INTRODUCTION

INTRODUCTION

CAUTION

CAUTION

* The service life of the cuff may vary by the frequency of washing, skin condition, and 

storage state. The typical service life is 10000 times.

* It is recommended that the performance should be checked every 2 years and after 

maintenance and repair, by retesting at least the requirements in limits of the error of the 

cuff pressure indication and air leakage (testing at least at 50mmHg and 200mmHg).

* Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according 

to the local guidelines.

* Manufacturer will make available on request circuit diagrams, component part lists, 

descriptions, calibration instructions,etc., to assist to service personnel in parts repair.

* The plug/adapter plug pins insulate the device from the main supply. Do not position the 

device in a position where it is difficult to disconnect from the supply mains to safely 

terminate operation of ME equipment.

* The operator shall not touch output of batteries/adapter and the patient simultaneously.

* Cleaning :Dust environment may affect the performance of the unit. Please use the soft 

cloth to clean the whole unit before and after use. Don’t use any abrasive or volatile 

the storage and operating conditions described in this booklet.

* The equipment is not AP/APG equipment and not suitable for use in the presence of a 

flammable anesthetic mixture with air of with oxygen or nitrous oxide.

* Warning: No servicing/maintenance while the ME equipment is in use.

* The patient is an intended operator.

* The patient can measure data and change batteries under normal circumstances and 

maintain the device and its accessories according to the user manual.

* To avoid measurement errors, please avoid the condition of strong electromagnetic field 

radiated interference signal or electrical fast transient/burst signal.

* The blood pressure monitor and the cuff are suitable for use within the patient 

environment. If you are allergic to polyester, nylon or plastic, please don't use this device. 

* During use, the patient will be in contact with the cuff. The materials of the cuff have been 

tested and found to comply with requirements of ISO 10993-5:2009 and ISO 

10993-10:2010. It will not cause any potential sensation or irritation reaction.

* Adaptor is specified as a part of ME EQUIPMENT.

* If you experience discomfort during a measurement, such as pain in the arm or other 

complaints, press the START/STOP button to release the air immediately from the cuff. 

Loosen the cuff and remove it from your arm.

* If the cuff pressure reaches 40 kPa (300 mmHg), the unit will automatically deflate. 

Should the cuff not deflate when pressure reaches 40 kPa (300 mmHg), detach the cuff 

from the arm and press the START/STOP button to stop inflation.

* Before use, make sure the device functions safely and is in proper working condition. 

Check the device, do not use the device if it is damaged in any way. The continuous use of 

a damaged unit may cause injury, improper results, or serious danger.

* Do not wash the cuff in a washing machine or dishwasher!

Summary of Contents for UAM-901

Page 1: ...of this product Keep the manual well for further reference in case you have problems 2020 08 18 1 2 3 4 5 6 70g 125 90 mm TMB 1872 T 014 2020 08 18 2020 08 18 2020 08 18 TMB 1872 T 014 GB 01 PO 2F647...

Page 2: ...ecords INFORMATION FOR USER 38 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 40 What are systolic pressure and diastolic pressure What is the standard blood pressure classification Irregular...

Page 3: ...ription The ZEWA Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and heartbeat rate with arm circumference ranging from 22cm to 42cm about 8 16 It is intended...

Page 4: ...atients undergoing intravascular therapy or arterio venous shunt or people who received a mastectomy Please consult your doctor prior to using the unit if you suffer from illnesses The device is not s...

Page 5: ...ed during measurement open the cuff immediately Prolonged high pressure cuff pressure 300mmHg or constant pressure 15mmHg for more than 3 minutes applied to the arm may lead to an ecchymosis Please ch...

Page 6: ...rning No servicing maintenance while the ME equipment is in use The patient is an intended operator The patient can measure data and change batteries under normal circumstances and maintain the device...

Page 7: ...must only be serviced repaired and opened by individuals at authorized sales service centers Please report to ZEWA if any unexpected operation or events occur Keep the unit out of reach of infants you...

Page 8: ...blood pressure Low battery Batteries are low and need to be replaced mmHg Pulse in beats per minute Pulse display Blood pressure monitor is detecting an irregular heartbeat during measurement Irregula...

Page 9: ...icator Indicate the blood pressure level Heartbeat Blood pressure monitor is detecting a heartbeat during measurement The voice switch Indicating voice switch Monitor Components Component list of pres...

