![ZOLL 330 Operator'S Manual Download Page 76](http://html1.mh-extra.com/html/zoll/330/330_operators-manual_3993212076.webp)
Specifications
A-4
www.zoll.com
906-0330-01-01 Rev. D
Emissions and Immunity Compliance
The ZOLL Aspirator is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Be prepared to use a manual aspiration technique in event of device interference or failure.
Warning!
Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they are operating normally.
The device does not have external sensors or transducers that are likely to affect product
compliance.
Warning!
Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper
operation.
Warning!
Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any party of
the aspirator, including cables specified by the manufacturer. Otherwise, degradation of
the performance of this equipment could result.
The aspirator should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the ZOLL
Aspirator. The device alarms may be issued in event the device is affected by electromagnetic
disturbance.