4-2
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
April 2019
Respiration Monitoring
Figure 19. RR Waveform and Value
To manually test RR alarm functionality on a daily basis, you may choose from
two approaches. First, if a patient’s respiration is currently being monitored,
make certain that the alarms are enabled, and then lower the RR upper alarm limit
setting below its current value. Verify that "RR < [upper limit]" annunciates as a
medium grade alarm. Press the ALARM SILENCE key, and then return the RR
upper alarm limit to its previous value. Verify that the alarm is no longer active,
and that RR monitoring continues normally. Alternatively, disconnect the RA or
LL electrode. Verify that "RR lead off" (interleaved with "HR lead off")
annunciates as a low grade alarm. Press ALARM SILENCE and reconnect the
electrode. Verify that the alarm ceases and that the RR waveform and parameter
return.
The Respiration Monitoring settings and specifications for the Nightingale PPM3
may be found in the PPM3 Monitor Settings and Technical Data chapters.
Procedures for changing configuration settings, such as enabling alarms, or
adjusting alarm limits, may be found in the Working With Menus chapter.
WARNING
– Operating the Nightingale PPM3 with RR limit alarms
disabled means that no low or high RR alarm conditions will produce
alarm notifications. Use this feature with extreme caution. Patients
must be closely observed if RR limit alarms are disabled.
WARNING
–
When using an ECG electrode to calculate respiration
rate via the thorax impedance method, movement artifacts
may create inaccurate results. Respiration rates derived from CO
2
parameter is not subject to such movement artifacts.
Note
– If the Respiration waveform is not currently selected for display on the PPM3,
follow the instructions in the chapter on Working With Menus to select this waveform for
display.
Summary of Contents for PPM3
Page 1: ...Nightingale Monitoring System PPM3 User s Guide ...
Page 4: ......
Page 16: ......
Page 28: ......
Page 32: ...2 4 Nightingale PPM3 User s Guide April 2019 Getting Started Figure 10 PPM3 Startup Screen ...
Page 52: ......
Page 96: ......
Page 100: ......
Page 138: ......
Page 160: ......
Page 164: ......
Page 184: ......
Page 194: ......
Page 196: ......
Page 198: ......