Page 10: ...tor which complies with local safety standard 1 Battery powered mode 6VDC 4 AAA batteries 2 CAUTION Please use the AC adaptor and USB cable just like the following picture not included AC adaptor powe...

Page 11: ...ire Batteries may explode or leak Remove batteries if the device is not likely to be used for some time Worn batteries are harmful to the environment Do not dispose with daily garbage Remove the old b...

Page 12: ...ight year press SET button to set down and turn to next step 4 Repeat steps 2 and 3 to set the MONTH and DAY 5 2 Press MEM button to change the YEAR Each press will increase the numeral by one in a cy...

Page 13: ...done and then it will turn off Select the User 1 2 3 When the monitor is off press and hold MEM button to enter user setting mode The user ID will blink Then press MEM button again select the user ID...

Page 14: ...to enter the mode for turn on or turn off the voice The trumpet will blink 2 Press MEM button again to select turn on or turn off the voice It will speak Voice on or Voice off at the same time Turn of...

Page 15: ...o expose the skin Make sure your sleeve is not too tight 2 3 Sit comfortably with your tested arm resting on a flat surface Place your elbow on a table so that the cuff is at the same level as your he...

Page 16: ...lay Inflating and measuring Display and save the results When the volume is on It will broadcast your measuring result Start the Measurement 2 Press Start Stop button to power off otherwise it will tu...

Page 17: ...splay the latest record instead Take user 1 for example When the volume is on it will read the number for you 2 Press MEM button or SET button to get the record you want When the volume is on it will...

Page 18: ...ent you can delete results by following steps below A Delete one record 1 Hold pressing set button except the AVG for 3 seconds when the monitor is in the memory recall mode the flash display dEL one...

Page 19: ...he memory recall mode the flash display dEL ALL User ID will show 2 Press and hold Start stop button to confirm deleting and the monitor will display dEL donE USER ID and then turn off B Delete all re...

Page 20: ...in the following circumstances Within 1 hour after dinner or drinking Within 20 minutes after taking a bath In a very cold environment Immediate measurement after tea coffee smoking When talking or m...

Page 21: ...sing wet cloths to remove dirt Avoid touching water clean it with a dry cloth in case Do not attempt to clean the reusable cuff with water and never immerse the cuff in water What is the standard bloo...

Page 22: ...value of them If there are two or more pulse intervals the difference between each interval and the average is more than the average value of 25 or there are four or more pulse intervals the differen...

Page 23: ...3 deep breaths before beginning will be better for measuring Advice Relax yourself for 4 5 minutes until you calm down This section includes a list of error messages and frequently asked questions for...

Page 24: ...asure again If the problem persists contact your physician Power supply Battery powered mode 6VDC 4 AAA batteries AC adaptor powered mode 5V 1A Not included Please only use the recommended AC adaptor...

Page 25: ...re Version A01 Device Classification Battery Powered Mode Internally Powered ME Equipment AC Adaptor Powered Mode Class II ME Equipment IP21 It means the device could protected against solid foreign o...

Page 26: ...devices EN 60601 1 2006 A1 2013 IEC 60601 1 2005 A1 2012 Medical electrical equipment Part 1 General requirements for basic safety and essential performance EN 60601 1 11 2015 IEC 60601 1 11 2015 Medi...

Page 27: ...ated non invasive sphygmomanometers ISO 81060 2 2013 Non invasive sphygmomanometers Part 2 Clinical validation of automated measurement type EMC Guidance The ME EQUIPMENT or ME SYSTEM is suitable for...

Page 28: ...gnetic Immunity Immunity Test 8 kV contact 2 kV 4kV 8 kV 15 kV air 8 kV contact 2 kV for power supply lines 0 5 kV 1 kV differential mode Table 2 Compliance level Electrostatic discharge ESD IEC 61000...

Page 29: ...n 0 15 MHz and 80 MHz 80 AM at 1 kHz 3 V 0 15 MHz 80 MHz 6 V in ISM and amateur radio bands between 0 15 MHz and 80 MHz 80 AM at 1 kHz 10 V m 80 MHz 2 7 GHz 80 AM at 1 kHz 10 V m 80 MHz 2 7 GHz 80 AM...

Page 30: ...modulation b 217Hz 2 0 3 28 2 0 3 28 2450 2400 2570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 Pulse modulation 217 Hz 5240 5500 5785 5100 5800 WLAN 802 11 a n Pulse modulation 217 Hz 0 2 0 3 9...

Page 31: ...59 60 WARRANTY WARRANTY...

Page 32: ......

Reviews: