12

  / 

35

 

 

Do not use the MagPurix 24 EVO System under direct sunlight or close to a heating 
device. It may shorten the life of the MagPurix 24 EVO System, or cause malfunction.

 

 

Do not open the maintenance doors and the front doors while operating the instrument.

 

3.3

 

Installation Procedures 

3.3.1

 

Unpacking 

The MagPurix 24 EVO System should be unpacked and installed by authorized Zinexts Life Science 
Corp. service personnel only. In this case, you need to take no further action until the on-site arrival 
of authorized Zinexts Life Science Corp. service personnel.

 

Should this not be possible, use the following procedure to install the instrument.

 

 

Caution: 

 

 

The MagPurix 24 EVO System weights 100 kg. It should be lifted with a pallet jack or 
fork lifter and moved by two persons at minimum.

 

 

Do not hold the front panel or the door when moving the instrument.

 

 

Improper movement might lead to instrument damages.

 

 

For correct and safe use of the MagPurix 24 EVO System, install the device in a location 
close to a power socket, and with enough space for installation procedures and main 
switch operation.

 

 

Open the wooden transport box by removing the top cover and front/side walls.

 

 

Carefully remove the outer wrapping. Move the MagPurix 24 EVO System with a pallet jack or 
fork lifter to a level ground. 

 

 

Move the instrument by holding the handles on both sides of the instrument. 

 

 

Move the MagPurix 24 EVO System to its desired location, and move down the leveling feet to 
fixate and level the instrument in its location. 

 

 

After system initialization, remove the fixing material (white foam). 

 

3.3.2

 

Control Unit 

Follow the steps below to release the fixing material inside the instrument:

 

 

Plug the power cord to the instrument and connect to the electric outlet.

 

 

Connect the barcode reader to the instrument.

 

 

Turn on the power of the MagPurix 24 EVO System with the switch on the right side of the 
instrument.

 

 

The  system  will  start  the  initialization  and  it  will  release  the  fixing  material  (white  foam) 

Summary of Contents for MagPurix 24 EVO

Page 1: ...rp User Manual Automated Nucleic Acid Purification System 24 EVO Catalog No ZP01019 Manual No QW IH01 20 Version 3 0 Obelis s a Bd G n ral Wahis 53 1030 Brussels Belgium Tel 32 2 732 59 54 Fax 32 2 73...

Page 2: ...Safety Symbols Labels 5 1 3 2 About the Instrument 5 1 3 3 About Reagent Kits 6 1 3 4 About Samples 7 1 3 5 About Infectious Waste 7 2 Intended Use General Description 8 2 1 Intended Use 8 2 2 Externa...

Page 3: ...Operation 19 5 1 Preparation 19 5 2 Start up the System 20 5 2 1 Basic Settings 20 5 2 2 User Management 20 5 2 3 Door Lock and Door Sensor Switch 20 5 3 Perform Extraction 21 5 4 System Fixing before...

Page 4: ...www zinexts com MagPurix 24 EVO User Manual 3 35 9 2 Technical Support 34 10 Warranty Information 34 Revision History 35...

Page 5: ...gPurix 24 EVO System is designed for the automated extraction and isolation of nucleic acids utilizing advanced magnetic bead separation technology 1 2 2 Applications The MagPurix 24 EVO System is int...

Page 6: ...indicates that certain precautions must be taken when working with potentially infectious material Hot Surface This symbol labels potentially hot surfaces on the instrument Electricity This symbol is...

Page 7: ...bjects on the MagPurix 24 EVO System Also refrain from bumping or knocking it as doing so may cause a failure or malfunction of the instrument If any liquid materials are left or have leaked inside th...

Page 8: ...3 5 About Infectious Waste Caution When handling or disposing of infectious materials follow the laboratory guidelines or the law regarding infectious waste to perform proper incineration fusion steri...

Page 9: ...ple types The MagPurix 24 EVO System is intended to be used only in combination with extraction kits indicated for use with MagPurix System for the applications described in the respective kit handboo...

Page 10: ...www zinexts com MagPurix 24 EVO User Manual 9 35 2 3 Internal Features...

Page 11: ...arts MagPurix 24 EVO System x 1 Power Cord x 1 Sample Rack for 2 ml Sample Tube x 2 Barcode Reader x 1 User Manual x 1 SD Card 32GB x 1 64GB x 1 Fuse x 3 1A 2A 4A Please make sure all components are f...

Page 12: ...up and down suddenly Warming a cold room suddenly or moving the MagPurix 24 EVO System from a room with low temperature to a warm room may cause condensation inside the device resulting in abnormal op...

