Specifications
A-12
www.zoll.com
9650-000820-01 Rev. K
Typical Determination time without Artifact:
Measurements on the deflation:
30 to 45 seconds
Measurements on the inflation (SureBP)*:
15 to 30 seconds
* using dual lumen cuffs
Maximum Determination Time - Measurement on the Inflation
Adult: 150 seconds
Pediatric:120 seconds
Neonatal:80 seconds
Blood Pressure Validation
:
Blood pressure measurements determined with this device are equivalent to those obtained by a
trained observer using the cuff/stethoscope auscultation method for adults and pediatric patients
and equivalent to an intra-arterial measurement for neonatal patients, within the limits
prescribed by the American National Standards Institute (ANSI-AAMI SP10). To receive a
copy of the report containing the AAMI SP10 results, contact the ZOLL Technical Service
Department.
NIBP Accuracy:
NIBP accuracy has been verified utilizing clinical test methods which have been determined to
meet the requirements of EN ISO 81060-2:2012.
Invasive Pressures
Number of Channels
: 3
Pressure range
: -30 to 300 mmHg
Pressure Accuracy
: +/- 2 mmHg or 2% of reading, whichever is greater, plus transducer error.
Pulse Rate Range:
25 to 250 BPM
Pulse Rate Accuracy:
+/- 3 BPM, or +/- 3% of value whichever is greater
Pulse Rate Display:
Average of last 4 beat-to-beat intervals.
Zero Adjust:
+ / - 200 mmHg
Transducer
:
Sensitivity: 5uV/V/mmHg
Offset: +/- 125 mmHg including transducer offset
Excitation Impedance Range: 150 to 10,000 ohms
Excitation Voltage: 4.75 +/- 0.25 VDC
Connector: 6-pin circular MS3100 series
Connect to:
A
B
C
D
E
Signal Type
Sig (-)
Exc (+)
Sig (+)
Exc (-)
shield