Operator
& Service Manual
A.T.S.
™
1200
TOURNIQUET SYSTEM
REF 60-1200-101-00
Page 1: ...Operator Service Manual A T S 1200 TOURNIQUET SYSTEM REF 60 1200 101 00...
Page 2: ...Inc be liable for any special incidental or consequential damages whether based on breach of warranty or other legal theory whether or not such damages are foreseeable Some states do not allow limita...
Page 3: ...ibration Check 8 2 5 Pressure and Time Defaults 9 2 6 Single Cuff Operation 10 2 7 Dual Cuff Operation 11 2 8 Bier Block Cuff Operation IVRA 11 2 9 Alarm Conditions 11 3 0 MAINTENANCE 3 1 General Main...
Page 4: ...erview front 23 2 A T S 1200 Tourniquet Overview rear 24 3 Calibration Setup Overview 25 4 Disassembly front case opening 26 5 Disassembly front case removal 26 6 Disassembly disconnecting wiring 26 7...
Page 5: ...pressure set point Can also be pressed to verify the set point TIME Button Used in conjunction with the ROTARY knob to adjust the time alarm set point Can also be pressed to verify the set point MAIN...
Page 6: ...nt light emitting diodes LED Displays pressure setting sensed cuff pressure and hardware failure conditions other messages TIME Display Red 14 segment light emitting diodes LED Displays time alarm set...
Page 7: ...ll cause bacteria exotoxins or malignant cells to spread to the general circulation or where it could dislodge thromboemboli that may have formed in the vessels The tourniquet cuff must be applied in...
Page 8: ...body level Whenever IVRA Bier Block anesthesia is used it is recommended that the tourniquet remain inflated for at least 20 minutes from the time of injection WARNING Cuffs will not deflate in STANDB...
Page 9: ...ely 2 seconds before the unit will allow a cuff to deflate 7 SECOND CUFF INFLATE button Inflation of the Second cuff is initiated by depressing the blue INFLATE button 8 SECOND CUFF DEFLATE button Def...
Page 10: ...ys followed by the text ATS 1200 on the respective display b The unit emits tones when 0 0 0 0 and text is displayed c The unit will display SELF TEST for an instant The unit is self testing specific...
Page 11: ...e the pressure to 250 mmHg The PRESSURE display should read 250 5 mmHg n Increase the pressure to 475 mmHg The PRESSURE display should read 475 5 mmHg o Remove the calibration hose from the unit The P...
Page 12: ...ressure during the procedure 4 Prepare the patient in accordance with your established procedures and cuff manufacturer s instructions The precautions of Section 1 and the following are offered as a g...
Page 13: ...re is a significant leak in the original cuff this feature could cause the inflation rate of the subsequent cuff to be longer and perhaps even cause the 30 second inflation alarm to sound The display...
Page 14: ...ude that a hardware failure has occurred rendering the unit unusable The appropriate action is to set the unit to STANDBY by pressing the ON STANDBY button Since this removes power from the internal i...
Page 15: ...the set value or greater at power up to determine if it needs resetting CUFF NOT DEFLATED normal CUFF Check for kinks in hose If alarm persists Pressure in deflated cuff is NOT disconnect hose from c...
Page 16: ...ice the unit combination occurred WATCHDOG FAILURE WDT FAIL Cycle the ON STANDBY Button If problem Windowing watchdog system persists service the unit detected a malfunction SYSTEM FAILURE SYS FAIL Cy...
Page 17: ...very six months 1 Cleaning The exterior of the unit may be cleaned with a cloth that has been dampened not dripping with a mild detergent The interior of the unit may be vacuumed or blown out as requi...
Page 18: ...50 mm Hg to the unit s Main cuff port Once the pressure has stabilized press the PRESSURE button so the unit can calibrate the 250 mm Hg point The unit will beep to let the user know the set point was...
Page 19: ...steps in Section 2 4 Step 5 Calibration Check before using this unit on a patient 3 5 LEAK TESTING The A T S 1200 Tourniquet is capable of keeping a cuff with a substantial leak inflated Naturally it...
Page 20: ...ilure that may occur is when an audible alarm occurs that cannot be silenced by the ALARM SILENCE button The valve and pump will be disabled which seals off the cuff to prevent pressure loss The displ...
Page 21: ...not properly plugged into P8 f Defective AC MAINS indicator g Defective AC MAINS indicator circuitry 4 No flashing orange BATTERY indicator light a Unit running on AC b Membrane Panel not properly plu...
Page 22: ...ve battery charging circuitry 11 AMP FAIL alarm a Transducer s amplifier out of range b Battery fully depleted or defective c Extremely high pressure exerted on transducers 12 Unit cannot be set to ST...
Page 23: ...ry use only TP19 0 VDC N A Digital ground TP20 N A N A Factory use only TP21 26 4 to 27 5 VDC 1 0 VDC Boost supply for battery charger TP22 5 VDC 0 5 VDC Supply voltage for battery charger TP23 5 VDC...
Page 24: ...263 Calibration Hose 0600 2002018 Plumbing Assembly 60 0938 003 00 A T S Stand 5 Wheel Caster Kit 60 2360 001 00 A T S Pole Clamp Knob w Screw 60 3000 001 00 A T S Cal Regulator Kit 60 7000 027 00 Bac...
Page 25: ...ivation Button G ON STANDBY Button H Green AC MAINS Indicator I Orange BACKUP BATTERY Indicator J SECOND CUFF INFLATE DEFLATE Buttons K Second Cuff Port L Main Cuff Port M MAIN CUFF INFLATE DEFLATE Bu...
Page 26: ...R Power Cord S Cord Retaining Clip T Factory Test Port U Pole Clamp V Mains Fuse Block 24 2...
Page 27: ...W Calibrated Pressure Meter with a range of 0 to 500 mm Hg X Pressure Regulator Source adjustable from 0 to 500 mm Hg Y Calibration Hose Included with Unit 25 3...
Page 28: ...the front case 1 Carefully slide the front off the rear Disconnecting the wiring 1 Use caution when disconnecting the wiring Mark the wires before disconnecting or see Section 3 2 for proper re connec...
Page 29: ...Rear Case separation 1 All components are easily accessed when the rear case is removed Reverse the process to reassemble Be extremely careful not to pinch any wiring or tubing when reassembling 8 27...
Page 30: ...28 9 Control Board Layout...
Page 31: ...Equipment with respect to electric shock fire and mechanical hazards only in accordance with UL 60601 1 CAN CSA C22 2 No 601 1 and IEC 60601 1 This product contains electrical or electronic materials...
Page 32: ...30 10...
Page 33: ...SCHEMATICS...
Page 34: ...Revised 04 06 Zimmer Orthopaedic Surgical Products 2002 2006 Zimmer Orthopaedic Surgical Products Inc 200 West Ohio Avenue Printed in U S A P O Box 10 62 8000 015 00 ENG Dover Ohio 44622 U S A 0123...