the metronome is active when
chest compressions are first detected during the CPR interval.
•
If set to Never, the metronome is silent.
Analysis/CPR Protocol or Manual Mode
•
If Enable Metronome is set to Always, the metronome is active only when the reported
compression rate falls below 80 cpm during the CPR period. If the chest compression sensor
is connected, the metronome is suspended after compressions have not been detected for at
least 2 seconds.
•
If set to Never or AED Mode Only, the metronome is silent.
The following table indicates when the CPR metronome is active.
FULLY RELEASE Prompt
The X Series unit can be configured to display the text prompt FULLY RELEASE, which
instructs rescuers to lift (fully release) their hands from the patient’s chest after compressions to
allow full recoil. This prompt occurs 45 seconds into the CPR interval.
The
FULLY RELEASE
text prompt is only available when using Adult CPR electrodes.
Enable
Metronome
Setting
AED Mode
Analysis/CPR
Protocol Mode
Manual Mode
Constant Metronome Enabled
AED Mode
Only
On when CPR detected
Off when CPR Idle
Off
Off
Always
On when CPR detected
Off when CPR Idle
On when CPR detected
Off when CPR Idle
On when CPR detected
Off when CPR Idle
Never
Off
Off
Off
Constant Metronome Disabled
AED Mode
Only
On at start of CPR period
Off
Off
Always
On at start of CPR period
On only at <80
compressions per minute
On only at <80
compressions per minute
Never
Off
Off
Off
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...