6/26/2019
1
ZOLL 731
Easy Set Up Guide
2
Vent Overview
Startup Switch
Vent parameter buttons
Confirmation button
Alarm Silence/ Cancel Action
Rotary Selection Knob
Single Manual Breath
Vent Status Indicator
Page 1: ...6 2019 1 ZOLL 731 Easy Set Up Guide 2 Vent Overview Startup Switch Vent parameter buttons Confirmation button Alarm Silence Cancel Action Rotary Selection Knob Single Manual Breath Vent Status Indicat...
Page 2: ...runs light should turn green 4 Select mode 1 Push button once to highlight mode 2 Rotate knob to AC 3 Press confirm button 4 Push button twice to highlight control method 5 Rotate knob to V 6 Press c...
Page 3: ...ss confirm button 4 Press button twice to highlight High alarm 5 Rotate knob to 10 above set 6 Press confirm button 7 Press button three times to highlight Low alarm 8 Press confirm button 6 Set Volum...
Page 4: ...ect high pressure 10 15 higher than outlined number not 40 3 Press confirm button 4 Press button twice 5 Rotate knob to select low pressure 10 15 below outlined number 6 Press confirm button Peep defa...
Page 5: ...knob to desired FiO2 3 Press confirm button ABG and Vent Changes Normal Ranges pH 7 35 7 45 PaCO2 35 45 HCO3 22 26 PaO2 80 100 Increase RR or VT to lower PaCO2 increase pH Decrease RR or VT to raise P...
Page 6: ...6 26 2019 6 Recommended adjunct placement 11 1 Suction 2 Emma 3 HME 4 MDI 5 Neb 1 2 3 4 5...