Chapter 16 Advisory Defibrillation
16-4
www.zoll.com
9650-001355-01 Rev. M
2 Press ANALYZE Button
WARNING! Keep patient motionless during ECG analysis. Do not touch the patient during analysis.
Cease all movement via stretcher or vehicle before analyzing the ECG.
Press the
ANALYZE
button to begin the analysis of the patient’s ECG rhythm and to determine
if a shockable rhythm is present.
An
ANALYZING
message is displayed at the top of the screen for 6 to 12 seconds while the
patient’s ECG is analyzed. Once the analysis is completed, the unit indicates whether or not a
shock is advised.
The analysis determines whether there is a shockable rhythm. If the analysis detects a
shockable rhythm the unit prompts the operator to shock the patient at the preconfigured energy
level. If the analysis does not detect a shockable rhythm, the unit alerts the operator that no
shock is advised.
WARNING! ECG rhythm analysis does not warn of patient asystole, which is not a shockable
rhythm.
When a nonshockable rhythm is detected, the unit displays a
NO SHOCK ADVISED
message.
Follow the local protocols to continue CPR or other life support, and re-analyze the ECG at
appropriate intervals.
Note:
When a nonshockable rhythm is detected, the X Series does not prevent the user from
manually defibrillating the patient.
R
06/06/2011 12:34:56
Adult
00:17:43
I, II,
III...
12
CO
2
SYNC
HR
bpm
80
121
79
(96)
NIBP
mmHg
38
12
BR
CO2
mmHg
SpO2
%
97
T1
ºF
98.6
Analyzing
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...