CLEANING, DECONTAMINATION AND STERILIZATION
16
Validated manual cleaning and disinfection procedure
STEP STEP DESCRIPTION STEP INSTRUCTION
ACCESSORIES
DURATION
Cleaning
Steps
1
Contamination
Removal
Rinse product with RO or
distilled water removing
any visible organic
material. Apply detergent
with a non-shedding cloth
and as necessary, with
assistance of a soft bristle
to brush any visible soil.
− Cold/Room temperature
RO or distilled water
−
Soft bristle brush
−
Non-shedding cloth
− Recommended pH-value
detergent/ disinfectant
Until all visible soil is
removed.
Minimum of two (2)
minutes is recommended.
2
Rinse
Rinse product with RO or
distilled water.
− Cold/Room temperature
RO or distilled water
Until all visible detergent
is removed.
3
Drying
Dry the device utilizing a
dry non-shedding cloth.
Medical quality filtered
air may be utilized if
available.
−
Non-shedding cloth
− Medical quality filtered
compressed air
Until product is visually
dry.
Disinfection
Steps
4
Disinfection
Application
Apply the recommended
pH-value detergent/
disinfectant
to a non-shedding cloth.
Cover the device’s surface
area with the saturated
cloth.
− Recommended pH-value
detergent/ disinfectant
−
Non-shedding cloth.
−
Spray bottle or other
manual applicator
Minimum of thirty
(30) minutes is
recommended.
5
Manual
Disinfection
After 30 minutes contact
time, wipe the device
contact surfaces, joints
and mated areas utilizing
the cover cloth.
− Recommended pH-
value detergent/
disinfectant
−
Non-shedding cloth
Manual disinfection
is complete when the
device’s surface, joints
and crevices have been
manually wiped.
6
Final Rinse
Rinse product under
room temperature RO or
distilled water.
−
Room temperature RO
or distilled water
Minimum of 30 seconds.
7
Final Drying
Dry the device utilizing a
dry non-shedding cloth.
Medical quality filtered
air may be utilized if
available.
−
Non-shedding cloth
− Medical quality filtered
compressed air
Until product is visually
dry.
WARNING
:
Zimmer
® UNIVERSAL Power System should not be immersed or submerged in water or cleaner.
PREPARATION OF AUTOMATED CLEANING MATERIALS
The automated cleaning and disinfection process has been validated according to AAMI TIR Number 30 and ISO 17664 guidelines.
CAUTION
: The manual cleaning and disinfection procedure must be performed
prior to the automated cleaning and
disinfection procedure.
Materials:
Detergent with neutral pH or lightly alkaline: pH < 9
Neutralizing agent
Rinse aid
Automated washer
The UNIVERSAL lift out tray (REF 89-8510-459-48) and the UNIVERSAL sterilization case (REF 89-8510-459-47) are also sold as
a kit (REF 89-8510-459-46). The use of the lift out tray (REF 89-8510-459-48) has been validated and is recommended for the
automated cleaning and disinfection procedure following the manual cleaning and disinfection procedure.