Analysis/CPR Protocol Defibrillation Procedure
9650-001355-01 Rev. M
X Series Operator’s Guide
17-3
1 Turn on unit
Press the green power switch located on the top of the unit. The green, yellow, and red lights on
the top of the unit flash on and off, and the unit displays the message
SELF TEST PASSED
.
If Unit is in AED Mode
1. Press the
Manual Mode
quick access key on the front panel of the unit to enter the Manual
mode of operation.
2.
Using the navigation keys, select the four digits in the Manual Mode pass code. Press
SAVE
when you are finished. Once you have entered your pass code, you will be able to enter
Manual mode.
Note:
If the unit has been not been configured to enter a pass code, the message
Exit to
Manual Mode
is displayed. Use the navigation keys to select
Yes
to enter the Manual
mode of operation. If you do not press
Yes
within 10 seconds, the unit will revert back
to AED operation.
If Unit is in Manual Mode
– No additional steps are necessary to proceed.
If no hands-free therapy electrodes have been attached to the patient and connected to the
X Series unit, the
ATTACH PADS
message and voice prompt will be issued.
Energy Select
The default energy selections for adult patients are:
Shock 1 - 120 joules
Shock 2 - 150 joules
Shock 3 - 200 joules
The default energy selections for pediatric patients are:
Shock 1 - 50 joules
Shock 2 - 70 joules
Shock 3 - 85 joules
Note:
Pediatric defibrillator energy levels should be selected based on site specific protocols.
WARNING! Use only pediatric electrodes to defibrillate patients under 8 years of age in Analysis/
CPR Protocol mode, and make sure the patient mode is set to pediatric. Use of adult
electrodes or adult mode with pediatric patients can result in the delivery of excessive
energy doses.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...