Page 13: ...might lead to instrument damages For correct and safe use of the MagPurix 24 EVO System install the device in a location close to a power socket and with enough space for installation procedures and...

Page 14: ...required for extraction are provided with the respective extraction reagent kits Contents of the kits will vary please refer to the resp extraction kit handbook for details 3 3 4 User Training User t...

Page 15: ...wering up the MagPurix 24 EVO System the MagPurix 24 EVO System will start the initialization process Click LOGIN and User ID and Password will appear for input Set up the local time System will start...

Page 16: ...nual 15 35 Camera time is suggested to be set for the first time use Follow the steps below to set the time Step 1 Remove the Camera as below Step 2 Turn on the instrument and click on Settings System...

Page 17: ...35 Step 4 Find the buttons from the left side of the camera Step 5 Camera will start recording automatically after sensor is turned off Press OK button to pause Step 6 Long press M button to access to...

Page 18: ...ser Manual 17 35 Step 8 Select Date Time and press OK Step 9 Use the button to set the time and press OK Step 10 Press M button to return to the main menu Step 11 On the panel click CAMERA to turn on...

Page 19: ...access to users to certain options Once logged in you will be able to set up at least 10 other users with their own passwords 4 2 Main Screen The Main Screen of the MagPurix 24 EVO system on the righ...

Page 20: ...he processed protocols and or some samples used with this instrument may contain infectious agents Caution Before start operating the instrument put on appropriate gloves a mask and safety goggles if...

Page 21: ...are version installed is displayed at the bottom of the screen by clicking the Status Bar 5 2 2 User Management After logging in as administrator the Main Screen opens Click the Settings button and on...

Page 22: ...MagPurix 24 EVO Software the Main Screen opens Click the Perform Run button and on the Perform Run Screen click the Extraction button Place reaction cartridges and all plastic consumables onto the wor...

Page 23: ...er to separate reagent cartridge strips reagent chambers and tip holders if you do not run full batches of samples The position of Piercing Pin on the right and Filter Tip on the left are as follows T...

Page 24: ...www zinexts com MagPurix 24 EVO User Manual 23 35 Note Zinexts does not guarantee with expired kit performance...

Page 25: ...al To amend ID information DEL the wrong ID information and then scan the right ID Please take the Sample Tube Rack with the eluates and the samples out of the extraction module and discard all used e...

Page 26: ...and 75 ethanol Clean the heating module with a soft paper tissue or soft cloth and use a cotton swab to remove any potential buffer residue from the hold of the heating module Run disinfection by tur...

Page 27: ...tatus of the piston units Cleaning of piston UV light change annually Service Engineer 6 1 Cleaning of Rubbish Tray and M Motor Elevate M motor before the start of cleaning Clean the rubbish tray and...

Page 28: ...24 EVO User Manual 27 35 6 2 Cleaning of Piston Enter Settings menu to select System and then click Piston Clean piston with 75 ethanol using soft cloth Change new piston if there is significant impur...

Page 29: ...urns on when the power is on 1 The packing materials are not removed 2 Technical problem 1 Turn off the power and remove the packing materials 2 Contact your local representative or agent Protocol sto...

Page 30: ...eparation of beads Too much sample material Reduce amount of sample material to the values recommended in the resp extraction kit handbook Carryover of magnetic beads into eluate Too much sample mater...

Page 31: ...leakage from the tips 1 Perform preventive maintenance annually to ensure proper motor movements 2 Re load the consumables according to the instructions in this manual Low DNA yield Incomplete lysis D...

Page 32: ...contain any particulate material that can clog the tip If needed centrifuge the sample prior to the extraction process No DNA recovered Magnetic beads stored or handled improperly Store cartridge cont...

Page 33: ...event bubble formation during mixing make sure the sample volume is at least at the recommended volume listed in the resp extraction kit handbook Purified DNA repeatedly frozen and thawed Aliquot puri...

Page 34: ...0 C assuming the room temperature at 25 C Protocol Input Touch Panel UV Light 30 minutes at 250 nm Built in Features 7 Touch Panel Instrument Dimensions 91 5 cm W x 66 5 cm D x 60 4 cm H Weight 100 kg...

Page 35: ...at each product described herein will be free from defects in materials and workmanship for a period of twelve 12 months from the date of delivery Zinexts Life Science Corp agrees as its sole responsi...

Page 36: ...ding on or obligate Zinexts Life Science Corp in any way In no event will Zinexts Life Science Corp be liable for any special incidental or consequential damages resulting from use or malfunction of t...

Reviews: