CO2 Monitoring Setup and Use
9650-001355-01 Rev. M
X Series Operator’s Guide
9-5
WARNING! •
The exhaust port for the Microstream sidestream CO
2
apparatus is an output for only
expired gases from the patient and any connected breathing apparatus. The exhaust
port is intended only for connection to gas collection equipment, such as gas scavenger
devices --
there should be no other connections to the exhaust port
.
•
When connecting the Microstream sidestream CO
2
accessory to patients who are
receiving or have recently received anesthetics, connect the CO
2
exhaust port to a
scavenging system, or to the patient’s anesthetic machine or ventilator to prevent
exposing medical staff to anesthetics
.
•
Do not lift the monitor by the FilterLine, as it could disconnect from the monitor,
causing the monitor to fall on the patient.
•
The FilterLine may ignite in the presence of high O
2
concentrations when directly
exposed to laser or ESU devices. Use caution when performing these procedures.
Note:
If you use a gas scavenging system, ensure that it is installed according to the
manufacturers instructions. The gas scavenging system should comply with
ISO 8835-3:1997 (E).
Note:
In order to avoid moisture buildup and sampling line occlusion during nebulization or
suction for intubated patients, remove the sampling line luer connector from the
monitor.
Applying a FilterLine Set
The FilterLine set is intended for the CO
2
monitoring of intubated patients.
Before attaching the airway adapter to the breathing circuit, verify that the adapter is clean, dry,
and undamaged. Replace if necessary.
Caution
The disposable FilterLine set is intended for single patient use. Do NOT reuse or sterilize any
part of the FilterLine Set, as the monitor may be damaged by cleaning or reuse.
1. Place the airway adapter at the proximal end of the airway circuit between the elbow and the
ventilator circuit wye. Do NOT place the airway adapter between the ET tube and the elbow
as this may allow patient secretions to accumulate in the adapter.
2.
If pooling does occur, the airway adapter may be removed from the circuit, rinsed with water
and reinserted into the circuit. To prevent moisture from draining into the sample tubing,
ensure that the sampling tube exits from the top of the airway adapter, not its bottom or sides.
See the following figure.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...