Warning!
▲
Use the AED Plus
®
unit only as described in this manual. Improper use of the
device can cause death or injury.
▲
DO NOT use or place the AED Plus unit in service until you have read the AED
Plus Operator’s and Administrator’s Guides.
▲
DO NOT use or place the AED Plus unit in service if the unit’s status indicator
window (located on the left side of the handle) displays a red “X”.
▲
DO NOT use or place the AED Plus unit in service if the unit emits a beeping tone.
▲
Connect the electrode cable to the AED Plus unit after installing batteries.
▲
Keep the electrode cable connected to the AED Plus unit at all times.
▲
This device should only be used by properly trained individuals.
▲
Only use electrodes labeled “Infant/Child” on children less than 8 years old or
weighing less than 55 lbs (25 kg). Use
CPR-D•padz
®
if patient is older than 8 years
or weighs more than 55 lbs (25 kg).
▲
Always stand clear of patient when delivering treatment. Defibrillation energy
delivered to the patient may be conducted through the patient’s body and cause a
lethal shock to those touching the patient.
▲
DO NOT touch the electrode surfaces, the patient, or any conductive material
touching the patient during ECG analysis or defibrillation.
▲
Move patient away from electrically conductive surfaces prior to use of equipment.
▲
DO NOT use the unit near or within puddles of water.
▲
Keep the patient as motionless as possible during ECG analysis.
▲
DO NOT use the unit near flammable agents, such as gasoline, oxygen-rich
atmospheres, or flammable anesthetics.
▲
Avoid radio frequency interference from high-power sources that might cause the
defibrillator to interpret cardiac rhythms incorrectly by turning off cell phones and
2-way radios.
▲
Disconnect non-defibrillation protected electronic devices or equipment from
patient before defibrillation.
▲
Dry victim’s chest, if wet, before attaching electrodes.
▲
Apply freshly opened and undamaged electrodes, within the electrode expiration
date, to clean and dry skin to minimize burning.
▲
DO NOT place electrodes directly over the patient’s implanted pacemaker.
Pacemaker stimuli may degrade the accuracy of ECG rhythm analyses or the
pacemaker may be damaged by defibrillator discharges.
▲
Check labeling inside the ZOLL
®
AED Plus cover before using the cover as a
Passive Airway Support System (PASS) device to ensure it is intended for this
use.
▲
DO NOT use the Passive Airway Support System (PASS) if there is a suspected
head or neck injury. Place the patient on a firm surface before performing CPR.
▲
DO NOT recharge, disassemble, or dispose of batteries in fire. Batteries may
explode if mistreated.
▲
DO NOT use or stack the AED Plus unit with other equipment. If the unit is used
or stacked with other equipment, verify proper operation prior to use.
Caution!
▲
DO NOT disassemble the unit. A shock hazard exists. Refer all servicing to
qualified personnel.
▲
Use only commercially available type 123A lithium manganese dioxide batteries.
Discard batteries properly after removal from unit. Use only batteries from
recommended manufacturers. See the AED Plus Administrator’s Guide (P/N
9650-0301-01) for a list of recommended battery manufacturers.
▲
If the device is stored outside the recommended environmental conditions, the
electrode pads and/or batteries may be damaged or their useful life reduced.
▲
The
CPR-D•padz
Electrode can be connected to other ZOLL defibrillators with
Multifunction Cables. Defibrillation can be administered when connected to other
ZOLL defibrillators.The CPR function does not operate with any device other than
the AED Plus defibrillator.
Important!
This symbol indicates that an AED Plus unit is equipped for treating adult and
pediatric patients. An AED Plus unit without this symbol is not equipped to treat
pediatric patients and will NOT work with
pedi•padz
®
II
pediatric electrodes. To
upgrade an AED Plus unit for use with ZOLL
pedi•padz II
pediatric electrodes,
contact ZOLL Medical Corporation or an authorized ZOLL distributor for information
on the ZOLL AED Plus Pediatric Upgrade Kit.
Set-up and Check-out Procedure:
1. Insert 10 new batteries into AED Plus unit.
2. Connect electrode cable to AED Plus unit and pack sealed electrodes inside unit
cover. Close cover.
3.
Turn unit on and wait for “Unit OK” audio message. Verify that unit issues
appropriate “Adult Pads” or “Pediatric Pads” audio message.
4. Turn unit off.
5. Wait 2 minutes. Verify that green check symbol (
) appears in status indicator
window (located on left side of handle) and that unit does not emit a beeping
tone.
6. Place AED Plus unit in service.
7. Check AED Plus unit periodically to ensure that green check symbol (
)
appears in status indicator window.
Battery Replacement
For AED Plus units running software version 5.32 or higher, replace batteries every 5
years or if unit prompts. For earlier software versions, replace batteries every 3
years and place a dated AED Plus battery replacement reminder label next to the
On/Off button. These labels are available from ZOLL Customer Service. Use only
type 123A lithium manganese dioxide batteries from recommended manufacturers.
■
Remove all batteries from battery compartment and discard before installing any
new batteries.
■
Insert 10 new batteries into battery well. Do not use old batteries.
■
Press button in battery well only after installation of new batteries.
Cleaning
■
Clean and disinfect unit with soft, damp cloth using 90% isopropyl alcohol or soap
and water, or chlorine bleach (30 ml/liter water).
■
Do not immerse any part of the unit in water.
■
Do not use ketones (MEK, acetone, etc.).
■
Avoid using abrasives (e.g., paper towels) on the LCD display, if so equipped.
■
Do not sterilize the unit.
Federal (U.S.A.) law restricts this device for sale to or on the order of a physician.
AED Plus
®
Automated External Defibrillator
Operator’s Guide
AHA 2010
For Technical support or repair:
ZOLL Medical Corporation
ZOLL Medical Europe
269 Mill Rd.
Newtonweg 18
Chelmsford, MA 01824-4105
6662 PV ELST
978-421-9655 • 800-348-9011
The Netherlands
Fax: 978-421-0010
+31 (0) 481 366410
Fax: +31 (0) 481 366411
International Service:
Contact your local distributor.
0123
©
2012 ZOLL Medical Corporation
REF
9650-0300-01 Rev. P
TROUBLESHOOTING
Problem
Recommended Action
Self-test failed.
Manually test by pressing and
holding the ON/OFF button for
more than 5 seconds. If unit fails
test again, remove from service.
“Change batteries”
prompt.
Replace all batteries at the same
time.
Red “X” in Status Indicator window
OR
beeping noise when unit is OFF.
Perform manual test. Check to
see if cable is attached properly
to unit. Replace batteries. If unit
still does not operate correctly,
remove from service.
Red “X” in Status Indicator window
when unit is ON.
Power cycle the unit. If Red “X” is
still present in Status Indicator
window, remove unit from
service.
www.AmericanAED.com