Operating Controls and Indicators
9650-1210-01 Rev. T
2-5
as well as space for patient name, date and comments. On the last event recorded, “SUMMARY COMPLETE” is
printed at the bottom left of the recorder strip.
Defibrillation Format
The summary report function records 6 seconds of pre-shock and 9 seconds of post-shock patient ECG data. Also
recorded are joules selected, joules delivered, sync if active (includes sync markers), ECG lead, ECG size, patient
impedance, actual event time and date. The date/time printed on the top of the strip corresponds to the ECG data
occurring 6 seconds before the event. AED units additionally include shock count and AED mode annotations.
Actual event date/time