How does NIBP Work?
9650-001355-01 Rev. M
X Series Operator’s Guide
8-3
How does NIBP Work?
The blood pressure cuff and hose connect to the X Series unit through the NIBP connector on
the side panel of the unit. The NIBP button on the front panel of the unit allows you to initiate
and terminate blood pressure measurements, which are displayed in the NIBP area of the
monitor.
The X Series unit’s NIBP module measures the oscillometric pulses transmitted through the
blood pressure cuff and hose, and calculates the blood pressure measurements accordingly. The
pressure measurement proceeds as follows:
1. The cuff inflates to a preconfigured pressure (determined by patient type) above the
patient’s systolic blood pressure. As the cuff inflates, the X Series unit measures the
oscillometric pulses transmitted from the cuff through the hose, and the SureBP monitoring
software calculates the patient’s systolic, diastolic, and mean blood pressure. Once these
measurements are taken, the cuff deflates. The X Series unit takes these measurements in
about 15 seconds.
The SureBP feature requires the use of a cuff and dual lumen hose.
2.
If high artifact prevents an accurate measurement during cuff inflation, or if you use a cuff
and single lumen hose, the cuff inflates to its target pressure to occlude blood flow through
the arteries in the monitored limb. The cuff deflates incrementally, allowing blood to flow
through the cuff and into the monitored limb. As blood flows through the partially deflated
cuff, it produces pressure oscillations that are transmitted to the X Series unit through the
hose.The X Series unit measures the oscillometric pulses and uses them to calculate the
corresponding systolic, diastolic, and mean blood pressure. This measurement takes about 30
seconds.
3.
The NIBP option automatically adjusts the blood pressure measurement procedure in
response to certain error conditions such as:
Condition
Adjustment/Response
The unit cannot detect systolic
pressure
The unit automatically increases the cuff inflation
pressure and completes the blood pressure mea-
surement
The unit cannot detect systolic,
diastolic or mean pressure after
180 seconds
The unit aborts the blood pressure measurement and
deflates the cuff
The unit detects a fault
The unit displays a corresponding error message on
the monitor and aborts the measurement
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...